A Treatment Planning Comparison of Passive-Scattering and Intensity-Modulated Proton Therapy for Typical Tumor Sites

被引:44
作者
Kase, Yuki [1 ,2 ]
Yamashita, Haruo [1 ,2 ]
Fuji, Hiroshi [2 ]
Yamamoto, Yuichi [3 ]
Pu, Yuehu [3 ]
Tsukishima, Chihiro [3 ]
Murayama, Shigeyuki [1 ,2 ]
机构
[1] Shizuoka Canc Ctr Res Inst, Proton Therapy Div, Shizuoka 4118777, Japan
[2] Shizuoka Canc Ctr Hosp, Proton Therapy Div, Shizuoka 4118777, Japan
[3] Mitsubishi Electr Corp, Energy Syst Ctr, Hyogo Ku, Kobe, Hyogo 6528555, Japan
关键词
IMPT; Proton therapy; Beam scanning; Treatment planning; SCANNING SYSTEM; BEAM THERAPY; RADIOTHERAPY; RANGE; UNCERTAINTIES; SENSITIVITY; ALGORITHM; DESIGN; TISSUE; MOTION;
D O I
10.1269/jrr.11136
中图分类号
Q [生物科学];
学科分类号
07 ; 0710 ; 09 ;
摘要
Intensity-modulated proton therapy (IMPT) is expected to improve treatment results with fewer side effects than other proton therapies. The purpose of this study was to evaluate the tumor sites for which IMPT was effective under the same beam calculation conditions by planning IMPT for typical cases treated with passive scattering proton therapy (PSPT). We selected 16 cases of nasal cavity, lung, liver or prostate cancers as typical tumor sites receiving PSPT. The dose distributions and dose volume histograms optimized by the IMPT were compared with those optimized by the PSPT. We took particular note of the doses to the skin and organs at risk (OAR) when PSPT was replaced by IMPT. Furthermore, an improvement of the beam angles was also performed to obtain better dose distributions in the IMPT. The IMPT with the same beam angles resulted in near-maximum doses to the skin of average 78%, 64%, 84% and 99% of the PSPT doses for nasal cavity, lung, liver, and prostate cancers, respectively. However, it was difficult to improve the dose homogeneity of the target volume. The change of the IMPT beam angles could reduce the doses to OARs and skin in the case of the nasal cavity, while it had limited effect in the other cases. We concluded that IMPT was effective for reducing the doses to some OARs when treating nasal cavity, lung, liver and prostate cancers. The selection of beam angles was important in the IMPT optimization, especially for nasal cavity cancers.
引用
收藏
页码:272 / 280
页数:9
相关论文
共 28 条
[1]   Ridge filter design for proton therapy at Hyogo Ion Beam Medical Center [J].
Akagi, T ;
Higashi, A ;
Tsugami, H ;
Sakamoto, H ;
Masuda, Y ;
Hishikawa, Y .
PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (22) :N301-N312
[2]   CHEST WALL TOXICITY AFTER STEREOTACTIC BODY RADIOTHERAPY FOR MALIGNANT LESIONS OF THE LUNG AND LIVER [J].
Andolino, David L. ;
Forquer, Jeffrey A. ;
Henderson, Mark A. ;
Barriger, Robert B. ;
Shapiro, Ronald H. ;
Brabham, Jeffrey G. ;
Johnstone, Peter A. S. ;
Cardenes, Higina R. ;
Fakiris, Achilles J. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2011, 80 (03) :692-697
[3]   Comparison of second cancer risk due to out-of-field doses from 6-MV IMRT and proton therapy based on 6 pediatric patient treatment plans [J].
Athar, Basit S. ;
Paganetti, Harald .
RADIOTHERAPY AND ONCOLOGY, 2011, 98 (01) :87-92
[4]   A treatment planning comparison of 3D conformal therapy, intensity modulated photon therapy and proton therapy for treatment of advanced head and neck tumours [J].
Cozzi, L ;
Fogliata, A ;
Lomax, A ;
Bolsi, A .
RADIOTHERAPY AND ONCOLOGY, 2001, 61 (03) :287-297
[5]   Intrafractional Respiratory Motion for Charged Particle Lung Therapy with Immobilization Assessed by Four-Dimensional Computed Tomography [J].
Dobashi, Suguru ;
Sugane, Toshio ;
Mori, Shinichiro ;
Asakura, Hiroshi ;
Yamamoto, Naoyoshi ;
Kumagai, Motoki ;
Kandatsu, Susumu ;
Baba, Masayuki .
JOURNAL OF RADIATION RESEARCH, 2011, 52 (01) :96-102
[6]   TOLERANCE OF NORMAL TISSUE TO THERAPEUTIC IRRADIATION [J].
EMAMI, B ;
LYMAN, J ;
BROWN, A ;
COIA, L ;
GOITEIN, M ;
MUNZENRIDER, JE ;
SHANK, B ;
SOLIN, LJ ;
WESSON, M .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1991, 21 (01) :109-122
[7]   RESIDUAL MOTION AND DUTY TIME IN RESPIRATORY GATING RADIOTHERAPY USING INDIVIDUALIZED OR POPULATION-BASED WINDOWS [J].
Fuji, Hiroshi ;
Asada, Yoshihiro ;
Numano, Masumi ;
Yamashita, Haruo ;
Nishimura, Tetsuo ;
Hashimoto, Takayuki ;
Harada, Hideyuki ;
Asakura, Hirofumi ;
Murayama, Shigeyuki .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 75 (02) :564-570
[8]   Performance of the NIRS fast scanning system for heavy-ion radiotherapy [J].
Furukawa, Takuji ;
Inaniwa, Taku ;
Sato, Shinji ;
Shirai, Toshiyuki ;
Takei, Yuka ;
Takeshita, Eri ;
Mizushima, Kota ;
Iwata, Yoshiyuki ;
Himukai, Takeshi ;
Mori, Shinichiro ;
Fukuda, Shigekazu ;
Minohara, Shinichi ;
Takada, Eiichi ;
Murakami, Takeshi ;
Noda, Koji .
MEDICAL PHYSICS, 2010, 37 (11) :5672-5682
[9]   Nuclear interactions of 160 MeV protons stopping in copper: A test of Monte Carlo nuclear models [J].
Gottschalk, B ;
Platais, R ;
Paganetti, H .
MEDICAL PHYSICS, 1999, 26 (12) :2597-2601
[10]   Repeated proton beam therapy for hepatocellular carcinoma [J].
Hashimoto, T ;
Tokuuye, K ;
Fukumitsu, N ;
Igaki, H ;
Hata, M ;
Kagei, K ;
Sugahara, S ;
Ohara, K ;
Matsuzaki, Y ;
Akine, Y .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2006, 65 (01) :196-202