Modification and application of a Pelorus Mark III Stereotactic System for CT-guided brain biopsy in 50 dogs

被引:53
作者
Koblik, PD
LeCouteur, RA [1 ]
Higgins, RJ
Fick, J
Kortz, GD
Sturges, BK
Pascoe, PJ
机构
[1] Univ Calif Davis, Sch Vet Med, Dept Surg & Radiol Sci, Davis, CA 95616 USA
[2] Univ Calif Davis, Sch Vet Med, Dept Pathol Microbiol & Immunol, Davis, CA 95616 USA
[3] Univ Calif Davis, Sch Vet Med, Vet Med Teaching Hosp, Davis, CA 95616 USA
[4] Med Coll Georgia, Neurosurg Sect, Dept Surg, Augusta, GA 30912 USA
关键词
biopsy; brain; stereotaxy; computed tomography; dog;
D O I
10.1111/j.1740-8261.1999.tb00370.x
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
The Pelorus Mark III Stereotactic System is a commercially available device for CT-guided stereotactic brain biopsy in people. With relatively minor modifications, this de,ice was used to safely and accurately perform CT-guided stereotactic brain biopsies in 50 dogs with intracranial lesions, Modifications were necessary to accommodate a 90 degrees shift in orientation of the canine head compared to the human head during the imaging phase of the procedure, and to facilitate other phases of the biopsy procedure that are affected by the uneven and variable topography of the canine skull. Description of a typical CT-guided brain biopsy procedure in dogs using the modified Pelorus Mark III Stereotactic System is provided. Accuracy of biopsy needle placement was determined by comparing the x, y and z coordinates of the biopsy target site with the actual coordinates of the needle tip on CT images, Mean needle placement error was 3.5 +/- 1.6 mm. Needle placement error was not significantly related to operator experience, dog size (body weight), or needle path length, however, needle placement error was significantly affected by lesion location.
引用
收藏
页码:424 / 433
页数:10
相关论文
共 14 条
[1]  
CAROL M, 1987, APPL NEUROPHYSIOL, V50, P133
[2]  
Engelhard H H 3rd, 1992, J Ky Med Assoc, V90, P24
[3]  
ENGLEHARD HH, 1993, SO MED J, V86, P760
[4]  
GALLOWAY RL, 1991, BIOMED INSTRUM TECHN, V25, P437
[5]  
GHUNG RY, 1998, SURG ONCOL CLIN N AM, V7, P589
[6]  
HARARI J, 1994, PROG VET NEUROL, V4, P41
[7]   CT-guided brain biopsy using a modified Pelorus Mark III Stereotactic System: Experience with 50 dogs [J].
Koblik, PD ;
LeCouteur, RA ;
Higgins, RJ ;
Bollen, AW ;
Vernau, KM ;
Kortz, GD ;
Ilkiw, JE .
VETERINARY RADIOLOGY & ULTRASOUND, 1999, 40 (05) :434-440
[8]  
LECOUTEUR RA, 1998, 16 ANN AM COLL VET I, V12, P207
[9]   MAGNETIC-RESONANCE AND COMPUTED TOMOGRAPHIC IMAGE-DIRECTED STEREOTAXY FOR ANIMAL RESEARCH [J].
MACIUNAS, RJ ;
GALLOWAY, RL .
STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 1989, 53 (03) :197-201
[10]   AN INDEPENDENT APPLICATION ACCURACY EVALUATION OF STEREOTAXIC FRAME SYSTEMS [J].
MACIUNAS, RJ ;
GALLOWAY, RL ;
LATIMER, J ;
COBB, C ;
ZACCHARIAS, E ;
MOORE, A ;
MANDAVA, VR .
STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 1992, 58 (1-4) :103-107