Effects of implant thread design on primary stability-a comparison between single- and double-threaded implants in an artificial bone model

被引:25
作者
Yamaguchi, Yoko [1 ]
Shiota, Makoto [2 ]
Fujii, Masaki [1 ]
Shimogishi, Masahiro [2 ]
Munakata, Motohiro [1 ]
机构
[1] Showa Univ, Dept Implant Dent, Sch Dent, Ota Ku, 2-1-1 Kitasenzoku, Tokyo 1458515, Japan
[2] Tokyo Med & Dent Univ, Oral Implantol & Regenerat Dent Med, Dept Masticatory Funct Rehabil, Div Oral Hlth Sci,Grad Sch,Bunkyo Ku, 1-5-45 Yushima, Tokyo 1138510, Japan
关键词
Primary stability; Single-threaded implant; Double-threaded implant; Artificial bone; Insertion torque; Removal torque; Low-density bone; INSERTION TORQUE; DENTAL IMPLANTS; IN-VITRO; DENSITY; SURFACE;
D O I
10.1186/s40729-020-00239-1
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Background Primary implant stability is essential for osseointegration. To increase stability without changing the implant size, the thread length must be extended by reducing pitch, using a double-threaded implant, or reducing pitch/lead and lead angle to half that of a single-threaded implant. Materials and methods We tested the stabilities of these configurations using artificial bone. A 1.2-mm pitch, single-threaded implant (12S) was the control. We tested a 0.6-mm pitch/1.2-mm-lead double-threaded implant (06D) and a 0.6-mm pitch/lead single-threaded implant (06S). We compared stabilities by measuring insertion torque, removal torque, and the implant stability quotient (ISQ). Damage to bone tissue caused by the implants was evaluated using microscopy and morphometric analysis. Results We show that 06D and 06S significantly improved stability compared with the 12S reference. The stability of 06S was significantly greater compared with that of 06D, except for ISQ. The three implants were associated with bone tissue damage characterized by debris and voids surrounding the implant/bone interface. The 06D caused the most tissue damage, followed by 06S and then 12S. Conclusion These findings indicate that primary stability was significantly improved by changing the implant size, extending the thread length with reduced pitch/lead, and reducing the lead angle to half that of a single-threaded implant compared with a double-threaded implant.
引用
收藏
页数:9
相关论文
共 30 条
[1]   Insertion Torque of Variable-Thread Tapered Implants in the Posterior Maxilla: A Clinical Study [J].
Abi-Aad, Habib L. ;
Daher, Fadi I. ;
Baba, Nadim Z. ;
Cordioli, Giampiero ;
Majzoub, Zeina A. K. .
JOURNAL OF PROSTHODONTICS-IMPLANT ESTHETIC AND RECONSTRUCTIVE DENTISTRY, 2019, 28 (02) :E788-E794
[2]   The effect of thread pattern upon implant osseointegration [J].
Abuhussein, Heba ;
Pagni, Giorgio ;
Rebaudi, Alberto ;
Wang, Hom-Lay .
CLINICAL ORAL IMPLANTS RESEARCH, 2010, 21 (02) :129-136
[3]   OSSEOINTEGRATED TITANIUM IMPLANTS - REQUIREMENTS FOR ENSURING A LONG-LASTING, DIRECT BONE-TO-IMPLANT ANCHORAGE IN MAN [J].
ALBREKTSSON, T ;
BRANEMARK, PI ;
HANSSON, HA ;
LINDSTROM, J .
ACTA ORTHOPAEDICA SCANDINAVICA, 1981, 52 (02) :155-170
[4]   Undersized Implant Site Preparation to Enhance Primary Implant Stability in Poor Bone Density: A Prospective Clinical Study [J].
Alghamdi, Hamdan ;
Anand, Pradeep S. ;
Anil, Sukumaran .
JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2011, 69 (12) :E506-E512
[5]  
Almutairi Abdullah Saleh, 2018, F1000Res, V7, P1898, DOI 10.12688/f1000research.17292.1
[6]   A biomechanical assessment of the relation between the oral implant stability at insertion and subjective bone quality assessment [J].
Alsaadi, Ghada ;
Quirynen, Marc ;
Michiels, Katleen ;
Jacobs, Reinhilde ;
van Steenberghe, Daniel .
JOURNAL OF CLINICAL PERIODONTOLOGY, 2007, 34 (04) :359-366
[7]  
Brunski J B, 1999, Adv Dent Res, V13, P99
[8]   Characterization of commercial rigid polyurethane foams used as bone analogs for implant testing [J].
Calvert, Kayla L. ;
Trumble, Kevin P. ;
Webster, Thomas J. ;
Kirkpatrick, Lynn A. .
JOURNAL OF MATERIALS SCIENCE-MATERIALS IN MEDICINE, 2010, 21 (05) :1453-1461
[9]   Implant-retained mandibular overdentures with immediate loading - A retrospective multicenter study on 226 consecutive cases [J].
Chiapasco, M ;
Gatti, C ;
Rossi, E ;
Haefliger, W ;
Markwalder, TH .
CLINICAL ORAL IMPLANTS RESEARCH, 1997, 8 (01) :48-57
[10]   The insertion torque-depth curve integral as a measure of implant primary stability: An in vitro study on polyurethane foam blocks [J].
Di Stefano, Danilo Alessio ;
Arosio, Paolo ;
Gastaldi, Giorgio ;
Gherlone, Enrico .
JOURNAL OF PROSTHETIC DENTISTRY, 2018, 120 (05) :706-714