A longitudinal study of electrical stimulation levels and electrode impedance in children using the Clarion cochlear implant

被引:37
作者
Henkin, Y [1 ]
Kaplan-Neeman, R
Kronenberg, J
Migirov, L
Hildesheimer, M
Muchnik, C
机构
[1] Tel Aviv Univ, Sackler Fac Med, Chaim Sheba Med Ctr, Dept Commun Disorders, IL-52621 Tel Hashomer, Israel
[2] Chaim Sheba Med Ctr, Speech & Hearing Ctr, Tel Hashomer, Israel
[3] Chaim Sheba Med Ctr, Dept Otolaryngol Head & Neck Surg, Tel Hashomer, Israel
关键词
Clarion cochlear implant; children; threshold (T) levels; comfortable (M) levels; dynamic range; electrode impedance;
D O I
10.1080/00016480500443391
中图分类号
R76 [耳鼻咽喉科学];
学科分类号
100213 ;
摘要
Conclusions. Electrical stimulation levels and electrode impedance values (EIVs) in children using the Clarion cochlear implant (CI) programmed with CIS strategy stabilized after 3 months of implant use. The data presented here may be useful as a general guideline for the programming of infants and young children and may further be of help for the identification of patients who fall outside the "average'' range. Objectives. The purpose of the present study was to evaluate changes in electrical stimulation levels, i.e. threshold (T) levels, comfortable (M) levels, dynamic range (DR), and EIVs during the first 18 months of implant use, in children using the Clarion CI. Materials and methods. The maps of 18 pre-lingual children (mean age at implantation 4.2 years; range 1-8), using the Enhanced Bipolar 1.2 or Bipolar standard electrode with the S-Series speech processor programmed with CIS strategy, were examined at five time points: connection, and 3, 6, 12, and 18 months post-initial stimulation. T levels, M levels, DR and EIVs were analyzed according to four cochlear segments: apical, apical-medial, medial-basal, and basal. Results. During the first 3 months of implant use T levels increased to some extent, whereas M levels and DR increased significantly. From 3 months and through the entire follow-up, T and M levels as well as DR were stable. EIVs of current carrying electrodes decreased significantly from connection to the 3-month visit; thereafter a stabilization of values was evident. Electrical stimulation levels and EIVs did not differ among the cochlear segments during the entire follow-up.
引用
收藏
页码:581 / 586
页数:6
相关论文
共 50 条
  • [1] Changes over time in electrical stimulation levels and electrode impedance values in children using the Nucleus 24M cochlear implant
    Henkin, Y
    Kaplan-Neeman, R
    Muchnik, C
    Kronenberg, J
    Hildesheimer, M
    INTERNATIONAL JOURNAL OF PEDIATRIC OTORHINOLARYNGOLOGY, 2003, 67 (08) : 873 - 880
  • [2] Electrode impedance changes over time in MED El cochlear implant children recipients: Relation to stimulation levels and behavioral measures
    Fayed, Eman Asaad
    Zaghloul, Hesham Saad
    Morgan, Ashraf Elsayed
    COCHLEAR IMPLANTS INTERNATIONAL, 2020, 21 (04) : 192 - 197
  • [3] Clinical trial of the CLARION® cochlear implant in children
    Osberger, MJ
    Zimmerman-Phillips, S
    Barker, M
    Geier, L
    ANNALS OF OTOLOGY RHINOLOGY AND LARYNGOLOGY, 1999, 108 (04) : 88 - 92
  • [4] Language development in children implanted with the CLARION® cochlear implant
    Robbins, AM
    Bollard, PM
    Green, J
    ANNALS OF OTOLOGY RHINOLOGY AND LARYNGOLOGY, 1999, 108 (04) : 113 - 118
  • [5] The Change in Electrical Stimulation Levels During 24 Months Postimplantation for a Large Cohort of Adults Using the Nucleus® Cochlear Implant
    Gajadeera, Emalka A.
    Galvin, Karyn L.
    Dowell, Richard C.
    Busby, Peter A.
    EAR AND HEARING, 2017, 38 (03) : 357 - 367
  • [6] Long-Term Evolution of the Electrical Stimulation Levels for Cochlear Implant Patients
    Luis Vargas, Jose
    Sainz, Manuel
    Roldan, Cristina
    Alvarez, Isaac
    de la Torre, Angel
    CLINICAL AND EXPERIMENTAL OTORHINOLARYNGOLOGY, 2012, 5 (04) : 194 - 200
  • [7] A Comparison of Electrical Stimulation Levels Across Ears for Children With Sequential Bilateral Cochlear Implants
    Galvin, Karyn L.
    Abdi, Roghayeh
    Dowell, Richard C.
    Nayagam, Bryony
    EAR AND HEARING, 2019, 40 (05) : 1174 - 1186
  • [8] Effects of Clarion electrode design on mapping levels in children
    Franck, KH
    Shah, UK
    Marsh, RR
    Potsic, WP
    ANNALS OF OTOLOGY RHINOLOGY AND LARYNGOLOGY, 2002, 111 (12) : 1128 - 1132
  • [9] Programming cochlear implant stimulation levels in infants and children with a combination of objective measures
    Gordon, K
    Papsin, BC
    Harrison, RV
    INTERNATIONAL JOURNAL OF AUDIOLOGY, 2004, 43 : S28 - S32
  • [10] Reference function of old electrical stimulation electrode in cochlear-reimplantation in children
    Xu, Y.
    Ren, H-B
    Jiang, L.
    Liu, L-Y
    Han, F-G
    Wang, S-F
    EUROPEAN ANNALS OF OTORHINOLARYNGOLOGY-HEAD AND NECK DISEASES, 2020, 137 (05) : 415 - 417