Kilovoltage Rotational External Beam Radiotherapy on a Breast Computed Tomography Platform: A Feasibility Study

被引:21
作者
Prionas, Nicolas D. [1 ]
McKenney, Sarah E. [1 ]
Stern, Robin L. [2 ]
Boone, John M. [1 ]
机构
[1] Univ Calif Davis, Med Ctr, Dept Radiol, Sacramento, CA 95817 USA
[2] Univ Calif Davis, Med Ctr, Dept Radiat Oncol, Sacramento, CA 95817 USA
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 2012年 / 84卷 / 02期
基金
美国国家卫生研究院;
关键词
Breast cancer; Rotational kilovoltage radiotherapy; Computed tomography; Monte Carlo simulation; CT;
D O I
10.1016/j.ijrobp.2011.12.042
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To demonstrate the feasibility of a dedicated breast computed tomography (bCT) platform to deliver rotational kilovoltage (kV) external beam radiotherapy (RT) for partial breast irradiation, whole breast irradiation, and dose painting. Methods and Materials: Rotational kV-external beam RT using the geometry of a prototype bCT platform was evaluated using a Monte Carlo simulator. A point source emitting 178 keV photons (approximating a 320-kVp spectrum with 4-mm copper filtration) was rotated around a 14-cm voxelized polyethylene disk (0.1 cm tall) or cylinder (9 cm tall) to simulate primary and primary plus scattered photon interactions, respectively. Simulations were also performed using voxelized bCT patient images. Beam collimation was varied in the x-y plane (1-14 cm) and in the z-direction (0.1-10 cm). Dose painting for multiple foci, line, and ring distributions was demonstrated using multiple rotations with varying beam collimation. Simulations using the scanner's native hardware (120 kVp filtered by 0.2-mm copper) were validated experimentally. Results: As the xey collimator was narrowed, the two-dimensional dose profiles shifted from a cupped profile with a high edge dose to an increasingly peaked central dose distribution with a sharp dose falloff. Using a 1-cm beam, the cylinder edge dose was <7% of the dose deposition at the cylinder center. Simulations using 120-kVp X-rays showed distributions similar to the experimental measurements. A homogeneous dose distribution (<2.5% dose fluctuation) with a 20% decrease in dose deposition at the cylinder edge (i.e., skin sparing) was demonstrated by weighted summation of four dose profiles using different collimation widths. Simulations using patient bCT images demonstrated the potential for treatment planning and image-guided RT. Conclusions: Rotational kV-external beam RT for partial breast irradiation, dose painting, and whole breast irradiation with skin sparing is feasible on a bCT platform with the potential for high-resolution image-guided RT. (C) 2012 Elsevier Inc.
引用
收藏
页码:533 / 539
页数:7
相关论文
共 16 条
[1]  
[Anonymous], 2005, XRAD 320 BIOL IRR TE
[2]   Accurate method for computer-generating tungsten anode x-ray spectra from 30 to 140 kV [J].
Boone, JM ;
Seibert, JA .
MEDICAL PHYSICS, 1997, 24 (11) :1661-1670
[3]   Computed tomography for imaging the breast [J].
Boone, John M. ;
Kwan, Alex L. C. ;
Yang, Kai ;
Burkett, George W. ;
Lindfors, Karen K. ;
Nelson, Thomas R. .
JOURNAL OF MAMMARY GLAND BIOLOGY AND NEOPLASIA, 2006, 11 (02) :103-111
[4]   Cone-beam volume CT breast imaging: Feasibility study [J].
Chen, B ;
Ning, R .
MEDICAL PHYSICS, 2002, 29 (05) :755-770
[5]   Image-guided radiotherapy: rationale, benefits, and limitations [J].
Dawson, Laura A. ;
Sharpe, Michael B. .
LANCET ONCOLOGY, 2006, 7 (10) :848-858
[6]  
HAMMERSTEIN GR, 1979, RADIOLOGY, V130, P485
[7]   The characterization of breast anatomical metrics using dedicated breast CT [J].
Huang, Shih-Ying ;
Boone, John M. ;
Yang, Kai ;
Packard, Nathan J. ;
McKenney, Sarah E. ;
Prionas, Nicolas D. ;
Lindfors, Karen K. ;
Yaffe, Martin J. .
MEDICAL PHYSICS, 2011, 38 (04) :2180-2191
[8]   A randomised trial of Supine versus Prone breast radiotherapy (SuPr study): Comparing set-up errors and respiratory motion [J].
Kirby, Anna M. ;
Evans, Philip M. ;
Helyer, Sarah J. ;
Donovan, Ellen M. ;
Convery, Helen M. ;
Yarnold, John R. .
RADIOTHERAPY AND ONCOLOGY, 2011, 100 (02) :221-226
[9]   Prone versus supine positioning for whole and partial-breast radiotherapy: A comparison of non-target tissue dosimetry [J].
Kirby, Anna M. ;
Evans, Philip M. ;
Donovan, Ellen M. ;
Convery, Helen M. ;
Haviland, Joanna S. ;
Yarnold, John R. .
RADIOTHERAPY AND ONCOLOGY, 2010, 96 (02) :178-184
[10]   Dedicated breast CT: Initial clinical experience [J].
Lindfors, Karen K. ;
Boone, John M. ;
Nelson, Thomas R. ;
Yang, Kai ;
Kwan, Alexander L. C. ;
Miller, DeWitt F. .
RADIOLOGY, 2008, 246 (03) :725-733