Near-infrared light to aid peripheral intravenous cannulation in children: a cluster randomised clinical trial of three devices

被引:63
作者
de Graaff, J. C. [1 ]
Cuper, N. J. [2 ]
Mungra, R. A. A. [3 ]
Vlaardingerbroek, K. [3 ]
Numan, S. C. [3 ]
Kalkman, C. J. [3 ]
机构
[1] Wilhelmina Childrens Hosp, Utrecht, Netherlands
[2] Univ Med Ctr, Dept Med Technol & Clin Phys, Utrecht, Netherlands
[3] Univ Med Ctr, Div Anaesthesia, Utrecht, Netherlands
关键词
FACILITATE BLOOD WITHDRAWAL; PREDICTION RULE; VEIN VIEWER; PLACEMENT; INSERTION; EFFICACY; ACCESS; SCORE; SKIN;
D O I
10.1111/anae.12294
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Intravenous cannulation can be difficult in children. Recently, new devices using near-infrared light to make blood vessels visible have become available. We aimed to evaluate the effectiveness of three such devices in facilitating peripheral intravenous cannulation in children. In this cluster randomised clinical trial, daily operating rooms at a tertiary childrens' hospital were randomised to the use of the VeinViewer((R)), AccuVein((R)) AV300, VascuLuminator Vision((R)) or to a control group. We included 1913 children between birth and 18years scheduled for elective surgery. Suitable veins for cannulation were more easily visible with the VeinViewer (307/322 (95.3%)) and AccuVein (239/254 (94.1%)) devices than with VascuLuminator (229/257 (89.1%)) (p=0.03). However, success at the first attempt was not significantly different among groups, ranging from 73.1% to 75.3% (p=0.93). We conclude that although vein visibility is enhanced, near-infrared devices do not improve cannulation.
引用
收藏
页码:835 / 845
页数:11
相关论文
共 24 条
[1]   Topical anaesthesia and intravenous cannulation success in paediatric patients: a randomized double-blind trial [J].
Arendts, G. ;
Stevens, M. ;
Fry, M. .
BRITISH JOURNAL OF ANAESTHESIA, 2008, 100 (04) :521-524
[2]   Pediatric intravenous insertion in the emergency department - Bevel up or bevel down? [J].
Black, KJL ;
Pusic, MV ;
Harmidy, D ;
McGillivray, D .
PEDIATRIC EMERGENCY CARE, 2005, 21 (11) :707-711
[3]   Vein Viewer-assisted Intravenous Catheter Placement in a Pediatric Emergency Department [J].
Chapman, Laura L. ;
Sullivan, Brenna ;
Pacheco, Amanda L. ;
Draleau, Charlene P. ;
Becker, Bruce M. .
ACADEMIC EMERGENCY MEDICINE, 2011, 18 (09) :966-971
[4]   Vein visualization: patient characteristic factors and efficacy of a new infrared vein finder technology [J].
Chiao, F. B. ;
Resta-Flarer, F. ;
Lesser, J. ;
Ng, J. ;
Ganz, A. ;
Pino-Luey, D. ;
Bennett, H. ;
Perkins, C., Jr. ;
Witek, B. .
BRITISH JOURNAL OF ANAESTHESIA, 2013, 110 (06) :966-971
[5]   The use of near-infrared light for safe and effective visualization of subsurface blood vessels to facilitate blood withdrawal in children [J].
Cuper, Natascha J. ;
Klaessens, John H. G. ;
Jaspers, Joris E. N. ;
de Roode, Rowland ;
Noordmans, Herke Jan ;
de Graaff, Jurgen C. ;
Verdaasdonk, Rudolf M. .
MEDICAL ENGINEERING & PHYSICS, 2013, 35 (04) :433-440
[6]   Near-Infrared Imaging in Intravenous Cannulation in Children: A Cluster Randomized Clinical Trial [J].
Cuper, Natascha J. ;
de Graaff, Jurgen C. ;
Verdaasdonk, Rudolf M. ;
Kalkman, Cor J. .
PEDIATRICS, 2013, 131 (01) :E191-E197
[7]   Predictive factors for difficult intravenous cannulation in pediatric patients at a tertiary pediatric hospital [J].
Cuper, Natascha J. ;
de Graaff, Jurgen C. ;
van Dijk, Atty T. H. ;
Verdaasdonk, Rudolf M. ;
van der Werff, Desiree B. M. ;
Kalkman, Cor J. .
PEDIATRIC ANESTHESIA, 2012, 22 (03) :223-229
[8]   Visualizing Veins With Near-Infrared Light to Facilitate Blood Withdrawal in Children [J].
Cuper, Natascha J. ;
Verdaasdonk, Rudolf M. ;
de Roode, Rowland ;
de Vooght, Karen M. K. ;
Viergever, Max A. ;
Kalkman, Cor J. ;
de Graaff, Jurgen C. .
CLINICAL PEDIATRICS, 2011, 50 (06) :508-512
[9]   SOLAR ULTRAVIOLET-RADIATION EFFECTS ON BIOLOGICAL-SYSTEMS [J].
DIFFEY, BL .
PHYSICS IN MEDICINE AND BIOLOGY, 1991, 36 (03) :299-328
[10]   THE VALIDITY AND PRACTICALITY OF SUN-REACTIVE SKIN TYPE-I THROUGH TYPE-VI [J].
FITZPATRICK, TB .