Development and Preliminary Evaluation of a Motorized Needle Guide Template for MRI-Guided Targeted Prostate Biopsy

被引:49
作者
Song, Sang-Eun [1 ]
Tokuda, Junichi [1 ]
Tuncali, Kemal [2 ]
Tempany, Clare M. [2 ]
Zhang, Elizabeth [1 ]
Hata, Nobuhiko [1 ]
机构
[1] Harvard Univ, Brigham & Womens Hosp, Sch Med, Surg Nav & Robot Lab,Dept Radiol, Boston, MA 02115 USA
[2] Harvard Univ, Brigham & Womens Hosp, Sch Med, Dept Radiol, Boston, MA 02115 USA
基金
美国国家卫生研究院;
关键词
Magnetic resonance imaging (MRI)-guided procedure; needle guide template; prostate biopsy; robotic intervention; ultrasonic motor; TO-NOISE RATIOS; PNEUMATIC ROBOT; CANCER; SYSTEM; RECONSTRUCTION; LOCALIZATION; PERFORMANCE; PLACEMENT; SOFTWARE; GUIDANCE;
D O I
10.1109/TBME.2013.2240301
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
To overcome the problems of limited needle insertion accuracy and human error in the use of a conventional needle guide template in magnetic resonance imaging (MRI)-guided prostate intervention, we developed a motorized MRI-compatible needle guide template that resembles a transrectal ultrasound-guided prostate template. The motorized template allows automated, gapless needle guidance in a 3T MRI scanner with minimal changes in the current clinical procedure. To evaluate the impact of the motorized template on MRI, signal-to-noise ratio and distortion were measured under various system configurations. A maximum of 44% signal-to-noise ratio decrease was found when the ultrasonic motors were running, and a maximum of 0.4% image distortion was observed due to the presence of the motorized template. To measure needle insertion accuracy, we performed four sets of five random target needle insertions mimicking four biopsy procedures, which resulted in an average in-plane targeting error of 0.94 mm with a standard deviation of 0.34 mm. The evaluation studies indicated that the presence and operation of the motorized template in the MRI bore create insignificant image degradation, and provide submillimeter targeting accuracy. The automated needle guide that is directly controlled by navigation software eliminates human error so that the safety of the procedure can be improved.
引用
收藏
页码:3019 / 3027
页数:9
相关论文
共 47 条
[11]   Image registration for targeted MRI-guided transperineal prostate biopsy [J].
Fedorov, Andriy ;
Tuncali, Kemal ;
Fennessy, Fiona M. ;
Tokuda, Junichi ;
Hata, Nobuhiko ;
Wells, William M. ;
Kikinis, Ron ;
Tempany, Clare M. .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2012, 36 (04) :987-992
[12]   A comparison of two methods for measuring the signal to noise ratio on MR images [J].
Firbank, MJ ;
Coulthard, A ;
Harrison, RM ;
Williams, ED .
PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (12) :N261-N264
[13]   MRI-compatible pneumatic robot for transperineal prostate needle placement [J].
Fischer, Gregory S. ;
Iordachita, Iulian ;
Csoma, Csaba ;
Tokuda, Junichi ;
DiMaio, Simon P. ;
Tempany, Clare M. ;
Hata, Nobuhiko ;
Fichtinger, Gabor .
IEEE-ASME TRANSACTIONS ON MECHATRONICS, 2008, 13 (03) :295-305
[14]   Areas Suspicious for Prostate Cancer: MR-guided Biopsy in Patients with at Least One Transrectal US-guided Biopsy with a Negative Finding-Multiparametric MR Imaging for Detection and Biopsy Planning [J].
Franiel, Tobias ;
Stephan, Carsten ;
Erbersdobler, Andreas ;
Dietz, Ekkehart ;
Maxeiner, Andreas ;
Hell, Nina ;
Huppertz, Alexander ;
Miller, Kurt ;
Strecker, Ralph ;
Hamm, Bernd .
RADIOLOGY, 2011, 259 (01) :162-172
[15]   An integrated visualization system for surgical planning and guidance using image fusion and an open MR [J].
Gering, DT ;
Nabavi, A ;
Kikinis, R ;
Hata, N ;
O'Donnell, LJ ;
Grimson, WEL ;
Jolesz, FA ;
Black, PM ;
Wells, WM .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2001, 13 (06) :967-975
[16]   Robotic system for closed-bore MRI-guided prostatic interventions [J].
Goldenberg, Andrew A. ;
Trachtenberg, John ;
Kucharczyk, Walter ;
Yi, Yang ;
Haider, Masoom. ;
Ma, Leo ;
Weersink, Robert ;
Raoufi, Cyrus .
IEEE-ASME TRANSACTIONS ON MECHATRONICS, 2008, 13 (03) :374-379
[17]   Prospective Assessment of Prostate Cancer Aggressiveness Using 3-T Diffusion-Weighted Magnetic Resonance Imaging-Guided Biopsies Versus a Systematic 10-Core Transrectal Ultrasound Prostate Biopsy Cohort [J].
Hambrock, Thomas ;
Hoeks, Caroline ;
Hulsbergen-van de Kaa, Christina ;
Scheenen, Tom ;
Futterer, Jurgen ;
Bouwense, Stefan ;
van Oort, Inge ;
Schroder, Fritz ;
Huisman, Henkjan ;
Barentsz, Jelle .
EUROPEAN UROLOGY, 2012, 61 (01) :177-184
[18]   MR imaging-guided prostate biopsy with surgical navigation software: Device validation and feasibility [J].
Hata, N ;
Jinzaki, M ;
Kacher, D ;
Cormak, R ;
Gering, D ;
Nabavi, A ;
Silverman, SG ;
D'Amico, AV ;
Kikinis, R ;
Jolesz, FA ;
Tempany, CMC .
RADIOLOGY, 2001, 220 (01) :263-268
[19]   MEASUREMENT OF SIGNAL INTENSITIES IN THE PRESENCE OF NOISE IN MR IMAGES [J].
HENKELMAN, RM .
MEDICAL PHYSICS, 1985, 12 (02) :232-233
[20]   Three-Tesla Magnetic Resonance-Guided Prostate Biopsy in Men With Increased Prostate-Specific Antigen and Repeated, Negative, Random, Systematic, Transrectal Ultrasound Biopsies: Detection of Clinically Significant Prostate Cancers [J].
Hoeks, Caroline M. A. ;
Schouten, Martijn G. ;
Bomers, Joyce G. R. ;
Hoogendoorn, Stefan P. ;
Hulsbergen-van de Kaa, Christina A. ;
Hambrock, Thomas ;
Vergunst, Henk ;
Sedelaar, J. P. Michiel ;
Futterer, Jurgen J. ;
Barentsz, Jelle O. .
EUROPEAN UROLOGY, 2012, 62 (05) :902-909