Alternatives to patient specific verification measurements in proton therapy: a comparative experimental study with intentional errors

被引:34
作者
Matter, M. [1 ,2 ]
Nenoff, L. [1 ,2 ]
Meier, G. [1 ]
Weber, D. C. [1 ,3 ,4 ]
Lomax, A. J. [1 ,2 ]
Albertini, F. [1 ]
机构
[1] Paul Scherrer Inst, Ctr Proton Therapy, Villigen, Switzerland
[2] Swiss Fed Inst Technol, ETH Hoenggerberg, Dept Phys, Zurich, Switzerland
[3] Univ Hosp Zurich, Dept Radiat Oncol, CH-8091 Zurich, Switzerland
[4] Univ Hosp Bern, Dept Radiat Oncol, CH-3000 Bern, Switzerland
关键词
proton therapy; quality assurance; patient specific verification; pencil beam scanning; log file analysis; independent dose calculation; QUALITY-ASSURANCE; LOG FILES; PER-BEAM; TOOL;
D O I
10.1088/1361-6560/aae2f4
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Patient specific verification (PSV) measurements for pencil beam scanning (PBS) proton therapy are resource-consuming and necessitate substantial beam time outside of clinical hours. As such, efforts to safely reduce the PSV-bottleneck in the clinical work-flow are of great interest. Here, capabilities of current PSV methods to ensure the treatment integrity were investigated and compared to an alternative approach of reconstructing the dose distribution directly from the machine control- or delivery log files with the help of an independent dose calculation (IDC). Scenarios representing a wide range of delivery or work-flow failures were identified (e.g. error in spot position, air gap or pre-absorber setting) and machine files were altered accordingly. This yielded 21 corrupted treatment files, which were delivered and measured with our clinical PSV protocol. IDC machine-and log file checks were also conducted and their sensitivity at detecting the errors compared to the measurements. Although some of the failure scenarios induced clinically relevant dose deviations in the patient geometry, the PSV measurement protocol only detected one out of 21 error scenarios. However, 11 and all 21 error scenarios were detected using dose reconstructions based on the log and machine files respectively. Our data suggests that, although commonly used in particle therapy centers, PSV measurements do a poor job detecting data transfer failures and imperfect delivery machine performance. Machine- and log-file IDCs have been shown to successfully detect erroneous work-flows and to represent a reliable addition to the QA procedure, with the potential to replace PSV.
引用
收藏
页数:12
相关论文
共 31 条
[1]   A comprehensive and efficient daily quality assurance for PBS proton therapy [J].
Actis, O. ;
Meer, D. ;
Koenig, S. ;
Weber, D. C. ;
Mayor, A. .
PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (05) :1661-1675
[2]   Precise on-line position measurement for particle therapy [J].
Actis, O. ;
Meer, D. ;
Koenig, S. .
JOURNAL OF INSTRUMENTATION, 2014, 9
[3]   Polymer gel dosimetry [J].
Baldock, C. ;
De Deene, Y. ;
Doran, S. ;
Ibbott, G. ;
Jirasek, A. ;
Lepage, M. ;
McAuley, K. B. ;
Oldham, M. ;
Schreiner, L. J. .
PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (05) :R1-R63
[4]   Treatment log files as a tool to identify treatment plan sensitivity to inaccuracies in scanned proton beam delivery [J].
Belosi, Maria Francesca ;
van der Meer, Robert ;
de Acilu Laa, Paz Garcia ;
Bolsi, Alessandra ;
Weber, Damien C. ;
Lomax, Antony J. .
RADIOTHERAPY AND ONCOLOGY, 2017, 125 (03) :514-519
[5]   XFeasibility of online IMPT adaptation using fast, automatic and robust dose restoration [J].
Bernatowicz, Kinga ;
Geets, Xavier ;
Barragan, Ana ;
Janssens, Guillaume ;
Souris, Kevin ;
Sterpin, Edmond .
PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (08)
[6]   3D DVH-based metric analysis versus per-beam planar analysis in IMRT pretreatment verification [J].
Carrasco, Pablo ;
Jornet, Nuria ;
Latorre, Artur ;
Eudaldo, Teresa ;
Ruiz, Agusti ;
Ribas, Montserrat .
MEDICAL PHYSICS, 2012, 39 (08) :5040-5049
[7]  
Chowdhuri Z, 2016, PSI WATER COLUMN PHA
[8]   Quality Control Quantification (QCQ): A Tool to Measure the Value of Quality Control Checks in Radiation Oncology [J].
Ford, Eric C. ;
Terezakis, Stephanie ;
Souranis, Annette ;
Harris, Kendra ;
Gay, Hiram ;
Mutic, Sasa .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2012, 84 (03) :E263-E269
[9]   Patient-specific QA and delivery verification of scanned ion beam at NIRS-HIMAC [J].
Furukawa, Takuji ;
Inaniwa, Taku ;
Hara, Yousuke ;
Mizushima, Kota ;
Shirai, Toshiyuki ;
Noda, Koji .
MEDICAL PHYSICS, 2013, 40 (12)
[10]   Characterizing a proton beam scanning system for Monte Carlo dose calculation in patients [J].
Grassberger, C. ;
Lomax, Anthony ;
Paganetti, H. .
PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (02) :633-645