ASSESSMENT OF ZERO DRIFT IN THE CODMAN INTRACRANIAL PRESSURE MONITOR: A STUDY FROM 2 NEUROINTENSIVE CARE UNITS

被引:28
作者
Al-Tamimi, Yahia Z. [1 ,2 ]
Helmy, Adel [1 ,2 ]
Bavetta, Seb [3 ]
Price, Stephen J. [1 ,2 ]
机构
[1] Univ Cambridge, Acad Neurosurg Unit, Cambridge, England
[2] Univ Cambridge, Addenbrookes Hosp, NHS Fdn Trust, Cambridge CB2 2QQ, England
[3] Queens Hosp, Dept Neurosurg, Romford, Essex, England
关键词
Codman microsensor; Intracranial pressure monitoring; Zero drift; CLINICAL-EVALUATION; TECHNICAL REPORT; MICROSENSOR; TRANSDUCER;
D O I
10.1227/01.NEU.0000328392.98602.5A
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
OBJECTIVE: Intraparenchymal monitoring devices play an important role in the daily management of head injury and other critically ill neurosurgical patients. Although zero drift data exist for the Camino system (Camino Laboratories, San Diego, CA), only in vitro data exist for the Codman system (Codman and Shurtleff, Inc., Raynham, MA). The aim of this study was to assess the extent of zero drift for the Codman intracranial pressure (ICP) monitor in patients being monitored in 2 neurointensive care units. METHODS: This was a prospective study conducted at 2 neurointensive care units. Eighty-eight patients who required ICP monitoring and who presented to the 2 neurosurgical departments, Center 1 (n = 48) and Center 2 (n = 40), were recruited for participation. The duration of ICP monitoring was noted, as was the resultant pressure reading in normal saline on removing the ICP monitor (zero drift). RESULTS: The median absolute zero drift for the group was 2.0 mm Hg (interquartile range, 1-3 mm Hg). The median time in situ was 108 hours (interquartile range, 69-201 hours). There was a positive correlation between the drift and time of the probe spent in situ (Spearman's correlation coefficient = 0.342; P = 0.001). Of the readings, 20 and 2% showed a drift greater than 5 and 10 mm Hg in magnitude, respectively. CONCLUSION: These data demonstrate that a small amount of zero drift exists in ICP monitors and that this drift increases with time. The wide range in the data demonstrates that some drift readings are quite excessive. This reinforces the school of thought that, although ICP readings contribute significantly to the management of neurosurgical patients, they should be interpreted carefully and in conjunction with clinical and radiological assessment of patients.
引用
收藏
页码:94 / 98
页数:5
相关论文
共 13 条
  • [1] Laboratory testing of the Pressio intracranial pressure monitor
    Allin, David
    Czosnyka, Marek
    Czosnyka, Zofia
    [J]. NEUROSURGERY, 2008, 62 (05) : 1158 - 1161
  • [2] Prospective study of zero drift in fiberoptic pressure monitors used in clinical practice
    Bavetta, S
    Norris, JS
    Wyatt, M
    Sutcliffe, JC
    Hamlyn, PJ
    [J]. JOURNAL OF NEUROSURGERY, 1997, 86 (06) : 927 - 930
  • [3] A prospective comparison of fibre-optic and fluid-filled single lumen bolt subdural pressure transducers in ventilated neurosurgical patients
    Bavetta, S
    Sutcliffe, JC
    Sparrow, OCE
    Hamlyn, PJ
    [J]. BRITISH JOURNAL OF NEUROSURGERY, 1996, 10 (03) : 279 - 284
  • [4] BULLOCK MR, 2007, NEUROTRAUMA, V24, pS45
  • [5] Clinical comparison of the Spiegelberg parenchymal transducer and ventricular fluid pressure
    Chambers, IR
    Siddique, MS
    Banister, K
    Mendelow, AD
    [J]. JOURNAL OF NEUROLOGY NEUROSURGERY AND PSYCHIATRY, 2001, 71 (03) : 383 - 385
  • [6] Laboratory testing of three intracranial pressure microtransducers: Technical report
    Czosnyka, M
    Czosnyka, Z
    Pickard, JD
    [J]. NEUROSURGERY, 1996, 38 (01) : 219 - 224
  • [7] Laboratory testing of the Spiegelberg brain pressure monitor: a technical report
    Czosnyka, M
    Czosnyka, Z
    Pickard, JD
    [J]. JOURNAL OF NEUROLOGY NEUROSURGERY AND PSYCHIATRY, 1997, 63 (06) : 732 - 735
  • [8] Fernandes HM, 1998, ACT NEUR S, V71, P44
  • [9] Clinical evaluation of intraparenchymal spiegelberg pressure sensor
    Lang, JM
    Beck, J
    Zimmermann, M
    Seifert, V
    Raabe, A
    [J]. NEUROSURGERY, 2003, 52 (06) : 1455 - 1459
  • [10] The problem of long-term ICP drift assessment: Improvement by use of the ICP drift index
    Morgalla, MH
    Dietz, K
    Deininger, M
    Grote, EH
    [J]. ACTA NEUROCHIRURGICA, 2002, 144 (01) : 57 - 61