Validation of the iHealth BP3 upper-arm blood pressure monitor, for clinic use and self-measurement, according to the European Society of Hypertension International Protocol revision 2010

被引:1
作者
Chen, Can [2 ]
Shang, Fujun [3 ]
Wang, Jiepin [4 ]
Chen, Jianghong [3 ]
Ji, Na [3 ]
Wan, Yi [1 ]
机构
[1] Fourth Mil Med Univ, Sch Publ Hlth, Dept Hlth Stat, Xian 710032, Peoples R China
[2] Guangdong Med Coll, Affiliated Hosp, Dept Cardiol, Zhanjiang, Peoples R China
[3] Fourth Mil Med Univ, Tangdu Hosp, Dept Cardiol, Xian 710032, Peoples R China
[4] Fourth Mil Med Univ, Tangdu Hosp, Dept Pharm, Xian 710032, Peoples R China
关键词
blood pressure; European Society of Hypertension; measurement; validation; MEASURING DEVICES; ADULTS;
D O I
10.1097/MBP.0b013e32835b9ecb
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Objective This study aimed to evaluate the performance of the iHealth BP3 upper-arm blood pressure monitor, which is designed for clinic use and self-measurement of blood pressure using Apple touch devices as an interface. Methods The European Society of Hypertension International Protocol (ESH-IP) revision 2010 for the validation of blood pressure measuring devices in adults was followed precisely. Ninty-nine couples of test device and reference blood pressure measurements were obtained during the study (three pairs for each of the 33 participants). Results The 33 participants, age 47.1 +/- 12.3 years (age range 27-69 years) and arm circumference 30.0 +/- 4.4 cm, had a mean systolic blood pressure (SBP) of 143.9 +/- 27.4 mmHg and a mean diastolic blood pressure (DBP) of 90.1 +/- 18.3 mmHg. The device passed all of the requirements fulfilling the standards of the protocol, and the mean +/- SD device-observer difference was 2.8 +/- 4.2 mmHg for SBP and -0.4 +/- 3.5 mmHg for DBP. Conclusion According to the results of the validation study on the basis of the ESH-IP revision 2010, the iHealth BP3 can be recommended for clinic use and self-measurement in an adult population. Blood Press Monit 17:253-256 (c) 2012 Wolters Kluwer Health vertical bar Lippincott Williams & Wilkins.
引用
收藏
页码:253 / 256
页数:4
相关论文
共 6 条
[1]   Does self-monitoring reduce blood pressure? Meta-analysis with meta-regression of randomized controlled trials [J].
Bray, Emma P. ;
Holder, Roger ;
Mant, Jonathan ;
McManus, Richard J. .
ANNALS OF MEDICINE, 2010, 42 (05) :371-386
[2]   Blood pressure self monitoring: questions and answers from a national conference [J].
McManus, Richard J. ;
Glasziou, Paul ;
Hayen, Andrew ;
Mant, Jonathan ;
Padfield, Paul ;
Potter, John ;
Bray, Emma P. ;
Mant, David .
BRITISH MEDICAL JOURNAL, 2008, 337 :38-42
[3]   European Society of hypertension recommendations for conventional, ambulatory and home blood pressure measurement [J].
O'Brien, E ;
Asmar, R ;
Beilin, L ;
Imai, Y ;
Mallion, JM ;
Mancia, G ;
Mengden, T ;
Myers, M ;
Padfield, P ;
Palatini, P ;
Parati, G ;
Pickering, T ;
Redon, J ;
Staessen, J ;
Stergiou, G ;
Verdecchia, P .
JOURNAL OF HYPERTENSION, 2003, 21 (05) :821-848
[4]   Working Group on Blood Pressure Monitoring of the European Society of Hypertension International Protocol for validation of blood pressure measuring devices in adults [J].
O'Brien, E ;
Pickering, T ;
Asmar, R ;
Myers, M ;
Parati, G ;
Staessen, J ;
Mengden, T ;
Imai, Y ;
Waeber, B ;
Palatini, P ;
Atkins, N ;
Gerin, W .
BLOOD PRESSURE MONITORING, 2002, 7 (01) :3-17
[5]   European Society of Hypertension International Protocol revision 2010 for the validation of blood pressure measuring devices in adults [J].
O'Brien, Eoin ;
Atkins, Neil ;
Stergiou, George ;
Karpettas, Nikos ;
Parati, Gianfranco ;
Asmar, Roland ;
Imai, Yutaka ;
Wang, Jiguang ;
Mengden, Thomas ;
Shennan, Andrew .
BLOOD PRESSURE MONITORING, 2010, 15 (01) :23-38
[6]   European Society of Hypertension Practice Guidelines for home blood pressure monitoring [J].
Parati, G. ;
Stergiou, G. S. ;
Asmar, R. ;
Bilo, G. ;
de Leeuw, P. ;
Imai, Y. ;
Kario, K. ;
Lurbe, E. ;
Manolis, A. ;
Mengden, T. ;
O'Brien, E. ;
Ohkubo, T. ;
Padfield, P. ;
Palatini, P. ;
Pickering, T. G. ;
Redon, J. ;
Revera, M. ;
Ruilope, L. M. ;
Shennan, A. ;
Staessen, J. A. ;
Tisler, A. ;
Waeber, B. ;
Zanchetti, A. ;
Mancia, G. .
JOURNAL OF HUMAN HYPERTENSION, 2010, 24 (12) :779-785