Comparison of defects in ProFile and ProTaper systems after clinical use

被引:147
作者
Shen, Y
Cheung, GSP
Bian, ZA
Peng, B
机构
[1] Wuhan Univ, Key Lab Oral Biomed Engn, Minist Educ, Sch Stomatol,Dept Operat Dent & Endodont, Wuhan 430079, Peoples R China
[2] Univ Hong Kong, Dept Conservat Dent & Endodont, Fac Dent, Hong Kong, Hong Kong, Peoples R China
关键词
clinical; defect; ProFile; ProTaper;
D O I
10.1016/j.joen.2005.10.017
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
The purpose of this study was to compare the incidence and mode of instrument separation of the two nickel-titanium (NiTi) rotary systems that were used according to a predefined schedule of clinical use by the same group of operators. A total of 166 ProFile and 325 ProTaper instruments, discarded from this endodontic clinic over 17 months, was analyzed. The incidences of instrument separation were 7% for ProFile and 14% for ProTaper (p < 0.05). The proportion of unwinding defects was 5% in ProFile and 0.3% in ProTaper instruments (p < 0.01). Flexural fatigue was implicated in the majority of separations in both groups. The results of this study indicated that while ProTaper was more likely to separate without warning, ProFile tended to exhibit unwinding of flutes more frequently.
引用
收藏
页码:61 / 65
页数:5
相关论文
共 30 条
[11]   A COMPARISON OF ROOT-CANAL PREPARATIONS USING NI-TI HAND, NI-TI ENGINE-DRIVEN, AND K-FLEX ENDODONTIC INSTRUMENTS [J].
GLOSSON, CR ;
HALLER, RH ;
DOVE, SB ;
DELRIO, CE .
JOURNAL OF ENDODONTICS, 1995, 21 (03) :146-151
[12]  
Guettier P., 2002, CONT ENDODONTICS, P21
[13]   Fatigue and mechanical properties of nickel-titanium endodontic instruments [J].
Kuhn, G ;
Jordan, L .
JOURNAL OF ENDODONTICS, 2002, 28 (10) :716-720
[14]   Factors influencing the fracture of nickel-titanium rotary instruments [J].
Martín, B ;
Zelada, G ;
Varela, P ;
Bahillo, JG ;
Magán, F ;
Ahn, S ;
Rodríguez, C .
INTERNATIONAL ENDODONTIC JOURNAL, 2003, 36 (04) :262-266
[15]   Factors influencina defects of rotary nickel-titanium endodontic instruments after clinical use [J].
Parashos, P ;
Gordon, I ;
Messer, HH .
JOURNAL OF ENDODONTICS, 2004, 30 (10) :722-725
[16]   A comparison of Greater Taper files, ProFiles, and stainless steel files to shape curved root canals [J].
Park, H .
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTICS, 2001, 91 (06) :715-718
[17]   The influence of a manual glide path on the separation rate of NiTi rotary instruments [J].
Patiño, PV ;
Biedma, BM ;
Liébana, CR ;
Cantatore, G ;
Bahillo, JG .
JOURNAL OF ENDODONTICS, 2005, 31 (02) :114-116
[18]   ProTaper rotary root canal preparation:: assessment of torque and force in relation to canal anatomy [J].
Peters, OA ;
Peters, CI ;
Schönenberger, K ;
Barbakow, F .
INTERNATIONAL ENDODONTIC JOURNAL, 2003, 36 (02) :93-99
[19]   Dynamic torque and apical forces of ProFile .04 rotary instruments during preparation of curved canals [J].
Peters, OA ;
Barbakow, F .
INTERNATIONAL ENDODONTIC JOURNAL, 2002, 35 (04) :379-389
[20]   Current challenges and concepts in the preparation of root canal systems: A review [J].
Peters, OA .
JOURNAL OF ENDODONTICS, 2004, 30 (08) :559-567