Screening Subclinical Keratoconus With Placido-Based Corneal Indices

被引:25
作者
Ramos-Lopez, Dario [1 ]
Martinez-Finkelshtein, Andrei [1 ,2 ]
Castro-Luna, Gracia M. [4 ]
Burguera-Gimenez, Neus [3 ,5 ]
Vega-Estrada, Alfredo [3 ,5 ]
Pinero, David [6 ]
Alio, Jorge L. [3 ,5 ]
机构
[1] Univ Almeria, Dept Math, Almeria 04120, Spain
[2] Univ Granada, Inst Carlos Theoret & Computat Phys 1, Granada, Spain
[3] VISSUM Corp, Alicante, Spain
[4] VISSUM Corp, Almeria, Spain
[5] Univ Miguel Hernandez, Div Ophthalmol, Alicante, Spain
[6] Univ Alicante, Dept Opt Farmacol & Anat, E-03080 Alicante, Spain
关键词
corneal irregularities; subclinical keratoconus; irregularity index; diagnosis; corneal topography; Placido disks; OCULAR RESPONSE ANALYZER; HIGHER-ORDER ABERRATIONS; IN-SITU KERATOMILEUSIS; BIOMECHANICAL PROPERTIES; RISK-FACTORS; TOPOGRAPHY; EYES; VIDEOKERATOGRAPHY; CLASSIFICATION; ELLIPSE;
D O I
10.1097/OPX.0b013e3182843f2a
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose. To assess in a sample of normal, keratoconic, and keratoconus (KC) suspect eyes the performance of a set of new topographic indices computed directly from the digitized images of the Placido rings. Methods. This comparative study was composed of a total of 124 eyes of 106 patients from the ophthalmic clinics Vissum Alicante and Vissum Almeria (Spain) divided into three groups: control group (50 eyes), KC group (50 eyes), and KC suspect group (24 eyes). In all cases, a comprehensive examination was performed, including the corneal topography with a Placido-based CSO topography system. Clinical outcomes were compared among groups, along with the discriminating performance of the proposed irregularity indices. Results. Significant differences at level 0.05 were found on the values of the indices among groups by means of Mann-Whitney-Wilcoxon nonparametric test and Fisher exact test. Additional statistical methods, such as receiver operating characteristic analysis and K-fold cross validation, confirmed the capability of the indices to discriminate between the three groups. Conclusions. Direct analysis of the digitized images of the Placido mires projected on the cornea is a valid and effective tool for detection of corneal irregularities. Although based only on the data from the anterior surface of the cornea, the new indices performed well even when applied to the KC suspect eyes. They have the advantage of simplicity of calculation combined with high sensitivity in corneal irregularity detection and thus can be used as supplementary criteria for diagnosing and grading KC that can be added to the current keratometric classifications. (Optom Vis Sci 2013;90:335-343)
引用
收藏
页码:335 / 343
页数:9
相关论文
共 53 条
[1]   Neural network-based system for early keratoconus detection from corneal topography [J].
Accardo, PA ;
Pensiero, S .
JOURNAL OF BIOMEDICAL INFORMATICS, 2002, 35 (03) :151-159
[2]  
Agresti A, 2013, Categorical data analysis, V3rd
[3]   Least-squares orthogonal distances fitting of circle, sphere, ellipse, hyperbola, and parabola [J].
Ahn, SJ ;
Rauh, W ;
Warnecke, HJ .
PATTERN RECOGNITION, 2001, 34 (12) :2283-2303
[4]  
Alió JL, 2006, J REFRACT SURG, V22, P539
[5]   Keratoconus-integrated characterization considering anterior corneal aberrations, internal astigmatism, and corneal biomechanics [J].
Alio, Jorge L. ;
Pinero, David P. ;
Aleson, Alicia ;
Teus, Miguel A. ;
Barraquer, Rafael I. ;
Murta, Joaquim ;
Maldonado, Miguel J. ;
Castro de Luna, Gracia ;
Gutierrez, Ramon ;
Villa, Cesar ;
Uceda-Montanes, Antonio .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2011, 37 (03) :552-568
[6]   Imaging of the Cornea: Topography vs Tomography [J].
Ambrosio, Renato, Jr. ;
Belin, Michael W. .
JOURNAL OF REFRACTIVE SURGERY, 2010, 26 (11) :847-849
[7]  
[Anonymous], 1970, STAT METHODS RES WOR
[8]  
[Anonymous], 1999, Nonparametric Statistical Methods
[9]  
Barbero S, 2002, J REFRACT SURG, V18, P263
[10]   CONSTRUCTING CONFIDENCE SETS USING RANK STATISTICS [J].
BAUER, DF .