Segmented crystalline scintillators: Empirical and theoretical investigation of a high quantum efficiency EPID based on an initial engineering prototype CsI(TI) detector

被引:34
作者
Sawant, A [1 ]
Antonuk, LE
El-Mohri, Y
Zhao, QH
Wang, Y
Li, YX
Du, H
Perna, L
机构
[1] Univ Michigan, Dept Radiat Oncol, Ann Arbor, MI 48103 USA
[2] St Gobain Crystals, Newbury, OH 44065 USA
关键词
portal imaging; MTF; DQE; high quantum efficiency; segmented crystals; image-guided radiotherapy;
D O I
10.1118/1.2178452
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Modern-day radiotherapy relies on highly sophisticated forms of image guidance in order to implement increasingly conformal treatment plans and achieve precise dose delivery. One of the most important goals of such image guidance is to delineate the clinical target volume from surrounding normal tissue during patient setup and dose delivery, thereby avoiding dependence on surrogates such as bony landmarks. In order to achieve this goal, it is necessary to integrate highly efficient imaging technology, capable of resolving soft-tissue contrast at very low doses, within the treatment setup. In this paper we report on the development of one such modality, which comprises a non-optimized, prototype electronic portal imaging device (EPID) based on a 40 mm thick, segmented crystalline CsI(T1) detector incorporated into an indirect-detection active matrix flat panel imager (AMFPI). The segmented detector consists of a matrix of 160X160 optically isolated, crystalline CSI(TI) elements spaced at 1016 /cm pitch. The detector was coupled to an indirect detection-based active matrix array having a pixel pitch of 508 mu m, with each detector element registered to 2 X 2 array pixels. The performance of the prototype imager was evaluated under very low-dose radiotherapy conditions and compared to that of a conventional megavoltage AMFPI based on a Lanex Fast-B phosphor screen. Detailed quantitative measurements were performed in order to determine the x-ray sensitivity, modulation transfer function, noise power spectrum, and detective quantum efficiency (DQE). In addition, images of a contrast-detail phantom and an anthropomorphic head phantom were also acquired. The prototype imager exhibited approximately 22 times higher zero-frequency DQE (similar to 22%) compared to that of the conventional AMFP1 (similar to 1%). The measured zero-frequency DQE was found to be lower than theoretical upper limits (similar to 27%) calculated from Monte Carlo simulations, which were based solely on the x-ray energy absorbed in the detector-indicating the presence of optical Swank noise. Moreover, due to the nonoptimized nature of this prototype, the spatial resolution was observed to be significantly lower than theoretical expectations. Nevertheless, due to its high quantum efficiency (similar to 55%), the prototype imager exhibited significantly higher DQE than that of the conventional AMFPI across all spatial frequencies. In addition, the frequency-dependent DQE was observed to be relatively invariant with respect to the amount of incident radiation, indicating x-ray quantum limited behavior. Images of the contrast-detail phantom and the head phantom obtained using the prototype system exhibit good visualization of relatively large, low-contrast features, and appear significantly less noisy compared to similar images from a conventional AMFPI. Finally, Monte Carlo-based theoretical calculations indicate that, with proper optimization, further, significant improvements in the DQE performance of such imagers could be achieved. It is strongly anticipated that the realization of optimized versions of such very high-DQE EPIDs would enable megavoltage projection imaging at very low doses, and tomographic imaging from a "beam's eye view" at clinically acceptable doses. (c) 2006 American Association of Physicists in Medicine.
引用
收藏
页码:1053 / 1066
页数:14
相关论文
共 65 条
  • [1] Megavoltage imaging with a large-area, flat-panel, amorphous silicon imager
    Antonuk, LE
    Yorkston, J
    Huang, WD
    Sandler, H
    Siewerdsen, JH
    ElMohri, Y
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1996, 36 (03): : 661 - 672
  • [2] Empirical investigation of the signal performance of a high-resolution, indirect detection, active matrix flat-panel imager (AMFPI) for fluoroscopic and radiographic operation
    Antonuk, LE
    ElMohri, Y
    Siewerdsen, JH
    Yorkston, J
    Huang, W
    Scarpine, VE
    Street, RA
    [J]. MEDICAL PHYSICS, 1997, 24 (01) : 51 - 70
  • [3] Initial performance evaluation of an indirect-detection, active matrix flat-panel imager (AMFPI) prototype for megavoltage imaging
    Antonuk, LE
    El-Mohri, Y
    Huang, WD
    Jee, KW
    Siewerdsen, JH
    Maolinbay, M
    Scarpine, VE
    Sandler, H
    Yorkston, J
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1998, 42 (02): : 437 - 454
  • [4] Electronic portal imaging devices: a review and historical perspective of contemporary technologies and research
    Antonuk, LE
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (06) : R31 - R65
  • [5] Daily targeting of intrahepatic tumors for radiotherapy
    Balter, JM
    Brock, KK
    Litzenberg, DW
    McShan, DL
    Lawrence, TS
    Ten Haken, R
    McGinn, CJ
    Lam, KL
    Dawson, LA
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2002, 52 (01): : 266 - 271
  • [6] Quantification of volumetric and geometric changes occurring during fractionated radiotherapy for head-and-neck cancer using an integrated CT/linear accelerator system
    Barker, JL
    Garden, AS
    Ang, KK
    O'Daniel, JC
    Wang, H
    Court, LE
    Morrison, WH
    Rosenthal, DI
    Chao, KSC
    Tucker, SL
    Mohan, R
    Dong, L
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 59 (04): : 960 - 970
  • [7] BARRETT H H, 1976, Computers in Biology and Medicine, V6, P307, DOI 10.1016/0010-4825(76)90068-8
  • [8] DESIGN PRINCIPLES AND CLINICAL POSSIBILITIES WITH A NEW GENERATION OF RADIATION-THERAPY EQUIPMENT - A REVIEW
    BRAHME, A
    [J]. ACTA ONCOLOGICA, 1987, 26 (06) : 403 - 412
  • [9] Performance of electronic portal imaging devices (EPIDs) used in radiotherapy:: Image quality and dose measurements
    Cremers, F
    Frenzel, T
    Kausch, C
    Albers, D
    Schönborn, T
    Schmidt, R
    [J]. MEDICAL PHYSICS, 2004, 31 (05) : 985 - 996
  • [10] Dobbins J. T., 2000, HDB MED IMAGING, V1, P163