A novel assessment of local impedance during catheter ablation: initial experience in humans comparing local and generator measurements

被引:50
作者
Gunawardene, Melanie [1 ]
Muenkler, Paula [1 ,2 ]
Eickholt, Christian [1 ]
Akbulak, Ruken Oe [1 ]
Jularic, Mario [1 ]
Klatt, Niklas [1 ]
Hartmann, Jens [1 ]
Dinshaw, Leon [1 ]
Jungen, Christiane [1 ,2 ]
Moser, Julia M. [1 ]
Merbold, Lydia [3 ]
Willems, Stephan [1 ,2 ]
Meyer, Christian [1 ,2 ]
机构
[1] Univ Hosp Hamburg Eppendorf, Univ Heart Ctr, Dept Cardiac Electrophysiol, Martinistr 52, D-20246 Hamburg, Germany
[2] DZHK German Ctr Cardiovasc Res, Partner Site Hamburg Kiel Luebeck, Berlin, Germany
[3] Boston Sci, Marlborough, MA USA
来源
EUROPACE | 2019年 / 21卷
关键词
Atrial fibrillation; Catheter ablation; Radiofrequency current; Impedance; Ultra high-density mapping; ELECTRODE-TISSUE CONTACT; RADIOFREQUENCY LESION SIZE; PULMONARY VEIN ISOLATION; ATRIAL-FIBRILLATION; FORCE; INDICATOR; DECREASE;
D O I
10.1093/europace/euy273
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Aims A novel measure of local impedance (LI) has been found to predict lesion formation during radiofrequency current (RFC) catheter ablation. The aim of this study was to investigate the utility of this novel approach, while comparing LI to the well-established generator impedance (GI). Methods and results In 25 consecutive patients with a history of atrial fibrillation, catheter ablation was guided by a 3D-mapping system measuring LI in addition to GI via an ablation catheter tip with three incorporated mini-electrodes. Local impedance and GI before and during RFC applications were studied. In total, 381 RFC applications were analysed. The baseline LI was higher in high-voltage areas (>0.5 mV; LI: 110.5 13.7 ) when compared with intermediate-voltage sites (0.1-0.5 mV; 90.9 +/- 10.1 , P < 0.001), low-voltage areas (<0.1 mV; 91.9 +/- 16.4 , P < 0.001), and blood pool LI (91.9 +/- 9.9 , P < 0.001). During ablation, mean LI drop (oLI; 13.1 +/- 9.1 ) was 2.15 times higher as mean GI drop (oGI) (6.1 +/- 4.2 , P < 0.001). Baseline LI correlated with oLI: a mean LI of 99.9 predicted a oLI of 12.9 [95% confidence interval (12.1-13.6), R-2 0.41; P < 0.001]. This relationship was weak for baseline GI predicting oGI (R-2 0.06, P < 0.001). Catheter movements were represented by rapid LI changes. The duration of an RFC application was not predictive for catheter-tissue coupling with no further change of oLI (P = 0.247) nor oGI (P = 0.376) during prolonged ablation. Conclusion Local impedance can be monitored during ablation. Compared with the sole use of GI, baseline LI is a better predictor of impedance drops during ablation and may provide useful insights regarding lesion formation. However, further studies are needed to investigate if this novel approach is useful to guide catheter ablation.
引用
收藏
页码:34 / 42
页数:9
相关论文
共 21 条
  • [1] Calkins H, 2017, HEART RHYTHM, V14, pE275, DOI [10.1093/europace/eux274, 10.1016/j.hrthm.2017.05.012, 10.1093/europace/eux275]
  • [2] MYOCARDIAL ELECTRICAL-IMPEDANCE MAPPING OF ISCHEMIC SHEEP HEARTS AND HEALING ANEURYSMS
    FALLERT, MA
    MIROTZNIK, MS
    DOWNING, SW
    SAVAGE, EB
    FOSTER, KR
    JOSEPHSON, ME
    BOGEN, DK
    [J]. CIRCULATION, 1993, 87 (01) : 199 - 207
  • [3] Electrogram signature of specific activation patterns: Analysis of atrial tachycardias at high-density endocardial mapping
    Frontera, Antonio
    Takigawa, Masateru
    Martin, Ruairidh
    Thompson, Nathaniel
    Cheniti, Ghassen
    Massoullie, Gregoire
    Duchateau, Josselin
    Wielandts, Jean Yves
    Teijeira, Elvis
    Kitamura, Takeshi
    Wolf, Michael
    Al-Jefairi, Nora
    Vlachos, Konstantinos
    Yamashita, Seigo
    Amraoui, Sana
    Denis, Arnaud
    Hocini, Meleze
    Cochet, Hubert
    Sacher, Frederic
    Jais, Pierre
    Haissaguerre, Michel
    Derval, Nicolas
    [J]. HEART RHYTHM, 2018, 15 (01) : 28 - 37
  • [4] Identification of pulmonary vein reconnection gaps with high-density mapping in redo atrial fibrillation ablation procedures
    Garcia-Bolao, Ignacio
    Ballesteros, Gabriel
    Ramos, Pablo
    Menendez, Diego
    Erkiaga, Ane
    Neglia, Renzo
    Jimenez Martin, Marta
    Vives-Rodriguez, Enrique
    [J]. EUROPACE, 2018, 20 : F351 - F358
  • [5] Use of electrical coupling information in AF catheter ablation: A prospective randomized pilot study
    Gaspar, Thomas
    Sih, Haris
    Hindricks, Gerhard
    Eitel, Charlotte
    Sommer, Philipp
    Kircher, Simon
    Rolf, Sascha
    Arya, Arash
    Teplitsky, Liane
    Piorkowski, Christopher
    [J]. HEART RHYTHM, 2013, 10 (02) : 176 - 181
  • [6] Influence of energy source on early atrial fibrillation recurrences: a comparison of cryoballoon vs. radiofrequency current energy ablation with the endpoint of unexcitability in pulmonary vein isolation
    Gunawardene, Melanie A.
    Hoffmann, Boris A.
    Schaeffer, Benjamin
    Chung, Da-Un
    Moser, Julia
    Akbulak, Ruken Oezge
    Jularic, Mario
    Eickholt, Christian
    Nuehrich, Jana
    Meyer, Christian
    Willems, Stephan
    [J]. EUROPACE, 2018, 20 (01): : 43 - 49
  • [7] IMPEDANCE MONITORING DURING RADIOFREQUENCY CATHETER ABLATION IN HUMANS
    HARVEY, M
    KIM, YN
    SOUSA, J
    ELATASSI, R
    MORADY, F
    CALKINS, H
    LANGBERG, JJ
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1992, 15 (01): : 22 - 27
  • [8] Relationship Between Catheter Contact Force and Radiofrequency Lesion Size and Incidence of Steam Pop in the Beating Canine Heart Electrogram Amplitude, Impedance, and Electrode Temperature Are Poor Predictors of Electrode-Tissue Contact Force and Lesion Size
    Ikeda, Atsushi
    Nakagawa, Hiroshi
    Lambert, Hendrik
    Shah, Dipen C.
    Fonck, Edouard
    Yulzari, Aude
    Sharma, Tushar
    Pitha, Jan V.
    Lazzara, Ralph
    Jackman, Warren M.
    [J]. CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2014, 7 (06) : 1174 - 1180
  • [9] Impact of impedance decrease during radiofrequency current application for atrial fibrillation ablation on myocardial lesion and gap formation
    Inaba, Osamu
    Nagata, Yasutoshi
    Sekigawa, Masahiro
    Miwa, Naoyuki
    Yamaguchi, Junji
    Miyamoto, Takamichi
    Goya, Masahiko
    Hirao, Kenzo
    [J]. JOURNAL OF ARRHYTHMIA, 2018, 34 (03) : 247 - 253
  • [10] Endocardial impedance mapping during circumferential pulmonary vein ablation of atrial fibrillation differentiates between atrial and venous tissue
    Lang, CCE
    Gugliotta, F
    Santinelli, V
    Mesas, C
    Tomita, T
    Vicedomini, G
    Augello, G
    Gulletta, S
    Mazzone, P
    De Cobelli, F
    Del Maschio, A
    Pappone, C
    [J]. HEART RHYTHM, 2006, 3 (02) : 171 - 178