Weighing Scale-Based Pulse Transit Time is a Superior Marker of Blood Pressure than Conventional Pulse Arrival Time

被引:115
作者
Martin, Stephanie L. -O. [1 ]
Carek, Andrew M. [2 ]
Kim, Chang-Sei [1 ]
Ashouri, Hazar [2 ]
Inan, Omer T. [2 ]
Hahn, Jin-Oh [1 ]
Mukkamala, Ramakrishna [3 ]
机构
[1] Univ Maryland, Dept Mech Engn, College Pk, MD 20742 USA
[2] Georgia Inst Technol, Sch Elect & Comp Engn, Atlanta, GA 30332 USA
[3] Michigan State Univ, Dept Elect & Comp Engn, E Lansing, MI 48824 USA
基金
美国国家卫生研究院;
关键词
STRESS;
D O I
10.1038/srep39273
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Pulse transit time (PTT) is being widely pursued for cuff-less blood pressure (BP) monitoring. Most efforts have employed the time delay between ECG and finger photoplethysmography (PPG) waveforms as a convenient surrogate of PTT. However, these conventional pulse arrival time (PAT) measurements include the pre-ejection period (PEP) and the time delay through small, muscular arteries and may thus be an unreliable marker of BP. We assessed a bathroom weighing scale-like system for convenient measurement of ballistocardiography and foot PPG waveforms-and thus PTT through larger, more elastic arteries-in terms of its ability to improve tracking of BP in individual subjects. We measured "scale PTT", conventional PAT, and cuff BP in humans during interventions that increased BP but changed PEP and smooth muscle contraction differently. Scale PTT tracked the diastolic BP changes well, with correlation coefficient of -0.80 +/- 0.02 (mean +/- SE) and root-mean-squared-error of 7.6 +/- 0.5 mmHg after a best-case calibration. Conventional PAT was significantly inferior in tracking these changes, with correlation coefficient of -0.60 +/- 0.04 and root-mean-squared-error of 14.6 +/- 1.5 mmHg (p < 0.05). Scale PTT also tracked the systolic BP changes better than conventional PAT but not to an acceptable level. With further development, scale PTT may permit reliable, convenient measurement of BP.
引用
收藏
页数:8
相关论文
共 22 条
[1]   Cardiovascular and neuroendocrine adjustment to public speaking and mental arithmetic stressors [J].
AlAbsi, M ;
Bongard, S ;
Buchanan, T ;
Pincomb, GA ;
Licinio, J ;
Lovallo, WR .
PSYCHOPHYSIOLOGY, 1997, 34 (03) :266-275
[2]  
[Anonymous], EXERCISE HEART
[3]   Unobtrusive Estimation of Cardiac Contractility and Stroke Volume Changes Using Ballistocardiogram Measurements on a High Bandwidth Force Plate [J].
Ashouri, Hazar ;
Orlandic, Lara ;
Inan, Omer T. .
SENSORS, 2016, 16 (06)
[4]  
ATTERHOG JH, 1981, BRIT HEART J, V46, P311
[5]   A NOTE ON THE USE OF THE INTRACLASS CORRELATION-COEFFICIENT IN THE EVALUATION OF AGREEMENT BETWEEN 2 METHODS OF MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
COMPUTERS IN BIOLOGY AND MEDICINE, 1990, 20 (05) :337-340
[6]   A survey on signals and systems in ambulatory blood pressure monitoring using pulse transit time [J].
Buxi, Dilpreet ;
Redoute, Jean-Michel ;
Yuce, Mehmet Rasit .
PHYSIOLOGICAL MEASUREMENT, 2015, 36 (03) :R1-R26
[7]   A Technical Assessment of Pulse Wave Velocity Algorithms Applied to Non-invasive Arterial Waveforms [J].
Gaddum, N. R. ;
Alastruey, J. ;
Beerbaum, P. ;
Chowienczyk, P. ;
Schaeffter, T. .
ANNALS OF BIOMEDICAL ENGINEERING, 2013, 41 (12) :2617-2629
[8]  
Gao M., 2016, IEEE T BIOM IN PRESS
[9]   Comparison of noninvasive pulse transit time estimates as markers of blood pressure using invasive pulse transit time measurements as a reference [J].
Gao, Mingwu ;
Olivier, N. Bari ;
Mukkamala, Ramakrishna .
PHYSIOLOGICAL REPORTS, 2016, 4 (10)
[10]   The cold pressor test for measuring the reactibility of the blood pressure data concerning 571 normal and hypertensive subjects [J].
Hines, EA ;
Brown, GE .
AMERICAN HEART JOURNAL, 1936, 11 :1-9