Comparison of the Nellcor N-200 and N-3000 pulse oximeters during simulated postoperative activities

被引:9
作者
Lie, C [1 ]
Kehlet, H [1 ]
Rosenberg, J [1 ]
机构
[1] HVIDOVRE UNIV HOSP,DEPT SURG GASTROENTEROL 235,DK-2650 HVIDOVRE,DENMARK
关键词
equipment; pulse oximeter; hypoxia;
D O I
10.1111/j.1365-2044.1997.027-az026.x
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Twenty-six healthy volunteers were monitored simultaneously with the Nellcor N-200 and N-3000 pulse oximeters during nonhypoxaemic simulated postoperative activity The overall number of registered events (hypoxaemic episodes or loss of signal) was fewer with the N-3000 than with the N-200 (8 vs. 32, p < 0.00005). Episodes of 'desaturation' of greater than or equal to 5% from baseline were significantly fewer with the N-3000 than with the N-200 (5 vs. 19, p = 0.0001), and lowest values below 90% occurred nine times on the N-200, but were not seen with the N-3000 (p < 0.00005). Furthermore, episodes owing to loss of signal were significantly rarer with the N-3000 than with the N-200 (3 vs. 13, p = 0.001). The Nellcor N-3000 oximeter may offer an advantage over the N-200 model when monitoring patients in the postoperative period.
引用
收藏
页码:450 / 452
页数:3
相关论文
共 8 条
[1]  
BARKER SJ, 1995, ANESTHESIOLOGY, V83, pA450
[2]   EFFECT OF MOTION ARTIFACT ON PULSE OXIMETERS - EVALUATION OF 4 INSTRUMENTS AND FINGER PROBES [J].
LANGTON, JA ;
HANNING, CD .
BRITISH JOURNAL OF ANAESTHESIA, 1990, 65 (04) :564-570
[3]   CRYING WOLF - FALSE ALARMS IN A PEDIATRIC INTENSIVE-CARE UNIT [J].
LAWLESS, ST .
CRITICAL CARE MEDICINE, 1994, 22 (06) :981-985
[4]  
*NELLC PUR BENN IN, 1994, 8 NELLC PUR BENN INC, P1
[5]   EVALUATION OF THE INFLUENCE OF MOVEMENT ON SATURATION READINGS FROM PULSE OXIMETERS [J].
PLUMMER, JL ;
ZAKARIA, AZ ;
ILSLEY, AH ;
FRONSKO, RRL ;
OWEN, H .
ANAESTHESIA, 1995, 50 (05) :423-426
[6]  
ROSENBERG J, 1994, EUR J SURG, V160, P657
[7]  
Stausholm K, 1994, Ugeskr Laeger, V156, P6675
[8]  
TAKAKI O, 1996, JAPANESE J CLIN MO S, V7