Super-global distortion correction for a rotational C-arm x-ray image intensifier

被引:32
作者
Liu, RR [1 ]
Rudin, S
Bednarek, DR
机构
[1] SUNY Buffalo, Dept Phys, Toshiba Stroke Res Ctr, Buffalo, NY 14214 USA
[2] SUNY Buffalo, Dept Radiol, Toshiba Stroke Res Ctr, Buffalo, NY 14214 USA
[3] SUNY Buffalo, Dept Neurosurg, Toshiba Stroke Res Ctr, Buffalo, NY 14214 USA
[4] SUNY Buffalo, Dept Physiol & Biophys, Toshiba Stroke Res Ctr, Buffalo, NY 14214 USA
关键词
distortion correction; image intensifier; cone-beam CT; digital radiography;
D O I
10.1118/1.598684
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Image intensifier (II) distortion changes as a function of C-arm rotation angle because of changes in the orientation of the II with the earth's or other stray magnetic fields. For cone-beam computed tomography (CT), distortion correction for all angles is essential. The new super-global distortion correction consists of a model to continuously correct II distortion not only at each location in the image but for every rotational angle of the C arm. Calibration bead images were acquired with a standard C arm in 9 in. II mode. The super-global (SG)model is obtained from the single-plane global correction of the selected calibration images with given sampling angle interval. The fifth-order single-plane global corrections yielded a residual rms error of 0.20 pixels, while the SG model yielded a rms error of 0.21 pixels, a negligibly small difference. We evaluated the accuracy dependence of the SG model on various factors, such as the single-plane global fitting order, SG order, and angular sampling interval. We found that a good SC model can be obtained using a sixth-order SG polynomial fit based on the fifth-order single-plane global correction, and that a 10 degrees sampling interval was sufficient. Thus, the SG model saves processing resources and storage space. The residual errors from the mechanical errors of the x-ray system were also investigated, and found comparable with the SG residual error. Additionally, a single-plane global correction was done in the cylindrical coordinate system, and physical information about pincushion distortion and S distortion were observed and analyzed; however, this method is not recommended due to a lack of calculational efficiency. In conclusion, the SC model provides an accurate, fast, and simple correction for rotational C-arm images, which may be used for cone-beam CT. (C) 1999 American Association of Physicists in Medicine. [S0094-2405(99)00509-X].
引用
收藏
页码:1802 / 1810
页数:9
相关论文
共 6 条
[1]   Three-dimensional computed tomographic reconstruction using a C-arm mounted XRII: Correction of image intensifier distortion [J].
Fahrig, R ;
Moreau, M ;
Holdsworth, DW .
MEDICAL PHYSICS, 1997, 24 (07) :1097-1106
[2]   The accuracy and reproducibility of a global method to correct for geometric image distortion in the x-ray imaging chain [J].
Gronenschild, E .
MEDICAL PHYSICS, 1997, 24 (12) :1875-1888
[3]  
PRESS WH, 1996, NUMERICAL RECIPES C, P411
[4]   ACCURATE CHARACTERIZATION OF IMAGE INTENSIFIER DISTORTION [J].
RUDIN, S ;
BEDNAREK, DR ;
WONG, R .
MEDICAL PHYSICS, 1991, 18 (06) :1145-1151
[5]  
SCHUELER B, 1995, P SOC PHOTO-OPT INS, V2432, P272, DOI 10.1117/12.208345
[6]   Accurate and efficient image intensifier distortion correction algorithm for volume tomographic angiography [J].
Wang, XH ;
Ning, RL .
OPTICAL ENGINEERING, 1998, 37 (03) :977-983