Fast Dynamic Reconstruction Algorithm with Joint Bilateral Filtering for Perfusion C-arm CT

被引:0
作者
Manhart, Michael [1 ]
Kowarschik, Markus [2 ]
Fieselmann, Andreas [1 ,3 ]
Yu Deuerling-Zheng [2 ]
Hornegger, Joachim [1 ,3 ]
机构
[1] Univ Erlangen Nurnberg, Pattern Recognit Lab, Dept Comp Sci, Martensstr 3, D-91058 Erlangen, Germany
[2] Siemens AG, Angiog & Intervent X Ray Syst, Forchheim, Germany
[3] Erlangen Grad Sch Adv Opt Technol SAOT, Erlangen, Germany
来源
2012 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE RECORD (NSS/MIC) | 2012年
关键词
Perfusion imaging; dynamic reconstruction; C-arm CT; stroke treatment; MODEL;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Tissue perfusion measurement using C-arm angiography systems is a novel technique with potential high benefit for catheter-guided treatment of stroke in the interventional suite. However, perfusion C-arm CT (PCCT) is challenging: the slow C-arm rotation speed only allows measuring samples of contrast time attenuation curves (TACs) every 5 - 6 s if reconstruction algorithms for static data are used. Furthermore, the peaks of the tissue TACs typically lie in a range of 5 - 30 HU, thus perfusion imaging is very sensitive to noise. We present a dynamic, iterative reconstruction (DIR) approach to reconstruct TACs described by a weighted sum of linear spline functions. The optimization problem is solved using an appropriate initialization and a Landweber-based optimization strategy with a modified backprojection step. To reduce noise a novel regularization technique based on joint bilateral filtering (JBF) is introduced. The algorithm is evaluated using simulation data created with a dynamic cylindrical phantom, a realistic digital brain phantom and real measured data from an animal study with a canine stroke model. Results indicate that the DIR algorithm qualitatively and quantitatively improves reconstructed TACs and perfusion maps compared to classical Feldkamp (FDK) reconstruction. For the brain phantom study the Pearson correlation (PC) of the reconstructed cerebral blood flow (CBF) maps to the ground truth increased from 0.82 (FDK) to 0.87 (DIR). For the canine study the PC of the CBF maps to co-registered perfusion CT maps increased from 0.61 (FDK) to 0.73 (DIR).
引用
收藏
页码:2304 / 2311
页数:8
相关论文
共 50 条
  • [1] Dynamic Iterative Reconstruction for Interventional 4-D C-Arm CT Perfusion Imaging
    Manhart, Michael T.
    Kowarschik, Markus
    Fieselmann, Andreas
    Deuerling-Zheng, Yu
    Royalty, Kevin
    Maier, Andreas K.
    Hornegger, Joachim
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2013, 32 (07) : 1336 - 1348
  • [2] A model based algorithm for perfusion estimation in interventional C-arm CT systems
    Wagner, Martin
    Deuerling-Zheng, Yu
    Moehlenbruch, Markus
    Bendszus, Martin
    Boese, Jan
    Heiland, Sabine
    MEDICAL PHYSICS, 2013, 40 (03)
  • [3] Interventional 4-D C-Arm CT Perfusion Imaging Using Interleaved Scanning and Partial Reconstruction Interpolation
    Fieselmann, Andreas
    Ganguly, Arundhuti
    Yu Deuerling-Zheng
    Zellerhoff, Michael
    Rohkohl, Christopher
    Boese, Jan
    Hornegger, Joachim
    Fahrig, Rebecca
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2012, 31 (04) : 892 - 906
  • [4] C-arm CT Perfusion Imaging in the Interventional Suite
    Fieselmann, Andreas
    Manhart, Michael
    CURRENT MEDICAL IMAGING, 2013, 9 (02) : 96 - 101
  • [5] ITERATIVE DENOISING ALGORITHMS FOR PERFUSION C-ARM CT WITH A RAPID SCANNING PROTOCOL
    Manhart, Michael T.
    Fieselmann, Andreas
    Deuerling-Zheng, Yu
    Kowarschik, Markus
    2013 IEEE 10TH INTERNATIONAL SYMPOSIUM ON BIOMEDICAL IMAGING (ISBI), 2013, : 1235 - 1239
  • [6] Performance evaluation of a C-Arm CT perfusion phantom
    Boese, Axel
    Gugel, Sebastian
    Serowy, Steffen
    Purmann, Jonas
    Rose, Georg
    Beuing, Oliver
    Skalej, Martin
    Kyriakou, Yiannis
    Yu Deuerling-Zheng
    INTERNATIONAL JOURNAL OF COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2013, 8 (05) : 799 - 807
  • [7] Performance evaluation of a C-Arm CT perfusion phantom
    Axel Boese
    Sebastian Gugel
    Steffen Serowy
    Jonas Purmann
    Georg Rose
    Oliver Beuing
    Martin Skalej
    Yiannis Kyriakou
    Yu Deuerling-Zheng
    International Journal of Computer Assisted Radiology and Surgery, 2013, 8 : 799 - 807
  • [8] Dynamic Measurement of Arterial Liver Perfusion With an Interventional C-Arm System
    Datta, Sanjit
    Mueller, Kerstin
    Moore, Terilyn
    Molvin, Lior
    Gehrisch, Sonja
    Rosenberg, Jarrett
    Saenz, Yamil
    Manhart, Michael
    Deuerling-Zheng, Yu
    Kothary, Nishita
    Fahrig, Rebecca
    INVESTIGATIVE RADIOLOGY, 2017, 52 (08) : 456 - 461
  • [9] C-Arm Perfusion Imaging with a Fast Penalized Maximum-Likelihood Approach
    Frysch, Robert
    Pfeiffer, Tim
    Bannasch, Sebastian
    Serowy, Steffen
    Gugel, Sebastian
    Skalej, Martin
    Rose, Georg
    MEDICAL IMAGING 2014: PHYSICS OF MEDICAL IMAGING, 2014, 9033
  • [10] Towards Clinical Application of a Laplace Operator-Based Region of Interest Reconstruction Algorithm in C-Arm CT
    Xia, Yan
    Hofmann, Hannes
    Dennerlein, Frank
    Mueller, Kerstin
    Schwemmer, Chris
    Bauer, Sebastian
    Chintalapani, Gouthami
    Chinnadurai, Ponraj
    Hornegger, Joachim
    Maier, Andreas
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2014, 33 (03) : 593 - 606