The Influence of Ultrasound Ciliary Plasty on Corneal Parameters

被引:3
作者
Bolek, Bartlomiej [1 ]
Wylegala, Adam [1 ]
Wylegala, Edward [1 ]
机构
[1] Med Univ Silesia, Dist Railway Hosp, Sch Med Zabrze, Clin Dept Ophthalmol, Katowice, Poland
关键词
glaucoma; ultrasound ciliary plasty; cyclodestruction; astigmatism; corneal topography; INTENSITY FOCUSED ULTRASOUND; DIODE-LASER CYCLOPHOTOCOAGULATION; TERM-FOLLOW-UP; OPEN-ANGLE GLAUCOMA; TRANSSCLERAL CYCLOPHOTOCOAGULATION; ENDOSCOPIC CYCLOPHOTOCOAGULATION; INTRAOCULAR-PRESSURE; HUMAN AUTOPSY; DEVICE; CYCLOCOAGULATION;
D O I
10.1097/IJG.0000000000001574
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Precis: The ultrasound ciliary plasty (UCP) procedure affects corneal parameters, although the effect is transient. Ultrasound energy affects the sclera, leading to induced astigmatism. The area of energy application correlates with the astigmatism axis. Purpose: To assess the postoperative corneal topography for 1 year after UCP. Materials and Methods: In total, 78 patients with primary and secondary refractory glaucoma were enrolled in the study. The primary outcome measures were the corneal parameters obtained with anterior segment swept-source optical coherence tomography. Intraocular pressure, number of antiglaucoma medications, and best-corrected logMAR visual acuities were analyzed as secondary outcome measures. Measurements were performed preoperatively and at 1 week, and 1, 3, 6, and 12 months postoperatively. A total of 39 patients (39 eyes) completed the follow-up period of 12 months. Results: There was a significant difference in the anterior/posterior steep/flat keratometry and anterior/posterior astigmatism values immediately after UCP. However, 3 months postoperatively these parameters returned to their initial values and there were no significant differences noted. The mean +/- SD values of anterior astigmatism preoperatively and at 1 week, and 1, 3, 6, and 12 months postoperatively were 1.12 +/- 0.66 D, 2.17 +/- 0.91 D (P<0.001), 2.05 +/- 0.93 D (P<0.001), 1.55 +/- 0.75 D (P=0.004), 1.31 +/- 0.70 D (P=0.024), and 1.20 +/- 0.73 D (P=0.406), respectively. The astigmatism axis tended to approach 90 degrees meridian with a significant difference 1 week postoperatively. Central and minimal corneal thickness remained unchanged, whereas the intraocular pressure and the number of antiglaucoma medications decreased significantly. Conclusions: UCP influenced the corneal topography parameters in the immediate postoperative period; however, with time, all parameters returned to their initial values.
引用
收藏
页码:899 / 904
页数:6
相关论文
共 33 条
  • [1] Aptel F, 2016, ACTA OPHTHALMOL, V94, pe268, DOI [10.1111/aos.12913, 10.1111/j.1755-3768.2016.0447]
  • [2] Treatment of refractory open-angle glaucoma using ultrasonic circular cyclocoagulation: a prospective case series
    Aptel, Florent
    Dupuy, Caroline
    Rouland, Jean-Francois
    [J]. CURRENT MEDICAL RESEARCH AND OPINION, 2014, 30 (08) : 1599 - 1605
  • [3] Therapeutic applications of ultrasound in ophthalmology
    Aptel, Florent
    Lafon, Cyril
    [J]. INTERNATIONAL JOURNAL OF HYPERTHERMIA, 2012, 28 (04) : 405 - 418
  • [4] Miniaturized High-Intensity Focused Ultrasound Device in Patients with Glaucoma: A Clinical Pilot Study
    Aptel, Florent
    Charrel, Thomas
    Lafon, Cyril
    Romano, Fabrice
    Chapelon, Jean-Yves
    Blumen-Ohana, Esther
    Nordmann, Jean-Philippe
    Denis, Philippe
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2011, 52 (12) : 8747 - 8753
  • [5] Histologic Effects of a New Device for High-Intensity Focused Ultrasound Cyclocoagulation
    Aptel, Florent
    Charrel, Thomas
    Palazzi, Xavier
    Chapelon, Jean-Yves
    Denis, Philippe
    Lafon, Cyril
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2010, 51 (10) : 5092 - 5098
  • [6] Effect of diode laser cyclophotocoagulation on the anterior segment - An Orbscan study
    Arikan, Gul
    Yaman, Aylin
    Ozbek, Zeynep
    Saatci, Ali Osman
    Durak, Ismet
    [J]. CORNEA, 2008, 27 (02) : 152 - 155
  • [7] Berke S., 2000, J GLAUCOMA, V9, P129, DOI [10.1097/00061198-200002000-00072, DOI 10.1097/00061198-200002000-00072]
  • [8] DEVELOPMENT OF A MINIATURIZED HIFU DEVICE FOR GLAUCOMA TREATMENT WITH CONFORMAL COAGULATION OF THE CILIARY BODIES
    Charrel, Thomas
    Aptel, Florent
    Birer, Alain
    Chavrier, Francoise
    Romano, Fabrice
    Chapelon, Jean-Yves
    Denis, Philippe
    Lafon, Cyril
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 2011, 37 (05) : 742 - 754
  • [9] COLEMAN AL, 1991, OPHTHALMIC SURG LAS, V22, P638
  • [10] Scleral Shape and Its Correlations With Corneal Astigmatism
    Consejo, Alejandra
    Rozema, Jos J.
    [J]. CORNEA, 2018, 37 (08) : 1047 - 1052