Recessed Ring-Disk Nanoelectrode Arrays Integrated in Nanofluidic Structures for Selective Electrochemical Detection

被引:42
作者
Ma, Chaoxiong [1 ]
Contento, Nicholas M. [2 ]
Gibson, Larry R., II [2 ]
Bohn, Paul W. [1 ,2 ]
机构
[1] Univ Notre Dame, Dept Chem & Biochem, Notre Dame, IN 46556 USA
[2] Univ Notre Dame, Dept Chem & Biomol Engn, Notre Dame, IN 46556 USA
基金
美国国家科学基金会;
关键词
ELECTRODE ARRAYS; ASCORBIC-ACID; MICROFLUIDIC SYSTEM; DOPAMINE DETECTION; REDOX; MICROCHIP; AMPLIFICATION; CHIP; MICROELECTRODE; FABRICATION;
D O I
10.1021/ac402417w
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Arrays of recessed ring-disk (RRD) electrodes with nanoscale spacing fabricated by multilayer deposition, nanosphere lithography, and multistep reactive ion etching were incorporated into nanofluidic channels. These arrays, which characteristically exhibit redox cycling leading to current amplification during cyclic voltammetry, can selectively analyze electroactive species based on differences in redox reversibility, redox potential, or both. Using Ru(NH3)(6)(3+) and ascorbic acid (AA) as model reversible and irreversible redox species, the selectivity for electrochemical measurement of Ru(NH3)(6)(3+) against a background of AA improves from similar to 10, for an array operated in a fluidically unconstrained geometry, to similar to 70 for an array integrated within nanofluidic channels. RRD arrays were also used for the detection of dopamine in the presence of AA by cyclic voltammetry. A linear response ranging from 100 nM to 1 mM with a detection limit of 20 nM was obtained for dopamine alone without nanofludic confinement. In nanochannel-confined arrays, AA was depleted by holding the ring electrodes at +0.5 V versus Ag/AgCl, allowing interference-free determination of dopamine at the disk electrodes in the presence of a 100-fold excess of AA. For selective detection of electrochemically reversible interfering species on an RRD array without nanochannel confinement, a ring potential can be chosen such that one species exhibits exclusively cathodic (anodic) current, allowing the other species to be determined from its anodic (cathodic) current. This approach for selective detection is demonstrated in a mixture of Ru(NH3)(6)(3+) and Fe(CN)(6)(3-), which have resolved redox potentials. The same principle was successfully applied to differentiate species with overlapping redox potentials, such as dopamine/Fe(CN)(6)(3-) and ferrocenemethanol/Fe(CN)(6)(4-).
引用
收藏
页码:9882 / 9888
页数:7
相关论文
共 48 条
[21]   Integrated microfluidic system for electrochemical sensing of glycosylated hemoglobin [J].
Huang, Chao-June ;
Chien, Hui-Ching ;
Chou, Tse-Chuan ;
Lee, Gwo-Bin .
MICROFLUIDICS AND NANOFLUIDICS, 2011, 10 (01) :37-45
[22]   A cyclo olefin polymer microfluidic chip with integrated gold microelectrodes for aqueous and non-aqueous electrochemistry [J].
Illa, Xavi ;
Ordeig, Olga ;
Snakenborg, Detlef ;
Romano-Rodriguez, Albert ;
Compton, Richard G. ;
Kutter, Jorg P. .
LAB ON A CHIP, 2010, 10 (10) :1254-1261
[23]   Nanocavity Redox Cycling Sensors for the Detection of Dopamine Fluctuations in Microfluidic Gradients [J].
Kaetelhoen, Enno ;
Hofmann, Boris ;
Lemay, Serge G. ;
Zevenbergen, Marcel A. G. ;
Offenhaeusser, Andreas ;
Wolfrum, Bernhard .
ANALYTICAL CHEMISTRY, 2010, 82 (20) :8502-8509
[24]   Signal Amplification in a Microchannel from Redox Cycling with Varied Electroactive Configurations of an Individually Addressable Microband Electrode Array [J].
Lewis, Penny M. ;
Sheridan, Leah Bullard ;
Gawley, Robert E. ;
Fritsch, Ingrid .
ANALYTICAL CHEMISTRY, 2010, 82 (05) :1659-1668
[25]   A large-area nanoscale gold hemisphere pattern as a nanoelectrode array [J].
Li, Huaqing ;
Wu, Nianqiang .
NANOTECHNOLOGY, 2008, 19 (27)
[26]   Redox Cycling in Nanoscale-Recessed Ring-Disk Electrode Arrays for Enhanced Electrochemical Sensitivity [J].
Ma, Chaoxiong ;
Contento, Nicholas M. ;
Gibson, Larry R., II ;
Bohn, Paul W. .
ACS NANO, 2013, 7 (06) :5483-5490
[27]   FABRICATION AND EVALUATION OF NANOELECTRODE ENSEMBLES [J].
MENON, VP ;
MARTIN, CR .
ANALYTICAL CHEMISTRY, 1995, 67 (13) :1920-1928
[28]   Microarrays of Ring-Recessed Disk Electrodes in Transient Generator-Collector Mode: Theory and Experiment [J].
Menshykau, Denis ;
O'Mahony, Aoife M. ;
Javier del Campo, F. ;
Xavier Munoz, Francesc ;
Compton, Richard G. .
ANALYTICAL CHEMISTRY, 2009, 81 (22) :9372-9382
[29]   Simultaneous measurement at dopamine and ascorbate at their physiological levels using voltammetric microprobe based on overoxidized poly(1,2-phenylenediamine)-coated carbon fiber [J].
Mo, JW ;
Ogorevc, B .
ANALYTICAL CHEMISTRY, 2001, 73 (06) :1196-1202
[30]   VOLTAMMETRIC MEASUREMENTS OF REVERSIBLE AND QUASI-REVERSIBLE REDOX SPECIES USING CARBON-FILM BASED INTERDIGITATED ARRAY MICROELECTRODES [J].
NIWA, O ;
TABEI, H .
ANALYTICAL CHEMISTRY, 1994, 66 (02) :285-289