Signal-to-thickness calibration and pixel-wise interpolation for beam-hardening artefact reduction in microCT

被引:4
作者
Gustscin, N. [1 ,2 ]
Gustschin, A. [1 ,2 ]
Epple, F. M. [3 ]
Allner, S. [1 ,2 ]
Achterhold, K. [1 ,2 ]
Herzen, J. [1 ,2 ]
Pfeiffer, F. [1 ,2 ,4 ]
机构
[1] Tech Univ Munich, Dept Phys, Chair Biomed Phys, D-85748 Garching, Germany
[2] Tech Univ Munich, Munich Sch BioEngn, D-85748 Garching, Germany
[3] MITOS GmbH, D-85748 Garching, Germany
[4] Tech Univ Munich, Klinikum Rechts Isar, Dept Diagnost & Intervent Radiol, D-81675 Munich, Germany
关键词
RAY COMPUTED-TOMOGRAPHY;
D O I
10.1209/0295-5075/125/38003
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
X-ray computed tomography (CT) reconstruction suffers from beam-hardening arte facts caused by the polychromaticity of virtually all lab-based X-ray sources. A method to correct for beam-hardening is a direct, pixel-wise signal-to-thickness calibration (STC). We compare reconstructions of conventionally flat-field corrected as well as STC preprocessed measurements of various samples performed on a commercial microCT device based on a flat-panel detector. We show that a good estimate between the transmission signal and the respective material thickness can be given by multiple exponential functions. We further compare the exponential interpolation approach to a hyperbolic model, which reduces the number of necessary calibration measurements significantly. Our method shows that typical beam-hardening artefacts like cupping and filling can be almost completely suppressed and a significant contrast increase is gained. The method can be applied with little additional calibration and computation effort and allows shorter acquisition times since beam filtration can be reduced or omitted. Copyright (C) EPLA, 2019
引用
收藏
页数:6
相关论文
共 22 条
[1]   Guidelines for Assessment of Bone Microstructure in Rodents Using Micro-Computed Tomography [J].
Bouxsein, Mary L. ;
Boyd, Stephen K. ;
Christiansen, Blaine A. ;
Guldberg, Robert E. ;
Jepsen, Karl J. ;
Mueller, Ralph .
JOURNAL OF BONE AND MINERAL RESEARCH, 2010, 25 (07) :1468-1486
[2]   BEAM HARDENING IN X-RAY RECONSTRUCTIVE TOMOGRAPHY [J].
BROOKS, RA ;
DICHIRO, G .
PHYSICS IN MEDICINE AND BIOLOGY, 1976, 21 (03) :390-398
[3]  
Davis G., 2008, P SOC PHOTO-OPT INS, V7078
[4]   A beam hardening correction for X-ray microtomography [J].
de Paiva, RF ;
Lynch, J ;
Rosenberg, E ;
Bisiaux, M .
NDT & E INTERNATIONAL, 1998, 31 (01) :17-22
[5]   POLYCHROMATIC STREAK ARTIFACTS IN COMPUTED TOMOGRAPHY IMAGES [J].
DUERINCKX, AJ ;
MACOVSKI, A .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1978, 2 (04) :481-487
[6]  
Hammersberg P, 1998, J Xray Sci Technol, V8, P75
[7]   CORRECTION FOR BEAM HARDENING IN COMPUTED-TOMOGRAPHY [J].
HERMAN, GT .
PHYSICS IN MEDICINE AND BIOLOGY, 1979, 24 (01) :81-106
[8]   Segmentation of X-ray computed tomography images of porous materials: A crucial step for characterization and quantitative analysis of pore structures [J].
Iassonov, Pavel ;
Gebrenegus, Thomas ;
Tuller, Markus .
WATER RESOURCES RESEARCH, 2009, 45
[9]   Data processing and image reconstruction methods for pixel detectors [J].
Jakubek, Jan .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 576 (01) :223-234
[10]   Empirical cupping correction: A first-order raw data precorrection for cone-beam computed tomography [J].
Kachelriess, M ;
Sourbelle, K ;
Kalender, WA .
MEDICAL PHYSICS, 2006, 33 (05) :1269-1274