Metal Ion Levels Not Sufficient as a Screening Measure for Adverse Reactions in Metal-on-Metal Hip Arthroplasties

被引:53
作者
Macnair, Rory D. [1 ]
Wynn-Jones, Henry [2 ]
Wimhurst, James A. [3 ]
Toms, Andoni [3 ]
Cahir, John [3 ]
机构
[1] Royal Cornwall Hosp, Dept Orthopaed Surg, Truro TR1 3LJ, Cornwall, England
[2] Wrightington Hosp, Ctr Hip Surg, Wigan, Lancs, England
[3] Norfolk & Norwich Univ Hosp, Norwich, Norfolk, England
关键词
metal ion levels; MAR MRI; metal-on-metal; hip; arthroplasty; RESURFACING ARTHROPLASTY; SURFACE REPLACEMENT; EARLY FAILURE; CHROMIUM; COBALT; INCLINATION; COMPONENT; MRI;
D O I
10.1016/j.arth.2012.05.029
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
This study aims to assess the accuracy of metal ion analysis in the diagnosis of adverse reaction to metal debris (ARMD) in patients with metal-on-metal hip arthroplasties by comparing the cobalt and chromium levels in 57 patients (62 hips) to findings on metal artifact reduction magnetic resonance imaging (MRI). An ARMD was detected using MRI in 18 (29%) of the hips. Forty patients had cobalt levels less than 7 mu g/L, and 33 had chromium levels less than 7 mu g/L, but 8 of these had an ARMD on MRI and only minimal symptoms (Oxford Hip Score >= 44/48). The incidence of ARMD was significantly higher when chromium concentration was above 7 mu g/L (P=.02), but normal metal ion levels can be misleading and metal artifact reduction MRI imaging is advised in all patients.
引用
收藏
页码:78 / 83
页数:6
相关论文
共 30 条
[1]   TREATMENT OF PRIMARY OSTEOARTHRITIS OF THE HIP - A COMPARISON OF TOTAL JOINT AND SURFACE REPLACEMENT ARTHROPLASTY [J].
AMSTUTZ, HC ;
THOMAS, BJ ;
JINNAH, R ;
KIM, W ;
GROGAN, T ;
YALE, C .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1984, 66A (02) :228-241
[2]   Grading the severity of soft tissue changes associated with metal-on-metal hip replacements: reliability of an MR grading system [J].
Anderson, Helen ;
Toms, Andoni Paul ;
Cahir, John G. ;
Goodwin, Richard W. ;
Wimhurst, James ;
Nolan, John F. .
SKELETAL RADIOLOGY, 2011, 40 (03) :303-307
[3]  
[Anonymous], MED DEVICE ALERT ALL
[4]   High failure rates with a large-diameter hybrid metal-on-metal total hip replacement CLINICAL, RADIOLOGICAL AND RETRIEVAL ANALYSIS [J].
Bolland, B. J. R. F. ;
Culliford, D. J. ;
Langton, D. J. ;
Millington, J. P. S. ;
Arden, N. K. ;
Latham, J. M. .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2011, 93B (05) :608-615
[5]   CT and MRI of hip arthroplasty [J].
Cahir, J. G. ;
Toms, A. P. ;
Marshall, T. J. ;
Wimhurst, J. ;
Nolan, J. .
CLINICAL RADIOLOGY, 2007, 62 (12) :1163-1171
[6]   The validity of serum levels as a surrogate measure of systemic exposure to metal ions in hip replacement [J].
Daniel, J. ;
Ziaee, H. ;
Pynsent, P. B. ;
McMinn, D. J. W. .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2007, 89B (06) :736-741
[7]   Correlation between inclination of the acetabular component and metal ion levels in metal-on-metal hip resurfacing replacement [J].
De Haan, R. ;
Pattyn, C. ;
Gill, H. S. ;
Murray, D. W. ;
Campbell, P. A. ;
De Smet, K. .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2008, 90B (10) :1291-1297
[8]   Metal Ion Measurement as a Diagnostic Tool to Identify Problems with Metal-on-Metal Hip Resurfacing [J].
de Smet, K. ;
de Haan, R. ;
Calistri, A. ;
Campbell, P. A. ;
Ebramzadeh, E. ;
Pattyn, C. ;
Gill, H. S. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2008, 90A :202-208
[9]   Early failure of the Ultima metal-on-metal total hip replacement in the presence of normal plain radiographs [J].
Donell, S. T. ;
Darrah, C. ;
Nolan, J. F. ;
Wimhurst, J. ;
Toms, A. ;
Barker, T. H. W. ;
Case, C. P. ;
Tucker, J. K. .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2010, 92B (11) :1501-1508
[10]   The Metasul metal-on-metal articulation in primary total hip replacement: Clinical and radiological results at ten years [J].
Eswaramoorthy, V. ;
Moonot, P. ;
Kalairajah, Y. ;
Biant, L. C. ;
Field, R. E. .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2008, 90B (10) :1278-1283