Microfluidic immunosensor for rapid and highly-sensitive salivary cortisol quantification

被引:42
作者
Pinto, V. [1 ]
Sousa, P. [1 ]
Catarino, S. O. [1 ]
Correia-Neves, M. [2 ,3 ]
Minas, G. [1 ]
机构
[1] Univ Minho, Microelectromech Syst Res Unit CMEMS UMinho, Campus Azurem, P-4800058 Guimaraes, Portugal
[2] Sch Med, Life & Hlth Sci Res Inst ICVS, Braga, Portugal
[3] Univ Minho, ICVS 3Bs, PT Govt Associate Lab, Braga, Portugal
关键词
Microfluidic immunosensor; Colorimetric detection; Lab-on-a-chip (LOC); Poly(dimethylsiloxane); Salivary cortisol; LABEL-FREE; STRESS; FABRICATION; HEALTH;
D O I
10.1016/j.bios.2016.11.067
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
This paper presents a novel poly(dimethylsiloxane) (PDMS) microfluidic immunosensor that integrates a complementary metal-oxide-semiconductor (CMOS) optical detection system for a rapid and highly-sensitive quantification of salivary cortisol. The simple and non-invasive method of saliva sampling provides an interesting alternative to the blood, allowing a fast sampling at short intervals, relevant for many clinical diagnostic applications. The developed approach is based on the covalent immobilization of a coating antibody (Ab), a polyclonal anti-IgG, onto a treated PDMS surface. The coating Ab binds the capture Ab, an IgG specific for cortisol, allowing its correct orientation. Horseradish peroxidase (HRP)-labelled cortisol is added to compete with the cortisol in the sample, for the capture Ab binding sites. The HRP-labelled cortisol, bonded to the capture Ab, is measured through the HRP enzyme and the tetramethylbenzidine (TMB) substrate reaction. The cortisol quantification is performed by colorimetric detection of HRP-labelled cortisol, through optical absorption at 450 nm, using a CMOS silicon photodiode as the photodetector. Under the developed optimized conditions presented here, e.g., microfluidic channels geometry, immobilization method and immunoassay conditions, the immunosensor shows a linear range of detection between 0.01-20 ng/mL, a limit of detection (LOD) of 18 pg/mL and an analysis time of 35 min, featuring a great potential for point-of-care applications requiring continuous monitoring of the salivary cortisol levels during a circadian cycle.
引用
收藏
页码:308 / 313
页数:6
相关论文
共 24 条
[1]  
Armbruster David A, 2008, Clin Biochem Rev, V29 Suppl 1, pS49
[2]   Microfluidic immunosensor systems [J].
Bange, A ;
Halsall, HB ;
Heineman, WR .
BIOSENSORS & BIOELECTRONICS, 2005, 20 (12) :2488-2503
[3]  
Ramírez NB, 2009, BRAZ J CHEM ENG, V26, P227, DOI 10.1590/S0104-66322009000200001
[4]   Piezoelectric actuators for acoustic mixing in microfluidic devices-Numerical prediction and experimental validation of heat and mass transport [J].
Catarino, S. O. ;
Silva, L. R. ;
Mendes, P. M. ;
Miranda, J. M. ;
Lanceros-Mendez, S. ;
Minas, G. .
SENSORS AND ACTUATORS B-CHEMICAL, 2014, 205 :206-214
[5]  
Catarino SO, 2016, IEEE ENG MED BIO, P5660, DOI 10.1109/EMBC.2016.7592011
[6]   Application of nanofiber-packed SPE for determination of salivary-free cortisol using fluorescence precolumn derivatization and HPLC detection [J].
Chen, Li-Qin ;
Kang, Xue-Jun ;
Sun, Jing ;
Deng, Jian-Jun ;
Gu, Zhong-Ze ;
Lu, Zu-Hong .
JOURNAL OF SEPARATION SCIENCE, 2010, 33 (15) :2369-2375
[7]   The effects of poly(dimethylsiloxane) surface silanization on the mesenchymal stem cell fate [J].
Chuah, Yon Jin ;
Kuddannaya, Shreyas ;
Lee, Min Hui Adeline ;
Zhang, Yilei ;
Kang, Yuejun .
BIOMATERIALS SCIENCE, 2015, 3 (02) :383-390
[8]   Current Status of Salivary Hormone Analysis [J].
Groeschl, Michael .
CLINICAL CHEMISTRY, 2008, 54 (11) :1759-1769
[9]   Methods for Improving Aptamer Binding Affinity [J].
Hasegawa, Hijiri ;
Savory, Nasa ;
Abe, Koichi ;
Ikebukuro, Kazunori .
MOLECULES, 2016, 21 (04)
[10]   A low cost design and fabrication method for developing a leak proof paper based microfluidic device with customized test zone [J].
Kakoti, Ankana ;
Siddiqui, Mohd Farhan ;
Goswami, Pranab .
BIOMICROFLUIDICS, 2015, 9 (02)