An experimental/numerical method for the efficiency calibration of low-energy germanium detectors

被引:0
|
作者
Anagnostakis, MJ
Simopoulos, SE
机构
关键词
D O I
暂无
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
Several radionuclides emit significant gamma rays at energies below 80 keV, which in many cases are essential for their detection in radioenvironmental assays. Gamma spectroscopic analysis at this energy region is conducted using planar Low Energy Germanium detectors, with high and almost constant efficiency in the energy region between 20 and 80 keV. In the analysis of solid and liquid samples, with these detectors, the count rate is highly affected by the intense serf-absorption of the low energy photons. Thus, the difference in the absorption properties between the calibration source and the sample requires the introduction of an efficiency correction factor. A method applicable for cylindrical geometries was adapted for the determination of the above correction factor using a newly developed experimental-numerical. technique and a Fortran program. This program, using as input the source-to-detector geometry and the values of the linear attenuation coefficient (mu) of both the calibration source and the material to be analysed, calculates the efficiency correction factor. The value of the linear attenuation coefficient (mu) needed for this calculation is being experimentally estimated for each material to be analysed. The technique has been cross-checked using standard materials. According to the results obtained, in the case of surface sail samples, lignite, and fly ash, the values of mu ranges from 0.2 to 0.9 cm(-1). This leads to efficiency correction factors, for the geometry used and a 4M HCl mixed radionuclide calibration source, in the range of 0.5 to 1.2 for the photons emitted by Pb-210 at 46.52 keV and Am-241 at 59.54 keV. The correction factor for 185.99 keV photons is slightly lower than 1.0, even for the most absorbing of the materials analysed. Copyright (C) 1996 Elsevier Science Ltd.
引用
收藏
页码:S93 / S99
页数:7
相关论文
共 50 条
  • [1] ACCURATE EFFICIENCY CALIBRATION AND PROPERTIES OF SEMICONDUCTOR DETECTORS FOR LOW-ENERGY PHOTONS
    HANSEN, JS
    MCGEORGE, JC
    NIX, D
    SCHMIDTO.WD
    UNUS, I
    FINK, RW
    NUCLEAR INSTRUMENTS & METHODS, 1973, 106 (02): : 365 - 379
  • [2] Efficiency calibration of low-energy electron detectors by means of β-ray emitters
    Nuclear Physics Institute, Acad. of Sci. of the Czech Republic, 250 68 Řež near Prague, Czech Republic
    Nucl Instrum Methods Phys Res Sect A, 2 (433-438):
  • [3] Efficiency calibration of low-energy electron detectors by means of β-ray emitters
    Spalek, A
    Dragoun, O
    Rysavy, M
    Dragounová, N
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1999, 438 (2-3) : 433 - 438
  • [4] New Approaches to Low-Energy Calibration of Cryogenic Detectors
    Ghaith, M.
    Rau, W.
    Peterson-Galema, M.
    Di Stefano, P.
    Fascione, E.
    Germond, R.
    Underwood, R.
    JOURNAL OF LOW TEMPERATURE PHYSICS, 2020, 200 (5-6) : 305 - 311
  • [5] New Approaches to Low-Energy Calibration of Cryogenic Detectors
    M. Ghaith
    W. Rau
    M. Peterson-Galema
    P. Di Stefano
    E. Fascione
    R. Germond
    R. Underwood
    Journal of Low Temperature Physics, 2020, 200 : 305 - 311
  • [6] THE APPLICATION OF GERMANIUM DETECTORS FOR INVIVO COUNTING OF LOW-ENERGY PHOTONS
    BOOTH, LF
    GROFF, DW
    BRONSON, FL
    HEALTH PHYSICS, 1987, 52 : S58 - S58
  • [7] A novel iterative energy calibration method for composite germanium detectors
    Pattabiraman, NS
    Chintalapudi, SN
    Ghugre, SS
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2004, 526 (03) : 439 - 446
  • [8] An automatic energy-calibration method for segmented germanium detectors
    Hu, Z
    Glasmacher, T
    Mueller, WF
    Wiedenhöver, I
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2002, 482 (03) : 715 - 719
  • [9] EFFICIENCY CALIBRATION OF GERMANIUM DETECTORS WITH INTERNAL STANDARDS
    FELSMANN, M
    DENK, HJ
    JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY-ARTICLES, 1992, 157 (01): : 47 - 55