Development and Investigation of a Mesh-Electrode Linear Ion Trap (ME-LIT) Mass Analyzer

被引:15
作者
Wang, Liang [1 ,2 ]
Xu, Fuxing [1 ,2 ]
Dai, Xinhua [3 ]
Fang, Xiang [3 ]
Ding, Chuan-Fan [1 ,2 ]
机构
[1] Fudan Univ, Dept Chem, Shanghai 200433, Peoples R China
[2] Fudan Univ, Laser Chem Inst, Shanghai 200433, Peoples R China
[3] Natl Inst Metrol, Beijing, Peoples R China
关键词
Linear ion trap; Mass analyzer; Mesh electrode; Tandem mass spectrometry; SPECTROMETRY; PERFORMANCE; PRESSURE;
D O I
10.1007/s13361-013-0803-x
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A mesh-electrode linear ion trap (ME-LIT) mass analyzer was developed and its performance was primarily characterized. In conventional linear ion trap mass analyzers, the trapped ions are mass-selected and then ejected in a radial direction by a slot on a trap electrode. The presence of slots can strongly affect the electric field distribution in the ion trapping region and distort the mass analysis performance. To compensate for detrimental electric field effects, the slot is usually designed and fabricated to be as small as possible, and also has very high mechanical accuracy and symmetry. A ME-LIT with several mesh electrodes was built to compensate for the effects caused by slots. Each mesh electrode was fabricated from a plate electrode with a relatively large slot and the slot was covered with a conductive mesh. Our preliminary experimental results show that the ME-LIT could considerably diminish the detrimental electric field effects caused by slots, and increase the mass resolving power and ion detection efficiency. Even with 4-mm-wide slots, a mass resolution in excess of 600 was obtained using the ME-LIT. Mass resolution could be remarkably improved using mesh electrodes in ion traps with asymmetric electrodes. The stability diagram of the ME-LIT was mapped, and highly efficient tandem mass spectrometry was demonstrated. The ME-LIT was qualified as a LIT mass analyzer. The ME-LIT can improve the mass resolution and decrease the requirements of mechanical accuracy and symmetry of slots, so it shows potential for a wide range of practical uses.
引用
收藏
页码:548 / 555
页数:8
相关论文
共 16 条
  • [1] Structure and Performance of Mesh Electrode Ion Trap Mass Analyzer
    Xu Fu-Xing
    Wang Liang
    Wang Yuan-Yuan
    Ding Chuan-Fan
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2013, 41 (05) : 781 - 786
  • [2] Structure and Property of Ladder Electrode Linear Ion Trap Mass Analyzer
    Xie Xiao-Dong
    Xu Fu-Xing
    Yang Hai-Yang
    Xiao Yu
    Jiang Jian
    Ding Chuan-Fan
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2016, 44 (06) : 994 - 1000
  • [3] Miniaturized Linear Wire Ion Trap Mass Analyzer
    Wu, Qinghao
    Li, Ailin
    Tian, Yuan
    Zare, Richard N.
    Austin, Daniel E.
    ANALYTICAL CHEMISTRY, 2016, 88 (15) : 7800 - 7806
  • [4] Simulation of Unidirectional Ion Ejection in an Asymmetric Half-Round Rod Electrode Linear Ion Trap Mass Analyzer
    Wu, HaiYan
    Zhang, LiPeng
    Zhang, ZaiYue
    Qian, Jie
    Zhang, ShuGuang
    Zhang, YingJun
    Ge, SaiJin
    Li, XiaoXu
    JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2018, 29 (10) : 2096 - 2104
  • [5] Development of a portable gas chromatography linear ion trap mass spectrometer (GC-LIT-MS) for VOCs analysis in water
    Qiu, Junwei
    Xu, Kai
    Zhang, Tao
    Zhu, Huijun
    Zhang, Shuo
    Lu, Xinxin
    Li, Xiaoxu
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2024, 497
  • [6] Simulation of double resonant excitation of ions in an asymmetric linear ion trap mass analyzer
    Xu, Kai
    Qiu, Jun Wei
    Zhang, Tao
    Zhu, Hui Jun
    Zhang, Shuo
    Lu, Xin Xin
    Li, Xiao Xu
    RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2023, 37 (23)
  • [7] Optimization of Linear Ion Trap Mass Analyzer with Triangular Electrodes by Simulation
    Yao Ru-Jiao
    He Yang
    Xiao Yu
    Ding Zheng-Zhi
    Yuan Guang-Zhou
    Zhang Zai-Yue
    Qian Jie
    Zhang Shu-Guang
    Li Xiao-Xu
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2016, 44 (11) : 1793 - 1797
  • [8] Effect of Dodecapole Electric Field on Performance of Linear Ion Trap Mass Analyzer
    Xu Fu-Xing
    Jin Liu-Yu
    Qian Bing-Jun
    Wang Wei-Min
    Ding Chuan-Fan
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2021, 49 (08) : 1282 - 1288
  • [9] Design and Performance Evaluation of a Linear Ion Trap Mass Analyzer Featuring Half Round Rod Electrodes
    Li, Xiaoxu
    Zhang, Xiaohua
    Yao, Rujiao
    He, Yang
    Zhu, Yongyong
    Qian, Jie
    JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2015, 26 (05) : 734 - 740
  • [10] Optimization of Resonant Excitation Electric Field in Discrete Electrode Printed Circuit Board Ion Trap Mass Analyzer
    Zhu Wen-Fei
    Wu Hui
    Ge Sai-Jin
    Zhang Ying-Jun
    Yao Ru-Jiao
    Jiang Gong-Yu
    Xiao Yu
    Li Xiao-Xu
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2021, 49 (05) : 693 - 701