Sequential detection of Salmonella typhimurium and Bacillus anthracis spores using magnetoelastic biosensors

被引:90
|
作者
Huang, S. [1 ]
Yang, H.
Lakshmanan, R. S.
Johnson, M. L.
Wan, J.
Chen, I. -H. [1 ]
Wikle, H. C., III
Petrenko, V. A. [2 ]
Barbaree, J. M. [1 ]
Chin, B. A.
机构
[1] Auburn Univ, Dept Bot & Microbiol, Auburn, AL 36849 USA
[2] Auburn Univ, Dept Pathobiol, Auburn, AL 36849 USA
来源
BIOSENSORS & BIOELECTRONICS | 2009年 / 24卷 / 06期
关键词
Multiple detection; Magnetoelastic; Biosensor; Phage; Salmonella typhimurium; Bacillus anthracis spores; PHAGE; WIRELESS;
D O I
10.1016/j.bios.2008.09.006
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Multiple phage-based magnetoelastic (ME) biosensors were simultaneously monitored for the detection of different biological pathogens that were sequentially introduced to the measurement system. The biosensors were formed by immobilizing phage and 1 mg/ml BSA (blocking agent) onto the magnetoelastic resonator's surface. The detection system included a reference sensor as a control, an E2 phage-coated sensor specific to S. typhimurium, and a JRB7 phage-coated sensor specific to B. anthracis spores. The sensors were free standing during the test, being held in place by a magnetic field. Upon sequential exposure to single pathogenic solutions. only the biosensor coated with the corresponding specific phage responded. As the cells/spores were captured by the specific phage-coated sensor. the mass of the sensor increased, resulting in a decrease in the sensor's resonance frequency. Additionally, non-specific binding was effectively eliminated by BSA blocking and was verified by the reference sensor, which showed no frequency shift. Scanning electron microscopy was used to visually verify the interaction of each biosensor with its target analyte. The results demonstrate that multiple magnetoelastic sensors may be simultaneously monitored to detect specifically targeted pathogenic species with good selectivity. This research is the first stage of an ongoing effort to simultaneously detect the presence of multiple pathogens in a complex analyte. (C) 2008 Elsevier B.V. All rights reserved.
引用
收藏
页码:1730 / 1736
页数:7
相关论文
共 50 条
  • [31] Phage immobilized magnetoelastic sensor for the detection of Salmonella typhimurium
    Lakshmanan, Ramji S.
    Guntupalli, Rajesh
    Hu, Jing
    Kim, Dong-Joo
    Petrenko, Valery A.
    Barbaree, James M.
    Chin, Bryan A.
    JOURNAL OF MICROBIOLOGICAL METHODS, 2007, 71 (01) : 55 - 60
  • [32] Rapid and Sensitive Detection of Salmonella Typhimurium on Eggshells by Using Wireless Biosensors
    Chai, Yating
    Li, Suiqiong
    Horikawa, Shin
    Park, Mi-Kyung
    Vodyanoy, Vitaly
    Chin, Bryan A.
    JOURNAL OF FOOD PROTECTION, 2012, 75 (04) : 631 - 636
  • [33] DETECTION OF SPORES OF BACILLUS-ANTHRACIS USING THE POLYMERASE CHAIN-REACTION
    CARL, M
    HAWKINS, R
    COULSON, N
    LOWE, J
    ROBERTSON, DL
    NELSON, WM
    TITBALL, RW
    WOODY, JN
    JOURNAL OF INFECTIOUS DISEASES, 1992, 165 (06): : 1145 - 1148
  • [34] Detection of Salmonella typhimurium using polyclonal antibody immobilized magnetostrictive biosensors
    Guntupalli, R.
    Hu, Jing
    Lakshmanan, Ramji S.
    Wan, Jiehui
    Huang, Shichu
    Yang, Hong
    Barbaree, James M.
    Huang, T. S.
    Chin, Bryan A.
    SENSORS, AND COMMAND, CONTROL, COMMUNICATIONS, AND INTELLIGENCE (C31)TECHNOLOGIES FOR HOMELAND SECURITY AND HOMELAND DEFENSE V, 2006, 6201
  • [35] A Multiple Magnetoelastic Sensor System for Detection of Salmonella Typhimurium Using Pulse Method
    Shen, W.
    Li, S.
    Horikawa, S.
    Chin, B.
    CHEMICAL SENSORS 9 -AND- MEMS/NEMS 9, 2010, 33 (08): : 77 - 84
  • [36] Rapid, enhanced detection of Salmonella Typhimurium on fresh spinach leaves using micron-scale, phage-coated magnetoelastic biosensors
    Horikawa, Shin
    Vaglenov, Kiril A.
    Gerken, Dana M.
    Chai, Yating
    Park, Mi-Kyung
    Li, Suiqiong
    Petrenko, Valery A.
    Chin, Bryan A.
    SENSING FOR AGRICULTURE AND FOOD QUALITY AND SAFETY IV, 2012, 8369
  • [37] A disclosure gel for visual detection of live Bacillus anthracis spores
    Robinson, C., V
    Bishop, A. H.
    JOURNAL OF APPLIED MICROBIOLOGY, 2019, 126 (06) : 1700 - 1707
  • [38] The effect of incubation time for Salmonella Typhimurium binding to phage-based magnetoelastic biosensors
    Park, Mi-Kyung
    Wikle, Howard C., III
    Chai, Yating
    Horikawa, Shin
    Shen, Wen
    Chin, Bryan A.
    FOOD CONTROL, 2012, 26 (02) : 539 - 545
  • [39] The effect of incubation temperature on the binding of Salmonella typhimurium to phage-based magnetoelastic biosensors
    Park, Mi-Kyung
    Oh, Jun-Hyun
    Chin, Bryan A.
    SENSORS AND ACTUATORS B-CHEMICAL, 2011, 160 (01): : 1427 - 1433
  • [40] Detection of Bacillus Anthracis Spores in Water Using a Phage-coated Piezoelectric Membrane
    Lu, X.
    Wang, L.
    Hou, L.
    Zhang, L.
    Chin, B. A.
    Cheng, Z-Y
    SELECTED PROCEEDINGS FROM THE 232ND ECS MEETING, 2017, 80 (10): : 1541 - 1547