A Novel Graft Fixation Technique for Anterior Cruciate Ligament Reconstruction Using Hamstring Tendon Grafts

被引:0
作者
Li, Guoan [1 ]
Hosseini, Ali [1 ]
Gadikota, Hemanth [1 ]
Gill, Thomas [1 ]
机构
[1] Harvard Med Sch, Newton Wellesley Hospital, Dept Orthopaed Surg, Orthopaed Biomech Lab, Newton, MA 02462 USA
来源
JOURNAL OF MEDICAL DEVICES-TRANSACTIONS OF THE ASME | 2018年 / 12卷 / 01期
关键词
anterior cruciate ligament (ACL); anatomical reconstruction; pullout strength; knee kinematics; robotic testing system; ACL RECONSTRUCTION; DOUBLE-BUNDLE; SINGLE-BUNDLE; IN-VITRO; STRENGTH; DEVICES; KINEMATICS;
D O I
10.1115/1.4038307
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This study evaluated the biomechanical efficacy of single-tunnel double-bundle anterior cruciate ligament (ACL) reconstruction technique. The graft construct is achieved using a novel fixation device that splits an ACL (SPACL) graft into two bundles, recreating the anteromedial (AM) and posterolateral (PL) bundles for ACL reconstruction. A pullout strength test of the SPACL was performed using a 7-mm bovine digital extensor tendon graft. The capability in restoration of knee kinematics after SPACL reconstruction was investigated using cadaveric human knees on a robotic testing system under an anterior tibial load of 134 N and a simulated quadriceps load of 400 N. The data indicated that the SPACL graft has a pullout strength of 823.7 +/- 172.3 N. Under the 134 N anterior tibial load, the anteroposterior joint laxity had increased constraint using the SPACL reconstruction but not significantly (p>0.05) at all selected flexion angles. Under the 400 N quadriceps load, no significant differences were observed between the anterior tibial translation of intact knee and SPACL conditions at all selected flexion angles, but the SPACL graft induced a significant increase in external tibial rotation compared to the intact knee condition at all selected flexion angles with a maximal external rotation of -3.20 deg +/- 3.6 deg at 90 deg flexion. These data showed that the SPACL technique is equivalent or superior to existing ACL reconstruction techniques in restoration of knee laxity and kinematics. The new SPACL reconstruction technique could provide a valuable alternation to contemporary ACL reconstruction surgery by more closely recreating native ACL kinematics.
引用
收藏
页数:6
相关论文
共 35 条
[1]   Tunnel widening in single- versus double-bundle anterior cruciate ligament reconstructed knees [J].
Aga, Cathrine ;
Wilson, Katharine J. ;
Johansen, Steinar ;
Dornan, Grant ;
La Prade, Robert F. ;
Engebretsen, Lars .
KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2017, 25 (04) :1316-1327
[2]   Mechanical properties of soft tissue femoral fixation devices for anterior cruciate ligament reconstruction [J].
Ahmad, CS ;
Gardner, TR ;
Groh, M ;
Arnouk, J ;
Levine, WN .
AMERICAN JOURNAL OF SPORTS MEDICINE, 2004, 32 (03) :635-640
[3]   MRI Volume and Signal Intensity of ACL Graft Predict Clinical, Functional, and Patient-Oriented Outcome Measures After ACL Reconstruction [J].
Biercevicz, Alison M. ;
Akelman, Matthew R. ;
Fadale, Paul D. ;
Hulstyn, Michael J. ;
Shalvoy, Robert M. ;
Badger, Gary J. ;
Tung, Glenn A. ;
Oksendahl, Heidi L. ;
Fleming, Braden C. .
AMERICAN JOURNAL OF SPORTS MEDICINE, 2015, 43 (03) :693-699
[4]   Indian hedgehog signaling and the role of graft tension in tendon-to-bone healing: Evaluation in a rat ACL reconstruction model [J].
Carbone, Andrew ;
Carballo, Camila ;
Ma, Richard ;
Wang, Hongsheng ;
Deng, Xianghua ;
Dahia, Chitra ;
Rodeo, Scott .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2016, 34 (04) :641-649
[5]  
Carulli C., 2016, ARTHROSCOPY, V25, P1264, DOI DOI 10.1007/s00167-016-4135-9
[6]   Results of ACL reconstruction with tibial Retroscrew fixation: Comparison of clinical outcomes and tibial tunnel widening [J].
Cohen, Steven B. ;
Pandarinath, Rajeev ;
O'Hagan, Thomas ;
Marchetto, Paul A. ;
Hyatt, Adam ;
Wascher, Jocelyn ;
Deluca, Peter F. .
PHYSICIAN AND SPORTSMEDICINE, 2015, 43 (02) :138-142
[7]   The effect of length on the structural properties of an achilles tendon graft as used in posterior cruciate ligament reconstruction [J].
DeFrate, LE ;
van der Ven, A ;
Gill, TJ ;
Li, GA .
AMERICAN JOURNAL OF SPORTS MEDICINE, 2004, 32 (04) :993-997
[8]   Increased platelet concentration does not improve functional graft healing in bio-enhanced ACL reconstruction [J].
Fleming, Braden C. ;
Proffen, Benedikt L. ;
Vavken, Patrick ;
Shalvoy, Matthew R. ;
Machan, Jason T. ;
Murray, Martha M. .
KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2015, 23 (04) :1161-1170
[9]  
Gadikota Hemanth R, 2015, Knee Surg Relat Res, V27, P69, DOI 10.5792/ksrr.2015.27.2.69
[10]   In Vitro and Intraoperative Laxities After Single-Bundle and Double-Bundle Anterior Cruciate Ligament Reconstructions [J].
Gadikota, Hemanth R. ;
Seon, Jong Keun ;
Chen, Chih-Hui ;
Wu, Jia-Lin ;
Gill, Thomas J. ;
Li, Guoan .
ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2011, 27 (06) :849-860