Similar polyethylene wear between cemented and cementless Oxford medial UKA: a 5-year follow-up randomized controlled trial on 79 patients using radiostereometry

被引:16
作者
Horsager, Kristian [1 ]
Madsen, Frank [2 ]
Odgaard, Anders [3 ]
Jepsen, Claus Fink [2 ]
Romer, Lone [4 ]
Kristensen, Per Wagner [5 ]
Kaptein, Bart L. [6 ]
Soballe, Kjeld [1 ]
Stilling, Maiken [1 ]
机构
[1] Aarhus Univ, Dept Clin Inst, Aarhus, Denmark
[2] Aarhus Univ Hosp, Dept Orthoped, Aarhus, Denmark
[3] Copenhagen Univ Hosp, Herlev Gentofte Hosp, Dept Orthopaed, Hellerup, Denmark
[4] Aarhus Univ Hosp, Dept Radiol, Aarhus, Denmark
[5] Vejle Hosp, Dept Orthoped, Vejle, Denmark
[6] Leiden Univ, Dept Orthoped Surg, Biomech & Imaging Grp, Med Ctr, Leiden, Netherlands
关键词
UNICOMPARTMENTAL KNEE REPLACEMENT; MENISCAL KNEE; HYDROXYAPATITE; OSTEOLYSIS; RETRIEVAL; ARTHROPLASTY; PARTICLES; PROSTHESES; MINIMUM; FAILURE;
D O I
10.1080/17453674.2018.1543757
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background and purpose - Hydroxyapatite (HA)-coated implants have been associated with high polyethylene wear in hip arthroplasties. HA coating as a promoter of wear in knee arthroplasties has not been investigated. We compared the wear-rate of the polyethylene bearing for cemented and cementless HA-coated Oxford medial unicondylar knee arthroplasties (UKA). Secondarily, we investigated whether wear-rates were influenced by overhang or impingement of the bearing. Patients and methods - 80 patients (mean age 64 years), treatment-blinded, were randomized to 1 of 3 Oxford medial UKA versions: cemented with double-pegged or single-pegged femoral component or cementless HA-coated with double-pegged femoral component (ratios 1:1:1). We compared wear between the cemented (n = 55) and cementless group (n = 25) (ratio 2:1). Wear, impingement, and overhang were quantified between surgery and 5-year follow-up using radiostereometry. Clinical outcome was evaluated with the Oxford Knee Score. Results - The mean wear-rate for patients without bearing overhang was 0.04 mm/year (95% CI 0.02-0.07) for the cemented group and 0.05 mm/year (CI 0.02-0.08) for the cementless group. The mean difference in wear was 0.008 mm/year (CI -0.04 to 0.03). No impingement was identified. Half of the patients had medial bearing overhang, mean 2.5 mm (1-5). Wear increased by 0.014 mm/year for each mm increment in overhang. The mean Oxford Knee Score was 39 for the cementless group and 38 for the cemented group at the 5-year follow-up. Interpretation - The wear-rates were similar for the 2 fixation methods, which supports further use of the cementless Oxford medial UKA. However, a caveat is a relatively large 95% CI of the mean difference in wear-rate. Component size and position is important as half of the patients presented with an additional increase in wear-rate due to medial bearing overhang.
引用
收藏
页码:67 / +
页数:8
相关论文
共 39 条
  • [1] POLYETHYLENE WEAR IN MENISCAL KNEE REPLACEMENT - A ONE TO 9-YEAR RETRIEVAL ANALYSIS OF THE OXFORD KNEE
    ARGENSON, JN
    OCONNOR, JJ
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1992, 74 (02): : 228 - 232
  • [2] Polyethylene wear in a non-congruous unicompartmental knee replacement: a retrieval analysis
    Ashraf, T
    Newman, JH
    Desai, VV
    Beard, D
    Nevelos, JE
    [J]. KNEE, 2004, 11 (03) : 177 - 181
  • [3] Bloebaum RD, 1997, CLIN ORTHOP RELAT R, P109
  • [4] BLOEBAUM RD, 1994, CLIN ORTHOP RELAT R, P19
  • [5] In Vitro Comparison of Fixed- and Mobile Meniscal-Bearing Unicondylar Knee Arthroplasties
    Burton, Andrew
    Williams, Sophie
    Brockett, Claire L.
    Fisher, John
    [J]. JOURNAL OF ARTHROPLASTY, 2012, 27 (08) : 1452 - 1459
  • [6] Dettori Joseph R, 2011, Evid Based Spine Care J, V2, P7, DOI 10.1055/s-0030-1267080
  • [7] Premature wear and osteolysis in an HA-coated, uncemented total hip arthroplasty
    Duffy, P
    Sher, JL
    Partington, PF
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2004, 86B (01): : 34 - 38
  • [8] The creep and wear of highly cross-linked polyethylene - A three-year randomised, controlled trial using radiostereometric analysis
    Glyn-Jones, S.
    McLardy-Smith, P.
    Gill, H. S.
    Murray, D. W.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2008, 90B (05): : 556 - 561
  • [9] Superior 11-year survival but higher polyethylene wear of hydroxyapatite-coated Mallory-Head cups
    Gottliebsen, Martin
    Rahbek, Ole
    Ottosen, Per F.
    Soballe, Kjeld
    Stilling, Maiken
    [J]. HIP INTERNATIONAL, 2012, 22 (01) : 35 - 40
  • [10] High wear rates and extensive osteolysis in 3 types of uncemented total hip arthroplasty - A review of the PCA, the Harris Galante and the profile/tri-lock plus arthroplasties with a minimum of 12 years median follow-up in 96 hips
    Hallan, Geir
    Lie, Stein A.
    Havelin, Leif I.
    [J]. ACTA ORTHOPAEDICA, 2006, 77 (04) : 575 - 584