Surface profiles of new-generation IOLs with improved intermediate vision

被引:48
作者
Tognetto, Daniele [1 ]
Cecchini, Paolo [1 ]
Giglio, Rosa [1 ]
Turco, Gianluca [1 ]
机构
[1] Univ Trieste, Dept Med Surg & Hlth Sci, Trieste, Italy
关键词
FOCUS INTRAOCULAR-LENS; EXTENDED DEPTH; VISUAL OUTCOMES; PERFORMANCE; IMPLANTATION; DISTANCE; EYES;
D O I
10.1097/j.jcrs.0000000000000215
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To evaluate the surface profile of different new-generation intraocular lenses (IOLs) designed to improve intermediate vision. Setting: Department of Medicine, Surgery and Health Sciences, University of Trieste, Trieste, Italy. Design: Prospective experimental laboratory study. Methods: One higher-order aspheric monofocal IOL (Tecnis Eyhance ICB00) and 2 extended depth-of-focus (EDOF) IOLs (Tecnis Symfony ZXR00, Mini Well Ready) were examined and compared with monofocal IOLs with the same platform (Sensar AAB00, Tecnis ZCB00, Mini 4 Ready) using contact profilometry. Raw profiles were obtained and compared. The best-fit circle was then subtracted from each raw profile to highlight differences. Results: The new higher-order aspheric IOL presented a peculiar profile, which appeared similar to the conventional monofocal IOL based on the same platform, with just minimal differences in the central design. One of the EDOF IOLs showed a diffractive design. The EDOF IOL based on spherical aberration presented a central steepening surrounded by a symmetric lateral change in depths. Conclusions: Improved intermediate vision can be achieved by means of different optic designs. Unlike the traditional EDOF and multifocal IOLs based on a diffractive design, the profiles of new-generation IOLs are characterized by smooth and progressive changes of the superficial geometry. Copyright (C) 2020 Published by Wolters Kluwer on behalf of ASCRS and ESCRS.
引用
收藏
页码:902 / 906
页数:5
相关论文
共 30 条
[1]   Extended depth of focus intraocular lenses for presbyopia [J].
Akella, Sruti S. ;
Juthani, Viral V. .
CURRENT OPINION IN OPHTHALMOLOGY, 2018, 29 (04) :319-323
[2]   Multifocal intraocular lenses: An overview [J].
Alio, Jorge L. ;
Plaza-Puche, Ana B. ;
Fernandez-Buenaga, Roberto ;
Pikkel, Joseph ;
Maldonado, Miguel .
SURVEY OF OPHTHALMOLOGY, 2017, 62 (05) :611-634
[3]   Visual Strehl Performance of IOL Designs with Extended Depth of Focus [J].
Arias Gallego, Augusto ;
Bara, Salvador ;
Jaroszewicz, Zbigniew ;
Kolodziejczyk, Andrzej .
OPTOMETRY AND VISION SCIENCE, 2012, 89 (12) :1702-1707
[4]  
Bellucci Roberto, 2005, Curr Opin Ophthalmol, V16, P33, DOI 10.1097/00055735-200502000-00006
[5]   Clinical and aberrometric evaluation of a new extended depth-of-focus intraocular lens based on spherical aberration [J].
Bellucci, Roberto ;
Cargnoni, Miriam ;
Bellucci, Carlo .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2019, 45 (07) :919-926
[6]   A New Extended Depth of Focus Intraocular Lens Based on Spherical Aberration [J].
Bellucci, Roberto ;
Curatolo, Maria C. .
JOURNAL OF REFRACTIVE SURGERY, 2017, 33 (06) :389-+
[7]   Chemical and physical analysis of phaco handpiece tip surfaces before and after cataract surgery [J].
Cecchini, Paolo ;
D'Aloisio, Rossella ;
Antonuccio, Manlio ;
Turco, Gianluca ;
Bondino, Federica ;
Magnano, Elena ;
Di Nicola, Marta ;
Tognetto, Daniele .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2017, 43 (08) :1107-1114
[8]   A Comparative Evaluation of a New Generation of Diffractive Trifocal and Extended Depth of Focus Intraocular Lenses [J].
Cochener, Beatrice ;
Boutillier, Guillaume ;
Lamard, Mathieu ;
Auberger-Zagnoli, Claire .
JOURNAL OF REFRACTIVE SURGERY, 2018, 34 (08) :507-+
[9]  
de Medeiros AL, 2017, CLIN OPHTHALMOL, V11, P1911, DOI 10.2147/OPTH.S145945
[10]   Dissatisfaction after implantation of multifocal intraocular lenses [J].
de Vries, Niels E. ;
Webers, Carroll A. B. ;
Touwslager, Wouter R. H. ;
Bauer, Noel J. C. ;
de Brabander, John ;
Berendschot, Tos T. ;
Nuijts, Rudy M. M. A. .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2011, 37 (05) :859-865