The next generation of fenestrated endografts: results of a clinical trial to support an expanded indication for aortic arch aneurysm treatment

被引:39
作者
Azuma, Takashi [1 ]
Yokoi, Yoshihiko [1 ]
Yamazaki, Kenji [1 ]
机构
[1] Tokyo Womens Med Univ, Dept Cardiovasc Surg, Tokyo 1628666, Japan
关键词
Endograft; Endoleak; Aneurysm; Precurved; Fenestrated; Aortic arch; REPAIR; EXPERIENCE; TEVAR; RISK;
D O I
10.1093/ejcts/ezt241
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
OBJECTIVES: Short- and mid-term data regarding the use of precurved, fenestrated endografts have shown that these devices are both safe and effective in carefully selected patients. The first generation of the product was limited to patients with proximal landing zones (LZs) of >20 mm. The next generation of these endografts has been refined to enable the treatment of patients with shorter proximal seal zones (<20 mm), using smaller fenestrations and a greater diversity of skeletons. We reviewed the clinical studies involving the next-generation product and analysed the morphological characteristics of aortic arch aneurysms that were successfully treated. METHODS: Next-generation endografts were used to treat 393 patients with aortic arch aneurysms at 35 medical institutions during 2010 and 2011. There were 371 (94%) patients with sealing zones <20 mm and 244 (62%) with sealing zones <15 mm. The proximal sealing length was 2-35 (14.2 +/- 5.1) mm. RESULTS: Technical success was achieved in 390 (99.2%) patients. Of the treated patient population, 6 patients died, 7 experienced strokes and 17 were subsequently identified to have Type I endoleaks. In cases with proximal LZs <15 mm, the aneurysm was more likely to develop an endoleak. The proximal sealing zones (11 +/- 12 vs 9 +/- 13 mm) were not significantly associated with the development of endoleaks, but the proximal aortic diameters were (34.0 +/- 13.3 vs 36.6 +/- 6.3 mm; P < 0.01), in the univariate analysis. In the discriminant analysis, the maximum length of the aneurysm was the only factor that was predictive of Type I endoleaks (73 +/- 55 vs 97 +/- 59 mm; P < 0.001). CONCLUSIONS: The next generation of precurved, fenestrated endografts shows promise as devices for aortic arch aneurysms with a <15-mm proximal sealing zone. These devices have a significant advantage in cases where the LZ has a short neck. However, more refinement is necessary to prevent Type I endoleaks, so that these devices can be used with aortic vessels with large proximal diameters and large aneurysms.
引用
收藏
页码:E156 / E163
页数:8
相关论文
共 13 条
[1]   Classic hybrid evolving approach to distal arch aneurysms: Toward the zone zero solution [J].
Bavaria, Joseph ;
Milewski, Rita K. ;
Baker, Joshua ;
Moeller, Patrick ;
Szeto, Wilson ;
Pochettino, Alberto .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2010, 140 (06) :S77-S80
[2]   Aging Impact on Thoracic Aorta 3D Morphometry in Intermediate-Risk Subjects: Looking Beyond Coronary Arteries with Non-Contrast Cardiac CT [J].
Craiem, Damian ;
Chironi, Gilles ;
Redheuil, Alban ;
Casciaro, Mariano ;
Mousseaux, Elie ;
Simon, Alain ;
Armentano, Ricardo L. .
ANNALS OF BIOMEDICAL ENGINEERING, 2012, 40 (05) :1028-1038
[3]  
Czerny M, 2011, CARDIOTHORAC SURG, DOI [10.1510/mmcts.2011.005108, DOI 10.1510/MMCTS.2011.005108]
[4]   TEVAR Using the Redesigned TAG Device (CTAG) For Traumatic Aortic Transection: A Nonrandomized Multicenter Trial [J].
Farber, Mark A. ;
Giglia, Joseph ;
Starnes, Benjamin ;
Stevens, Scott ;
Holleman, Jeremiah ;
Chaer, Rabih ;
Matsumura, Jon .
JOURNAL OF VASCULAR SURGERY, 2012, 55 (02) :622-622
[5]   Fenestrated and branched devices in the pipeline [J].
Greenberg, Roy K. ;
Qureshi, Moqueet .
JOURNAL OF VASCULAR SURGERY, 2010, 52 :15S-21S
[6]  
Kawaguchi S, 1998, Jpn J Thorac Cardiovasc Surg, V46, P971
[7]   Thoracic endovascular aneurysm repair in Japan: Experience with fenestrated stent grafts in the treatment of distal arch aneurysms [J].
Kawaguchi, Satoshi ;
Yokoi, Yoshihiko ;
Shimazaki, Taro ;
Koide, Kenji ;
Matsumoto, Masataka ;
Shigematsu, Hiroshi .
JOURNAL OF VASCULAR SURGERY, 2008, 48 :24S-29S
[8]   First United States experience of the TX2 Pro-Form thoracic delivery system [J].
Lee, W. Anthony ;
Martin, Tomas D. ;
Hess, Philip J., Jr. ;
Beaver, Thomas M. ;
Klodell, Charles T. .
JOURNAL OF VASCULAR SURGERY, 2010, 52 (06) :1459-1463
[9]   Thoracic Endovascular Repair (TEVAR) in the Management of Aortic Arch Pathology [J].
Murphy, Erin H. ;
Stanley, Gregory A. ;
Ilves, Mihaiela ;
Knowles, Martyn ;
Dimaio, J. Michael ;
Jessen, Michael E. ;
Arko, Frank R., III .
ANNALS OF VASCULAR SURGERY, 2012, 26 (01) :55-66
[10]   The Chimney Graft: A Technique for Preserving or Rescuing Aortic Branch Vessels in Stent-Graft Sealing Zones [J].
Ohrlander, Tomas ;
Sonesson, Bjorn ;
Ivancev, Krasnodar ;
Resch, Timothy ;
Dias, Nuno ;
Malina, Martin .
JOURNAL OF ENDOVASCULAR THERAPY, 2008, 15 (04) :427-432