Development rationale for an articular surface replacement: a science-based evolution

被引:21
作者
Isaac, GH [1 ]
Siebel, T [1 ]
Schmalzried, TP [1 ]
Cobb, AG [1 ]
O'Sullivan, T [1 ]
Oakeshott, RD [1 ]
Flett, M [1 ]
Vail, TP [1 ]
机构
[1] DePuy Int Ltd, Leeds LS11 0EA, W Yorkshire, England
关键词
surface replacement; resurfacing; conservative hip replacement; metal-on-metal articulations;
D O I
10.1243/095441105X69132
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Hip resurfacing has an enduring appeal because of the advantages of bone conservation and maximal joint stability. However, a far from satisfactory experience with earlier resurfacing designs led to its virtual disappearance in the 1980s. The concept was reintroduced in the late 1990s. The current generation of resurfacing devices generally consisted of a large-diameter metal-on-metal articulation, the femoral components being cemented and the acetabular components utilizing various forms of cementless fixation. The encouraging medium-term results, with a follow-up of up to 8 years using the current generation of surface replacement joints, combined with favourable reports related to long-term performance of some metal bearings have led to a rapid increase in the use of such components with these devices. This trend is most marked in younger, more active patients who have expectations of restoration of lifestyle in addition to improved mobility and pain relief and in whom failure with conventional total hip replacement is much higher than previously reported with more sedentary patients. The aim of this paper is, firstly, to highlight a number of areas of improvement and, secondly, to explain how these may be addressed by making modifications to the design of both implants and instrumentation and to the surgical technique. The areas identified for improvement were tissue preservation (thinner components, and reduced steps between sizes), acetabular cup issues (fixation, insertion, and positioning), femoral component issues (design, loading, and cementation), improved bearing surface characteristics, and simplified precise instrumentation with a low-trauma surgical technique.
引用
收藏
页码:253 / 268
页数:16
相关论文
共 52 条
[1]  
Amstutz HC, 2004, ORTHOPEDICS, V27, P1266
[2]   Metal-on-metal hybrid surface arthroplasty:: Two to six-year follow-up study [J].
Amstutz, HC ;
Beaulé, PE ;
Dorey, FJ ;
Le Duff, MJ ;
Campbell, PA ;
Gruen, TA .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2004, 86A (01) :28-39
[3]   Fracture of the neck of the femur after surface arthroplasty of the hip [J].
Amstutz, HC ;
Campbell, PA ;
Le Duff, MJ .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2004, 86A (09) :1874-1877
[4]   Early results of primary Birmingham hip resurfacings - An independent prospective study of the first 230 hips [J].
Back, DL ;
Dalziel, R ;
Young, D ;
Shimmin, A .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2005, 87B (03) :324-329
[5]   Metal-on-metal surface arthroplasty with a cemented femoral component -: A 7-10 year follow-up study [J].
Beaulé, PE ;
Le Duff, M ;
Campbell, P ;
Dorey, FJ ;
Park, SH ;
Amstutz, HC .
JOURNAL OF ARTHROPLASTY, 2004, 19 (08) :17-22
[6]   EBRA-FCA for measurement of migration of the femoral component in surface arthroplasty of the hip [J].
Beaulé, PE ;
Krismer, M ;
Mayrhofer, P ;
Wanner, S ;
Le Duff, M ;
Mattesich, M ;
Stoeckl, B ;
Amstutz, HC ;
Biedermann, R .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2005, 87B (05) :741-744
[7]   SYSTEMIC EFFECTS OF BIOMATERIALS [J].
BLACK, J .
BIOMATERIALS, 1984, 5 (01) :11-18
[8]  
BOBYN JD, 1980, CLIN ORTHOP RELAT R, P263
[9]  
BOWSHER JG, 2003, T 49 ANN M ORTH RES, P1398
[10]   Cup inclination and serum concentration of cobalt and chromium after metal-on-metal total hip arthroplasty [J].
Brodner, W ;
Grübl, A ;
Jankovsky, R ;
Meisinger, V ;
Lehr, S ;
Gottsauner-Wolf, F .
JOURNAL OF ARTHROPLASTY, 2004, 19 (08) :66-70