Update: Toward image-guided robotic surgery: determining the intrinsic accuracy of the daVinci-S robot

被引:27
|
作者
Kwartowitz, David M. [1 ]
Herrell, S. Duke [2 ]
Galloway, Robert L. [1 ,3 ,4 ]
机构
[1] Vanderbilt Univ, Dept Biomed Engn, Nashville, TN 37235 USA
[2] Vanderbilt Univ, Dept Urol, Nashville, TN 37235 USA
[3] Vanderbilt Univ, Dept Surg, Nashville, TN 37240 USA
[4] Vanderbilt Univ, Dept Neurol Surg, Nashville, TN 37235 USA
关键词
Robot; Image-guided surgery; Image registration; Surgery; Robotic assisted surgery;
D O I
10.1007/s11548-006-0064-2
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Early in 2006, Intuitive Surgical (Sunnyvale, CA, USA) released their next generation robotic surgery system, the daVinci-S. While the S is an update on the da Vinci surgical system it has significantly different structures. This led to the belief that there might be differences in the accuracy and precision of the daVinci-S as compared to the daVinci "classic". In a previous study, we measured the accuracy of the daVinci "classic" to be 1.02 mm throughout its work volume (Kwartowitz et al. in Int J Comput Assist Radiol Surg 1: 159-165, 2006) In this study we evaluated the accuracy of the daVinci-S using the same protocol as was used for the daVinci "classic". The localization error is found to be 1.05 mm, which is essentially the same as the measured error for the daVinci "classic". It is concluded that the daVinci-S is an appropriate localizer for a robotic image guided surgery (RIGS) system.
引用
收藏
页码:301 / 304
页数:4
相关论文
共 46 条
  • [1] Update: Toward image-guided robotic surgery: determining the intrinsic accuracy of the daVinci-S robot
    David M. Kwartowitz
    S. Duke Herrell
    Robert L. Galloway
    International Journal of Computer Assisted Radiology and Surgery, 2007, 1 : 301 - 304
  • [2] Toward image-guided robotic surgery: determining intrinsic accuracy of the daVinci robot
    Kwartowitz, David M.
    Herrell, S. Duke
    Galloway, Robert L.
    INTERNATIONAL JOURNAL OF COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2006, 1 (03) : 157 - 165
  • [3] Toward image-guided robotic surgery: determining intrinsic accuracy of the da Vinci robot
    David M. Kwartowitz
    S. Duke Herrell
    Robert L. Galloway
    International Journal of Computer Assisted Radiology and Surgery, 2006, 1 : 157 - 165
  • [4] The Rise of Molecular Image-Guided Robotic Surgery
    van Leeuwen, Fijs W. B.
    Buckle, Tessa
    van Oosterom, Matthias N.
    Rietbergen, Daphne D. D.
    JOURNAL OF NUCLEAR MEDICINE, 2024, 65 (10) : 1505 - 1511
  • [5] Augmented Reality and Image-Guided Robotic Liver Surgery
    Giannone, Fabio
    Felli, Emanuele
    Cherkaoui, Zineb
    Mascagni, Pietro
    Pessaux, Patrick
    CANCERS, 2021, 13 (24)
  • [6] An image-guided robotic surgery system for spinal fusion
    Chung, GB
    Kim, S
    Lee, SG
    Yi, BJ
    Kim, W
    Oh, SM
    Kim, YS
    So, BR
    Park, JI
    Oh, SH
    INTERNATIONAL JOURNAL OF CONTROL AUTOMATION AND SYSTEMS, 2006, 4 (01) : 30 - 41
  • [7] Robot Ontologies for Sensor- and Image-Guided Surgery
    Haidegger, Tamas
    Barreto, Marcos
    Goncalves, Paulo J. S.
    Habib, Maki K.
    Ragavan, S. Veera
    Li, Howard
    Vaccarella, Alberto
    Perrone, Roberta
    Prestes, Edson
    2013 IEEE INTERNATIONAL SYMPOSIUM ON ROBOTIC AND SENSORS ENVIRONMENTS (ROSE 2013), 2013,
  • [8] Design and Accuracy Assessment of an Automated Image-Guided Robotic Osteotomy System
    Bhagvath, Prathamesh V.
    Mercier, Philippe
    Hall, Andrew F.
    IEEE TRANSACTIONS ON MEDICAL ROBOTICS AND BIONICS, 2024, 6 (01): : 96 - 109
  • [9] Toward Image Guided Robotic Surgery: System Validation
    Herrell, Stanley D.
    Kwartowitz, David Morgan
    Milhoua, Paul M.
    Galloway, Robert L.
    JOURNAL OF UROLOGY, 2009, 181 (02) : 783 - 789
  • [10] Technical advances toward interactive image-guided laparoscopic surgery
    Herline, A
    Stefansic, JD
    Debelak, J
    Galloway, RL
    Chapman, WC
    SURGICAL ENDOSCOPY-ULTRASOUND AND INTERVENTIONAL TECHNIQUES, 2000, 14 (07): : 675 - 679