Control Module Design for a Continuous Positive Airway Pressure Ventilator

被引:1
|
作者
Dai, Min [1 ]
Zhang, Zhisheng [1 ]
Liu, Zhigang
Yin, Dongfu
机构
[1] Southeast Univ, Sch Mech Engn, Nanjing 211189, Jiangsu, Peoples R China
来源
MECHATRONICS AND INDUSTRIAL INFORMATICS, PTS 1-4 | 2013年 / 321-324卷
关键词
CPAP ventilator; Digital PID algorithm; Control module; AVR microcontroller;
D O I
10.4028/www.scientific.net/AMM.321-324.1657
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Continuous positive airway pressure (CPAP) ventilator is an effective treatment device for obstructive sleep apnea syndrome (OSAS). A new control-module design for CPAP ventilator is discussed in this paper. The key technologies, which have improved the performance and reliability of the equipment, such as AVR-based module control, high-precision air pressure detection, improved digital PID control algorithm and smart motor-drive regulation, are described. As indicated by the results from our experiments, the proposed design promises a stable output pressure of the ventilator, the error being less than 0.5cmH(2)O.