Evaluation of clip localization for different kilovoltage imaging modalities as applied to partial breast irradiation setup

被引:12
作者
Buehler, Andreas [1 ]
Ng, Sook-Kien [1 ]
Lyatskaya, Yulia [2 ,3 ,4 ]
Stsepankou, Dzmitry [1 ]
Hesser, Jurgen
Zygmanski, Piotr [3 ,4 ]
机构
[1] Univ Heidelberg, Univ Med Ctr Mannheim, D-68167 Mannheim, Germany
[2] Dept Radiat Oncol, Boston, MA 02115 USA
[3] Brigham & Womens Hosp, Boston, MA 02115 USA
[4] Harvard Univ, Sch Med, Boston, MA 02115 USA
关键词
biological organs; computerised tomography; image reconstruction; iterative methods; mammography; medical image processing; phantoms; surgery; CONE-BEAM CT; RADIATION-THERAPY; SURGICAL CLIPS; DIGITAL TOMOSYNTHESIS; COMPUTED-TOMOGRAPHY; LUMPECTOMY CAVITY; VERIFICATION; TARGET;
D O I
10.1118/1.3075904
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Surgical clip localization and image quality were evaluated for different types of kilovoltage cone beam imaging modalities as applied to partial breast irradiation (PBI) setup. These modalities included (i) clinically available radiographs and cone beam CT (CB-CT) and (ii) various alternative modalities based on partial/sparse/truncated CB-CT. An anthropomorphic torso-breast phantom with surgical clips was used for the imaging studies. The torso phantom had artificial lungs, and the attached breast phantom was a mammographic phantom with realistic shape and tissue inhomogeneities. Three types of clips of variable size were used in two orthogonal orientations to assess their in-/cross-plane characteristics for image-guided setup of the torso-breast phantom in supine position. All studies were performed with the Varian on-board imaging (OBI (TM), Varian) system. CT reconstructions were calculated with the standard Feldkamp-Davis-Kress algorithm. First, the radiographs were studied for a wide range of viewing angles to characterize image quality for various types of body anatomy in the foreground/background of the clips. Next, image reconstruction quality was evaluated for partial/sparse/truncated CB-CT. Since these modalities led to reconstructions with strong artifacts due to insufficient input data, a knowledge-based CT reconstruction method was also tested. In this method, the input data to the reconstruction algorithm were modified by combining complementary data sets selected from the treatment and reference projections. Different partial/sparse/truncated CB-CT scan types were studied depending on the total arc angle, angular increment between the consequent views (CT projections), orientation of the arc center with respect to the imaged breast and chest wall, and imaging field size. The central angles of the viewing arcs were either tangential or orthogonal to the chest wall. Several offset positions of the phantom with respect to the reference position were studied. The acquired and reconstructed image data sets were analyzed using home-built software focusing on the ability to localize clips in 3D. Streaking and leakage reconstruction artifacts and spatial distortions of breast surface were analyzed as well. Advantages and disadvantages of each kilovoltage CB imaging modality as applied to partial breast setup evaluation based on clips are presented. Because clips were found to be difficult to recognize in radiographs, 3D reconstructions were preferred. Even though it was possible to localize clips with about +/- 1 mm accuracy based on reconstructions for short arcs of 40 degrees and incremental angle up to about 5 degrees, identification of clips in such reconstructions is difficult. Reconstructions obtained for arcs of as low as 80 degrees and incremental angle of as high as 3 degrees were suggested for easier clip identification. For more severely undersampled data, iterative CB-CT reconstruction is recommended to decrease the artifacts.
引用
收藏
页码:821 / 834
页数:14
相关论文
共 21 条
  • [1] Accelerated partial breast irradiation using 3D conformal radiation therapy (3D-CRT)
    Baglan, KL
    Sharpe, MB
    Jaffray, D
    Frazier, RC
    Fayad, J
    Kestin, LL
    Remouchamps, V
    Martinez, AA
    Wong, J
    Vicini, FA
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2003, 55 (02): : 302 - 311
  • [2] Accurate patient dosimetry of kilovoltage cone-beam CT in radiation therapy
    Ding, George X.
    Duggan, Dennis M.
    Coffey, Charles W.
    [J]. MEDICAL PHYSICS, 2008, 35 (03) : 1135 - 1144
  • [3] PRACTICAL CONE-BEAM ALGORITHM
    FELDKAMP, LA
    DAVIS, LC
    KRESS, JW
    [J]. JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1984, 1 (06): : 612 - 619
  • [4] Prone accelerated partial breast irradiation after breast-conserving surgery: Preliminary clinical results and dose-volume histogram analysis
    Formenti, SC
    Truong, MT
    Goldberg, JD
    Mukhi, V
    Rosenstein, B
    Roses, D
    Shapiro, R
    Guth, A
    Dewyngaert, JK
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 60 (02): : 493 - 504
  • [5] FULLER CD, 2006, US ONCOLOGICAL DIS, P1
  • [6] Comparison of target registration errors for multiple image-guided techniques in accelerated partial breast irradiation
    Gierga, David P.
    Riboldi, Marco
    Turcotte, Julie C.
    Sharp, Greg C.
    Jiang, Steve B.
    Taghian, Alphonse G.
    Chen, George T. Y.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 70 (04): : 1239 - 1246
  • [7] Digital tomosynthesis with an on-board kilovoltage imaging device
    Godfrey, DJ
    Yin, FF
    Oldham, M
    Yoo, S
    Willett, C
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2006, 65 (01): : 8 - 15
  • [8] Gonzales R., 2007, DIGITAL IMAGE PROCES, V3rd
  • [9] Flat-panel cone-beam computed tomography for image-guided radiation therapy
    Jaffray, DA
    Siewerdsen, JH
    Wong, JW
    Martinez, AA
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2002, 53 (05): : 1337 - 1349
  • [10] Application of radiosurgery principles to a target in the breast: A dosimetric study
    Jozsef, G
    Luxton, G
    Formenti, SC
    [J]. MEDICAL PHYSICS, 2000, 27 (05) : 1005 - 1010