Number of pegs influence focal stress distributions and micromotion in glenoid implants: a finite element study
被引:11
作者:
Wahab, Abdul Hadi Abdul
[1
]
Kadir, Mohammed Rafiq Abdul
[1
]
Harun, Muhammad Noor
[2
]
Kamarul, Tunku
[3
]
Syahrom, Ardiyansyah
[2
]
机构:
[1] Univ Teknol Malaysia, Fac Biosci & Med Engn, Med Devices Technol Grp MediTeg, Skudai 81310, Johor, Malaysia
[2] Univ Teknol Malaysia, IHCE, SITC, Skudai 81310, Johor, Malaysia
[3] Univ Teknol Malaysia, Dept Orthopaed Surg NOCERAL, Tissue Engn Grp, Fac Med, Kuala Lumpur, Malaysia
关键词:
Total shoulder arthroplasty;
Stress critical volume;
Biomechanics;
Orthopedics;
Finite element analysis;
TOTAL SHOULDER ARTHROPLASTY;
CEMENT THICKNESS;
COMPONENTS;
BONE;
PROSTHESIS;
DESIGN;
STABILITY;
OSTEOARTHRITIS;
COMPLICATIONS;
RETROVERSION;
D O I:
10.1007/s11517-016-1525-6
中图分类号:
TP39 [计算机的应用];
学科分类号:
081203 ;
0835 ;
摘要:
The present study was conducted to compare the stability of four commercially available implants by investigating the focal stress distributions and relative micromotion using finite element analysis. Variations in the numbers of pegs between the implant designs were tested. A load of 750 N was applied at three different glenoid positions (SA: superior-anterior; SP: superior-posterior; C: central) to mimic off-center and central loadings during activities of daily living. Focal stress distributions and relative micromotion were measured using Marc Mentat software. The results demonstrated that by increasing the number of pegs from two to five, the total focal stress volumes exceeding 5 MPa, reflecting the stress critical volume (SCV) as the threshold for occurrence of cement microfractures, decreased from 8.41 to 5.21 % in the SA position and from 9.59 to 6.69 % in the SP position. However, in the C position, this change in peg number increased the SCV from 1.37 to 5.86 %. Meanwhile, micromotion appeared to remain within 19-25 A mu m irrespective of the number of pegs used. In conclusion, four-peg glenoid implants provide the best configuration because they had lower SCV values compared with lesser-peg implants, preserved more bone stock, and reduced PMMA cement usage compared with five-peg implants.
引用
收藏
页码:439 / 447
页数:9