Detector dose response in megavoltage small photon beams. II. Pencil beam perturbation effects

被引:55
作者
Bouchard, Hugo [1 ]
Kamio, Yuji [2 ]
Palmans, Hugo [1 ,3 ]
Seuntjens, Jan [4 ]
Duane, Simon [1 ]
机构
[1] Natl Phys Lab, Acoust & Ionising Radiat Team, Teddington TW11 0LW, Middx, England
[2] CHUM, Montreal, PQ H2L 4M1, Canada
[3] EBG MedAustron GmbH, Med Phys, A-2700 Wiener Neustadt, Austria
[4] McGill Univ, Med Phys Unit, Montreal, PQ H3G 1A4, Canada
关键词
beam calibration; cavity theory; Monte Carlo simulations; nonstandard beams; perturbation effects; photon beams; quality correction factors; radiation detectors; radiation dosimetry; small fields; MONTE-CARLO-SIMULATION; WALLED IONIZATION CHAMBERS; SMALL-FIELD DETECTORS; ELECTRON-TRANSPORT; ION CHAMBERS; REFERENCE DOSIMETRY; SCATTER FACTORS; CAVITY THEORY; K(Q) FACTORS; DERIVATION;
D O I
10.1118/1.4930798
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To quantify detector perturbation effects in megavoltage small photon fields and support the theoretical explanation on the nature of quality correction factors in these conditions. Methods: In this second paper, a modern approach to radiation dosimetry is defined for any detector and applied to small photon fields. Fano's theorem is adapted in the form of a cavity theory and applied in the context of nonstandard beams to express four main effects in the form of perturbation factors. The pencil-beam decomposition method is detailed and adapted to the calculation of perturbation factors and quality correction factors. The approach defines a perturbation function which, for a given field size or beam modulation, entirely determines these dosimetric factors. Monte Carlo calculations are performed in different cavity sizes for different detection materials, electron densities, and extracameral components. Results: Perturbation effects are detailed with calculated perturbation functions, showing the relative magnitude of the effects as well as the geometrical extent to which collimating or modulating the beam impacts the dosimetric factors. The existence of a perturbation zone around the detector cavity is demonstrated and the approach is discussed and linked to previous approaches in the literature to determine critical field sizes. Conclusions: Monte Carlo simulations are valuable to describe pencil beam perturbation effects and detail the nature of dosimetric factors in megavoltage small photon fields. In practice, it is shown that dosimetric factors could be avoided if the field size remains larger than the detector perturbation zone. However, given a detector and beam quality, a full account for the detector geometry is necessary to determine critical field sizes. (C) 2015 American Association of Physicists in Medicine.
引用
收藏
页码:6048 / 6061
页数:14
相关论文
共 58 条
[1]   A new formalism for reference dosimetry of small and nonstandard fields [J].
Alfonso, R. ;
Andreo, P. ;
Capote, R. ;
Huq, M. Saiful ;
Kilby, W. ;
Kjaell, P. ;
Mackie, T. R. ;
Palmans, H. ;
Rosser, K. ;
Seuntjens, J. ;
Ullrich, W. ;
Vatnitsky, S. .
MEDICAL PHYSICS, 2008, 35 (11) :5179-5186
[2]  
[Anonymous], 1983, MED PHYS, V10, P741
[3]   PENELOPE - AN ALGORITHM FOR MONTE-CARLO SIMULATION OF THE PENETRATION AND ENERGY-LOSS OF ELECTRONS AND POSITRONS IN MATTER [J].
BARO, J ;
SEMPAU, J ;
FERNANDEZVAREA, JM ;
SALVAT, F .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 1995, 100 (01) :31-46
[4]   Monte Carlo calculated and experimentally determined output correction factors for small field detectors in Leksell Gamma Knife Perfexion beams [J].
Benmakhlouf, H. ;
Johansson, J. ;
Paddick, I. ;
Andreo, P. .
PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (10) :3959-3973
[5]   Output correction factors for nine small field detectors in 6 MV radiation therapy photon beams: A PENELOPE Monte Carlo study [J].
Benmakhlouf, Hamza ;
Sempau, Josep ;
Andreo, Pedro .
MEDICAL PHYSICS, 2014, 41 (04)
[6]  
Berger Martin J., 1963, Methods Comput. Phys., V135
[7]  
Berger MJ., 2005, ESTAR, PSTAR, and ASTAR: Computer programs for calculating Stopping-Power and range Tables for electrons. Protons, and Helium Ions (version 1.2.3)
[8]   AN ANALYTIC THEORY OF THE POINT-SOURCE NONUNIFORMITY CORRECTION FACTOR FOR THICK-WALLED IONIZATION CHAMBERS IN PHOTON BEAMS [J].
BIELAJEW, AF .
PHYSICS IN MEDICINE AND BIOLOGY, 1990, 35 (04) :517-538
[9]   CORRECTION FACTORS FOR THICK-WALLED IONIZATION CHAMBERS IN POINT-SOURCE PHOTON BEAMS [J].
BIELAJEW, AF .
PHYSICS IN MEDICINE AND BIOLOGY, 1990, 35 (04) :501-516