A Robotic Flexible Drill and Its Navigation System for Total Hip Arthroplasty

被引:15
作者
Fuad, Ahmad Nazmi Bin Ahmad [1 ]
Elangovan, Hariprashanth [1 ]
Deep, Kamal [2 ]
Yao, Wei [1 ]
机构
[1] Univ Strathclyde, Dept Biomed Engn, 106 Rottenrow, Glasgow G4 0NW, Lanark, Scotland
[2] Golden Jubilee Natl Hosp, Agamemnon St, Clydebank G81 4DY, Scotland
关键词
Robotics; Flexible; Steerable; Tracking; Navigation; Total hip arthroplasty (THA); Orthopaedics; ASSISTED ORTHOPEDIC-SURGERY; STEM IMPLANTATION; KNEE REPLACEMENT; COMPUTER; ACCURACY; TOOL;
D O I
10.1007/s10439-017-1959-5
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This paper presents a robotic flexible drill and its navigation system for total hip arthroplasty (THA). The new robotic system provides an unprecedented and unique capability to perform curved femoral milling under the guidance of a multimodality navigation system. The robotic system consists of three components. Firstly, a flexible drill manipulator comprises multiple rigid segments that act as a sheath to a flexible shaft with a drill/burr attached to the end. The second part of the robotic system is a hybrid tracking system that consists of an optical tracking system and a position tracking system. Optical tracking units are used to track the surgical objects and tools outside the drilling area, while a rotary encoder placed at each joint of the sheath is synchronized to provide the position information for the flexible manipulator with its virtual object. Finally, the flexible drill is integrated into a computer-aided navigation system. The navigation system provides real time guidance to a surgeon during the procedure. The flexible drill system is then able to implement THA by bone milling. The final section of this paper is an evaluation of the flexible and steerable drill and its navigation system for femoral bone milling in sawbones.
引用
收藏
页码:464 / 474
页数:11
相关论文
共 32 条
[11]  
Jakopec M, 2001, Comput Aided Surg, V6, P329, DOI 10.3109/10929080109146302
[12]   Precision freehand sculpting for unicondylar knee replacement: design and experimental validation [J].
Jaramaz, Branislav ;
Nikou, Constantinos .
BIOMEDICAL ENGINEERING-BIOMEDIZINISCHE TECHNIK, 2012, 57 (04) :293-299
[13]  
Langlotz F, 1998, CLIN ORTHOP RELAT R, P92
[14]  
Lavallee S, 1995, J Image Guid Surg, V1, P65, DOI 10.1002/(SICI)1522-712X(1995)1:1<65::AID-IGS10>3.0.CO
[15]  
2-7
[16]   Spatial motion constraints using virtual fixtures generated by anatomy [J].
Li, Ming ;
Ishii, Masaru ;
Taylor, Russell H. .
IEEE TRANSACTIONS ON ROBOTICS, 2007, 23 (01) :4-19
[17]  
Lonner Jess H, 2009, Am J Orthop (Belle Mead NJ), V38, P3
[18]   Computer- and robot-assisted total knee replacement: Analysis of a new surgical procedure [J].
Martelli, M ;
Marcacci, M ;
Nofrini, L ;
La Palombara, F ;
Malvisi, A ;
Iacono, F ;
Vendruscolo, P ;
Pierantoni, M .
ANNALS OF BIOMEDICAL ENGINEERING, 2000, 28 (09) :1146-1153
[19]   Low accuracy of stem implantation in THR using the CASPAR-system - Anteversion measurements in 10 hips [J].
Mazoochian, F ;
Pellengahr, C ;
Huber, A ;
Kircher, J ;
Refior, HJ ;
Jansson, V .
ACTA ORTHOPAEDICA SCANDINAVICA, 2004, 75 (03) :261-264
[20]   Fiber grating sensors in medicine: Current and emerging applications [J].
Mishra, Vandana ;
Singh, Nahar ;
Tiwari, Umesh ;
Kapur, Pawan .
SENSORS AND ACTUATORS A-PHYSICAL, 2011, 167 (02) :279-290