A Clinical Study Comparing the Diagnostic Performance of Assist Strain Ratio Against Manual Strain Ratio in Ultrasound Breast Elastography

被引:3
作者
Barr, Richard G. [1 ,2 ]
Managuli, Ravi A. [3 ,4 ]
机构
[1] Southwood Imaging, 7623 Market St, Youngstown, OH 44512 USA
[2] Northeastern Ohio Med Univ, Rootstown, OH USA
[3] Hitachi Healthcare Amer, Twinsburg, OH USA
[4] Univ Washington, Seattle, WA 98195 USA
关键词
breast cancer; elastography; strain ratio; strain; fat-to-lesion ratio; FLR; ELASTICITY; RECOMMENDATIONS; GUIDELINES; SOCIETY; CANCER; US;
D O I
10.1097/RUQ.0000000000000398
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective Strain ratio (SR) is a semiquantitative parameter in differentiating benign from malignant tumors in breast ultrasound elastography. Currently, SR is computed manually and, thus, user dependent. The objective of this study was to evaluate the performance of a new tool assist strain ratio (ASR) and determine how it performs compared with an expert sonologist. Methods Ninety-one patients scheduled for breast biopsy were included in this institutional review board-approved/Health Insurance Portability and Accountability Act-compliant study. For manual strain ratio (MSR), fat and lesion were manually outlined, whereas for ASR, the clinician indicated the lesion center and the fat-to-lesion ratio is computed automatically. Three measurements were obtained for each lesion. The same raw data were used to calculate the MSR and ASR. Results The SR thresholds to differentiate benign from malignant tumors were determined using the Youden index. For MSR, the cutoff was 2.7, and for ASR was 2.8. The MSR showed a sensitivity of 88%, specificity of 64%, accuracy of 77%, positive predictive value of 72%, and negative predictive value of 92.1%. Corresponding ASR showed a sensitivity of 86%, specificity of 76%, accuracy of 81%, positive predictive value of 79%, and negative predictive value of 84%. The areas under the curve for the MSR and ASR were 0.83 and 0.85, respectively. The average coefficients of variation for the MSR and ASR measurements were 43% and 30%, respectively. Conclusion Assist strain ratio demonstrated similar diagnostic performance compared with MSR. In addition, the coefficient of variation of ASR is lower, implying lower intraoperator dependency. Thus, ASR may aid less-experienced scanners in obtaining improved results.
引用
收藏
页码:82 / 87
页数:6
相关论文
共 14 条
  • [1] SEMI-QUANTITATIVE AND QUALITATIVE ASSESSMENT OF BREAST ULTRASOUND ELASTOGRAPHY IN DIFFERENTIATING BETWEEN MALIGNANT AND BENIGN LESIONS
    Alhabshi, Sharifah Majedah Idrus
    Rahmat, Kartini
    Halim, Nurazidawati Abdul
    Aziz, Suraya
    Radhika, Sridharan
    Gan, Gek Choo
    Vijayananthan, Anushya
    Westerhout, Caroline Judy
    Mohd-Shah, Mohammad Nazri
    Jaszle, Saladina
    Latar, Nani Harlina Mohd
    Muhammad, Rohaizak
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 2013, 39 (04) : 568 - 578
  • [2] WFUMB GUIDELINES AND RECOMMENDATIONS FOR CLINICAL USE OF ULTRASOUND ELASTOGRAPHY: PART 2: BREAST
    Barr, Richard G.
    Nakashima, Kazutaka
    Amy, Dominique
    Cosgrove, David
    Farrokh, Andre
    Schafer, Fritz
    Bamber, Jeffrey C.
    Castera, Laurent
    Choi, Byung Ihn
    Chou, Yi-Hong
    Dietrich, Christoph F.
    Ding, Hong
    Ferraioli, Giovanna
    Filice, Carlo
    Friedrich-Rust, Mireen
    Hall, Timothy J.
    Nightingale, Kathryn R.
    Palmeri, Mark L.
    Shiina, Tsuyoshi
    Suzuki, Shinichi
    Sporea, Ioan
    Wilson, Stephanie
    Kudo, Masatoshi
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 2015, 41 (05) : 1148 - 1160
  • [3] Effects of Precompression on Elasticity Imaging of the Breast Development of a Clinically Useful Semiquantitative Method of Precompression Assessment
    Barr, Richard G.
    Zhang, Zheng
    [J]. JOURNAL OF ULTRASOUND IN MEDICINE, 2012, 31 (06) : 895 - 902
  • [4] Real-Time Ultrasound Elasticity of the Breast Initial Clinical Results
    Barr, Richard G.
    [J]. ULTRASOUND QUARTERLY, 2010, 26 (02) : 61 - 66
  • [5] Shear-wave Elastography Improves the Specificity of Breast US: The BE1 Multinational Study of 939 Masses
    Berg, Wendie A.
    Cosgrove, David O.
    Dore, Caroline J.
    Schaefer, Fritz K. W.
    Svensson, William E.
    Hooley, Regina J.
    Ohlinger, Ralf
    Mendelson, Ellen B.
    Balu-Maestro, Catherine
    Locatelli, Martina
    Tourasse, Christophe
    Cavanaugh, Barbara C.
    Juhan, Valerie
    Stavros, A. Thomas
    Tardivon, Anne
    Gay, Joel
    Henry, Jean-Pierre
    Cohen-Bacrie, Claude
    [J]. RADIOLOGY, 2012, 262 (02) : 435 - 449
  • [6] Cancerorg B, US BREAST CANC STAT
  • [7] Diagnostic Value of Strain Ratio Measurement in the Differentiation of Malignant and Benign Breast Lesions
    Farrokh, A.
    Wojcinski, S.
    Degenhardt, F.
    [J]. ULTRASCHALL IN DER MEDIZIN, 2011, 32 (04): : 400 - 405
  • [8] Breast disease: Clinical application of US elastography for diagnosis
    Itoh, A
    Ueno, E
    Tohno, E
    Kamma, H
    Takahashi, H
    Shiina, T
    Yamakawa, M
    Matsumura, T
    [J]. RADIOLOGY, 2006, 239 (02) : 341 - 350
  • [9] Elastic moduli of breast and prostate tissues under compression
    Krouskop, TA
    Wheeler, TM
    Kallel, F
    Garra, BS
    Hall, T
    [J]. ULTRASONIC IMAGING, 1998, 20 (04) : 260 - 274
  • [10] Breast Cancer Screening With Imaging: Recommendations From the Society of Breast Imaging and the ACR on the Use of Mammography, Breast MRI, Breast Ultrasound, and Other Technologies for the Detection of Clinically Occult Breast Cancer
    Lee, Carol H.
    Dershaw, David
    Kopans, Daniel
    Evans, Phil
    Monsees, Barbara
    Monticciolo, Debra
    Brenner, R. James
    Bassett, Lawrence
    Berg, Wendie
    Feig, Stephen
    Hendrick, Edward
    Mendelson, Ellen
    D'Orsi, Carl
    Sickles, Edward
    Burhenne, Linda Warren
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF RADIOLOGY, 2010, 7 (01) : 18 - 27