Real-time MRI guided atrial septal puncture and balloon septostomy in swine

被引:36
作者
Raval, AN
Karmarkar, PV
Guttman, MA
Ozturk, C
DeSilva, R
Aviles, RJ
Wright, VJ
Schenke, WH
Atalar, E
McVeigh, ER
Lederman, RJ
机构
[1] NHLBI, Cardiovasc Branch, Div Intramural Res, NIH, Bethesda, MD 20892 USA
[2] Johns Hopkins Univ, Dept Radiol, Baltimore, MD USA
[3] NHLBI, Cardiac Energet Lab, Div Intramural Res, NIH, Bethesda, MD 20892 USA
关键词
atrial septal puncture; congenital heart disease; real-time magnetic resonance imaging; interventional MRI; magnetic resonance;
D O I
10.1002/ccd.20579
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Cardiac perforation during atrial septal puncture (ASP) might be avoided by improved image guidance. X-ray fluoroscopy (XRF), which guides ASP, visualizes tissue poorly and does not convey depth information. Ultrasound is limited by device shadows and constrained imaging windows. Alternatively, real-time MRI (rtMRI) provides excellent tissue contrast in any orientation and may enable ASP and balloon atrial septostomy (BAS) in swine. Custom MRI catheters incorporated "active" (receiver antenna) and "passive" (iron or gadolinium) elements. Wholly rtMRI-guided transfemoral ASP and BAS were performed in 10 swine in a 1.5T interventional suite. Hemodynamic results were measured with catheters and velocity encoded MRI. Successful ASP was performed in all 10 animals. Necropsy confirmed septostomy confined within the fossa ovalis in all. BAS was successful in 9/10 animals. Antenna failure in a re-used needle led to inadvertent vena cava tear prior to BAS in 1 animal. ASP in the same animal was easily performed using a new needle. rtMRI illustrated clear device-tissue-lumen relationships in multiple orientations, and facilitated simple ASP and BAS. The mean procedure time was 19 +/- 10 minutes. Septostomy achieved a mean left to right shunt ratio of 1.3:1 in these healthy animals. Interactive rtMRI permits rapid transcatheter ASP and BAS in swine. Further technical development may enable novel applications. Published 2006 Wiley-Liss, inc.
引用
收藏
页码:637 / 643
页数:7
相关论文
共 63 条
[1]  
Allen HD, 1998, CIRCULATION, V97, P609
[2]   Magnetic resonance image-guided trans-septal puncture in a swine heart [J].
Arepally, A ;
Karmarkar, PV ;
Weiss, C ;
Rodriguez, ER ;
Lederman, RJ ;
Atalar, E .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2005, 21 (04) :463-467
[3]  
Bartels LW, 2000, J MAGN RESON IMAGING, V12, P599, DOI 10.1002/1522-2586(200010)12:4<599::AID-JMRI13>3.3.CO
[4]  
2-Z
[5]  
Block Peter C, 2003, J Interv Cardiol, V16, P93, DOI 10.1046/j.1540-8183.2003.08007.x
[6]   LIMBIC LEDGE - A LANDMARK FOR TRANSSEPTAL LEFT HEART CATHETERIZATION [J].
BLOOMFIE.DA ;
SINCLAIR.BC .
CIRCULATION, 1965, 31 (01) :103-&
[7]   TRANS-SEPTAL LEFT HEART CATHETERIZATION - A REVIEW OF 278 STUDIES [J].
BLUNDQVIST, C ;
OLSSON, SB ;
VARNAUSKAS, E .
CLINICAL CARDIOLOGY, 1986, 9 (01) :21-26
[8]  
BRAUNWALD E, 1968, CIRCULATION, V37, pII74
[9]   TRANSSEPTAL LEFT HEART CATHETERIZATION - A REVIEW OF 450 STUDIES AND DESCRIPTION OF AN IMPROVED TECHNIC [J].
BROCKENBROUGH, EC ;
BRAUNWALD, E ;
ROSS, J .
CIRCULATION, 1962, 25 (01) :15-&
[10]   Real-time MR guidance for inferior vena cava filter placement in an animal model [J].
Bücker, A ;
Neuerburg, JM ;
Adam, GB ;
Glowinski, A ;
Schaeffter, T ;
Rasche, V ;
van Vaals, JJ ;
Günther, RW .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2001, 12 (06) :753-756