Ergonomic Deficits in Robotic Gynecologic Oncology Surgery: A Need for Intervention

被引:59
作者
Craven, Renatta [1 ]
Franasiak, Jason [2 ]
Mosaly, Prithima [3 ]
Gehrig, Paola A. [4 ]
机构
[1] Univ N Carolina, Sch Med, Chapel Hill, NC 27599 USA
[2] Univ N Carolina, Dept Obstet & Gynecol, Chapel Hill, NC 27599 USA
[3] Univ N Carolina, Dept Radiat Oncol, Chapel Hill, NC 27599 USA
[4] Univ N Carolina, Div Gynecol Oncol, Dept Obstet & Gynecol, Chapel Hill, NC 27599 USA
关键词
Ergonomics; Occupational strain; Rapid upper limb assessment; Robotic surgery; Strain Index; STRAIN INDEX; SURGICAL SYSTEM; DISORDERS; WORK; RISK;
D O I
10.1016/j.jmig.2013.04.008
中图分类号
R71 [妇产科学];
学科分类号
100211 ;
摘要
Study Objective: To evaluate surgeon strain using validated ergonomic assessment tools. Design: Observational study (Canadian Task Force classification III). Setting: Academic medical center. Participants: Robotic surgeons performing gynecologic oncology surgical procedures. Interventions: Videotape footage of surgeons performing robotic gynecologic oncology procedures was obtained. A human factors engineer experienced with health care ergonomics analyzed the video recordings and performed ergonomic evaluations of the surgeons. Measurements and Main Results: An initial evaluation was conducted using the Rapid Upper Limb Assessment (RULA) survey, an ergonomic assessment and prioritization method for determining posture, force, and frequency concerns with focus on the upper limbs. A more detailed analysis followed using the Strain Index (SI) method, which uses multiplicative interactions to identify jobs that are potentially hazardous. Seventeen hours of video recordings were analyzed, and descriptive data based on RULA/SI analysis were collected. Ergonomic evaluation of surgeon activity resulted in a mean RULA score of 6.46 (maximum possible RULA score, 7), indicating a need for further investigation. The mean SI grand score was 24.34. SI scores >10 suggest a potential for hazard to the operator. Thus, the current use of the surgical robot is potentially dangerous with regards to ergonomic positioning and should be modified. Conclusion: At a high-volume robotics center, there are ergonomics deficits that are hazardous to gynecologic surgeons and suggest the need for modification and intervention. A training strategy must be developed to address these ergonomic issues and knowledge deficiencies. (C) 2013 AAGL. All rights reserved.
引用
收藏
页码:648 / 655
页数:8
相关论文
共 33 条
[1]   The da Vinci telerobotic surgical system: the virtual operative field and telepresence surgery [J].
Ballantyne, GH ;
Moll, F .
SURGICAL CLINICS OF NORTH AMERICA, 2003, 83 (06) :1293-+
[2]   Surgery and ergonomics [J].
Berguer, R .
ARCHIVES OF SURGERY, 1999, 134 (09) :1011-1016
[3]   Surgical technology and the ergonomics of laparoscopic instruments [J].
Berguer, R .
SURGICAL ENDOSCOPY-ULTRASOUND AND INTERVENTIONAL TECHNIQUES, 1998, 12 (05) :458-462
[4]  
BORG G, 1970, Scandinavian Journal of Rehabilitation Medicine, V2, P92
[5]  
Chaffin D.B., 1991, OCCUPATIONAL BIOMECH
[6]   The history of robotics in urology [J].
Challacombe, Ben J. ;
Khan, Mohammad Shamim ;
Murphy, Declan ;
Dasgupta, Prokar .
WORLD JOURNAL OF UROLOGY, 2006, 24 (02) :120-127
[7]   Developments in robotic cardiac surgery [J].
Falk, V ;
Diegler, A ;
Walther, T ;
Autschbach, R ;
Mohr, FW .
CURRENT OPINION IN CARDIOLOGY, 2000, 15 (06) :378-387
[8]   Physical strain and urgent need for ergonomic training among gynecologic oncologists who perform minimally invasive surgery [J].
Franasiak, Jason ;
Ko, Emily M. ;
Kidd, Juli ;
Secord, Angeles Alvarez ;
Bell, Maria ;
Boggess, John F. ;
Gehrig, Paola A. .
GYNECOLOGIC ONCOLOGY, 2012, 126 (03) :437-442
[9]  
JAMES TM, 1995, HUM FAC ERG SOC P, P573
[10]  
Lawson EH, 2007, J ROBOT SURG, V1, P61, DOI 10.1007/s11701-007-0016-z