Surgical Drill Bit Design and Thermomechanical Damage in Bone Drilling: A Review

被引:45
作者
Akhbar, Mohd Faizal Ali [1 ,2 ]
Sulong, Akmal Wani [3 ]
机构
[1] Univ Malaysia Pahang UMP, Coll Engn, Kuantan 26300, Pahang, Malaysia
[2] Univ Malaysia Pahang UMP, Fac Civil Engn Technol FTKA, Kuantan 26300, Pahang, Malaysia
[3] Univ Sultan Zainal Abidin UniSZA, Fac Med, Med Campus,Jalan Sultan Mahmud, Terengganu 20400, Malaysia
关键词
Thermal osteonecrosis; Bone necrosis; Torque; Force; Bone drilling; Bone temperature; Drill bit geometry; Drill bit design; Osteonecrosis region; IMPLANT SITE PREPARATION; CORTICAL BONE; HEAT-PRODUCTION; IN-VITRO; THERMAL-CONDUCTIVITY; TEMPERATURE RISE; STAINLESS-STEEL; OSTEOTOMY PREPARATION; INSTRUMENT BREAKAGE; CUTTING PARAMETERS;
D O I
10.1007/s10439-020-02600-2
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
As drilling generates substantial bone thermomechanical damage due to inappropriate cutting tool selection, researchers have proposed various approaches to mitigate this problem. Among these, improving the drill bit design is one of the most feasible and economical solutions. The theory and applications in drill design have been progressing, and research has been published in various fields. However, pieces of information on drill design are dispersed, and no comprehensive review paper focusing on this topic. Systemizing this information is crucial and, therefore, the impetus of this review. Here, we review not only the state-of-the-art in drill bit designs-advances in surgical drill bit design-but also the influences of each drill bit geometries on bone damage. Also, this work provides future directions for this topic and guidelines for designing an improved surgical drill bit. The information in this paper would be useful as a one-stop document for clinicians, engineers, and researchers who require information related to the tool design in bone drilling surgery.
引用
收藏
页码:29 / 56
页数:28
相关论文
共 199 条
[91]   Thermal Variation During Osteotomy With Different Dental Implant Drills: A Standardized Study in Bovine Ribs [J].
Hochscheidt, Celso Joao ;
Shimizu, Roberto Hideo ;
Andrighetto, Augusto Ricardo ;
Moura, Luis Mauro ;
Golin, Alexsander Luiz ;
Hochscheidt, Regina Celia .
IMPLANT DENTISTRY, 2017, 26 (01) :73-79
[92]  
Huiskes R, 1980, ACTA ORTHOP SCAND, V51, P208
[93]   Osseointegration of alumina with a bioactive coating under load-bearing and unloaded conditions [J].
Ignatius, A ;
Peraus, M ;
Schorlemmer, S ;
Augat, P ;
Burger, W ;
Leyen, S ;
Claes, L .
BIOMATERIALS, 2005, 26 (15) :2325-2332
[94]   Thermal necrosis after simultaneous tibial osteotomy and total knee arthroplasty using press-fit extension-stem [J].
Ishida, Kazunari ;
Tsumura, Nobuhiro ;
Takayama, Koji ;
Fukase, Naomasa ;
Kitagawa, Atsushi ;
Chin, Takaaki ;
Iguchi, Tetsuhiro ;
Matsumoto, Tomoyuki ;
Kubo, Seiji ;
Kurosaka, Masahiro ;
Kuroda, Ryosuke .
KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2011, 19 (01) :112-114
[95]   STUDY OF BONE MACHINING PROCESS-DRILLING [J].
JACOB, CH ;
BERRY, JT ;
POPE, MH ;
HOAGLUND, FT .
JOURNAL OF BIOMECHANICS, 1976, 9 (05) :343-&
[96]   STUDY OF BONE MACHINING PROCESS - ORTHOGONAL CUTTING [J].
JACOBS, CH ;
POPE, MH ;
BERRY, JT ;
HOAGLUND, F .
JOURNAL OF BIOMECHANICS, 1974, 7 (02) :131-&
[97]   Comprehensive analysis on orthopedic drilling: A state-of-the-art review [J].
Jamil, Muhammad ;
Rafique, Saima ;
Khan, Aqib Mashood ;
Hegab, Hussien ;
Mia, Mozammel ;
Gupta, Munish Kumar ;
Song, Qinghua .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2020, 234 (06) :537-561
[98]   Modeling, multi-objective optimization and cost estimation of bone drilling under micro-cooling spray technique: an integrated analysis [J].
Jamil, Muhammad ;
Khan, Aqib Mashood ;
Hegab, Hussien ;
Mia, Mozammel ;
Gupta, Munish Kumar .
INTERNATIONAL JOURNAL OF INTERACTIVE DESIGN AND MANUFACTURING - IJIDEM, 2020, 14 (02) :435-450
[99]   Evaluating the effect of micro-lubrication in orthopedic drilling [J].
Jamil, Muhammad ;
Khan, Aqib Mashood ;
Mia, Mozammel ;
Iqbal, Asif ;
Gupta, Munish Kumar ;
Sen, Binayak .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2019, 233 (10) :1024-1041
[100]   Thermal damage of osteocytes during pig bone drilling: an in vivo comparative study of currently available and modified drills [J].
Kanaya, Haruhisa ;
Enokida, Makoto ;
Uehara, Kazutake ;
Ueki, Masaru ;
Nagashima, Hideki .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2019, 139 (11) :1599-1605