Evaluation of a real-time blood glucose monitor in children with diabetic ketoacidosis

被引:10
作者
Coss-Bu, JA [1 ]
Jefferson, LS [1 ]
Stone-McCord, S [1 ]
Ou, CN [1 ]
Watrin, C [1 ]
Sachdeva, R [1 ]
Copeland, KC [1 ]
机构
[1] Baylor Coll Med, Dept Pediat, Houston, TX 77030 USA
关键词
glucose monitoring; ketoacidosis; pediatrics; intensive care;
D O I
10.1016/S0168-8227(99)00033-9
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Use of a real-time bedside glucose monitor was analyzed during the course of management of diabetic ketoacidosis (DKA) in children. Simultaneous determinations of blood glucose were obtained, using three methods: bedside glucose meter (One Touch(R) II), laboratory glucose analyzer (YSI 2300 STAT), and a real-time bedside glucose monitor (MA 1-01G Blood Chemistry monitor). Study patients included seventeen patients <18 years of age admitted to a Pediatric Intensive Care Unit, with blood samples obtained during treatment of DKA by continuous insulin infusion. Four patients did not complete the study. Three experienced temporary technical problems with the monitor, and four required repeat IV placement. Duration of monitor use ranged between 6 and 47 h (mean 24 +/- 4 h). Blood glucose values ranged between 2.6 and 22.5 mmol/l. Overall correlation of blood glucose values were as follows: 0.965, 0.965, 0.973, VIA 1-01G vs. One Touch(R) II, VIA 1-01G vs. YSI 2300 STAT, and One Touch(R) II vs. YSI 2300 STAT, respectively (all P-values < 0.0001). This real-time bedside glucose monitor is accurate at glucose values < 13.8 mmol/l, and reliable for rapid, repetitive analyses. Results indicate that blood glucose values obtained using this real-time monitor are comparable to those using standard methods of measurement, and that this device is clinically applicable for use in management of children with DKA. (C) 1999 Published by Elsevier Science Ireland Ltd. All rights reserved.
引用
收藏
页码:175 / 181
页数:7
相关论文
共 28 条
[1]  
Altimier L B, 1996, Neonatal Netw, V15, P15
[2]  
[Anonymous], 1987, DIABETES CARE, V10, P95
[3]  
Blackett P R, 1990, J Okla State Med Assoc, V83, P594
[4]  
CARAWAY WTW, 1986, TXB CLIN CHEM
[5]   BEDSIDE BLOOD-GLUCOSE DETERMINATIONS IN CRITICAL CARE MEDICINE - A COMPARATIVE-ANALYSIS OF 2 TECHNIQUES [J].
CHERNOW, B ;
DIAZ, M ;
CRUESS, D ;
BALESTRIERI, F ;
UDDIN, D ;
RAINEY, TG ;
OBRIAN, JT .
CRITICAL CARE MEDICINE, 1982, 10 (07) :463-465
[6]   EVALUATING CLINICAL ACCURACY OF SYSTEMS FOR SELF-MONITORING OF BLOOD-GLUCOSE [J].
CLARKE, WL ;
COX, D ;
GONDERFREDERICK, LA ;
CARTER, W ;
POHL, SL .
DIABETES CARE, 1987, 10 (05) :622-628
[7]  
Edge JA, 1996, BRIT J HOSP MED, V55, P508
[8]   DIRECT READING GLUCOSE ELECTRODES DETECT THE MOLALITY OF GLUCOSE IN PLASMA AND WHOLE-BLOOD [J].
FOGHANDERSEN, N ;
WIMBERLEY, PD ;
THODE, J ;
SIGGAARDANDERSEN, O .
CLINICA CHIMICA ACTA, 1990, 189 (01) :33-38
[9]  
GINSBERG BH, 1992, CLIN CHEM, V38, P1596
[10]   A PRECISION INDEX FOR EVALUATION OF TECHNIQUES FOR SELF-MONITORING OF BLOOD-GLUCOSE [J].
HALLE, JP ;
LAMBERT, J ;
RENAUD, E ;
TOUPIN, A ;
LEGAULT, C ;
LALUMIERE, G .
DIABETES CARE, 1986, 9 (03) :319-320