Evaluation of a Tactile Display Around the Waist for Physiological Monitoring under Different Clinical Workload Conditions

被引:7
作者
Ng, G. [1 ]
Barralon, P. [1 ]
Schwarz, S. K. W. [2 ]
Dumont, G. [1 ]
Ansermino, J. M. [2 ]
机构
[1] Univ British Columbia, Dept Elect & Comp Engn, Vancouver, BC V5Z 1M9, Canada
[2] Univ British Columbia, Dept Anesthesiol Pharmacol & Therapeut, Vancouver, BC V5Z 1M9, Canada
来源
2008 30TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-8 | 2008年
关键词
D O I
10.1109/IEMBS.2008.4649399
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In this study, we have assessed the usability of a tactile belt prototype for clinical monitoring of physiologic patient data in the operating room under low workload (LW) and high workload (HW) conditions. In previous investigations, we have evaluated tactile technology in clinical settings and demonstrated that anesthesiologists have enhanced situational awareness towards adverse clinical events when a tactile display prototype is used as a supplemental monitoring device. To further evaluate the effectiveness of our tactile belt prototype, we compared the effects of workload on the performance of anesthesiologists in terms of accuracy and response time in tactile alert identification. We also administered a post-study questionnaire to evaluate the usability of the tactile belt as well as users' opinions about the device. We found that the response time to tactile alert identification to be faster under LW than under HW, however the accuracy of identification was not statisticatly different. Participants rated the tactile belt prototype as comfortable to use and the tactile alert scheme as easy to learn. Our findings further support the feasibility and efficacy of vibrotactile devices for enhancing physiological monitoring of patients in clinical environments.
引用
收藏
页码:1288 / +
页数:2
相关论文
共 11 条
[1]  
[Anonymous], 1981, Statistical Tables
[2]  
BARRALON P, 2007, MOBILEHCI 07, P186
[3]  
BROWN LM, 2006, MOBILEHCI 06, P231
[4]  
DANIELS J, 2007, SOC TECHNOLOGY ANEST
[5]  
FINK G, 2003, 1 ANN C SYST INT MAR
[6]   A novel vibrotactile display to improve the performance of anesthesiologists in a simulated critical incident [J].
Ford, Simon ;
Daniels, Jeremy ;
Lim, Joanne ;
Koval, Valentyna ;
Dumont, Guy ;
Schwarz, Stephan K. W. ;
Ansermino, J. Mark .
ANESTHESIA AND ANALGESIA, 2008, 106 (04) :1182-1188
[7]   IBM COMPUTER USABILITY SATISFACTION QUESTIONNAIRES - PSYCHOMETRIC EVALUATION AND INSTRUCTIONS FOR USE [J].
LEWIS, JR .
INTERNATIONAL JOURNAL OF HUMAN-COMPUTER INTERACTION, 1995, 7 (01) :57-78
[8]  
LOEB RG, 1993, ANESTH ANALG, V76, P337
[9]   Response times to visual and auditory alarms during anaesthesia [J].
Morris, RW ;
Montano, SR .
ANAESTHESIA AND INTENSIVE CARE, 1996, 24 (06) :682-684
[10]  
NG G, 2007, EMBC 07, P4204