Coil setup optimization for 2D-SENSE whole-heart coronary imaging

被引:10
作者
Buehrer, M
Huber, ME
Wiesinger, F
Boesiger, P
Kozerke, S
机构
[1] Univ Zurich, Inst Biomed Engn, CH-8092 Zurich, Switzerland
[2] Swiss Fed Inst Technol, CH-8092 Zurich, Switzerland
关键词
parallel imaging; 2D-SENSE; coil configuration; coronary imaging; whole-heart;
D O I
10.1002/mrm.20776
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Accelerated "whole-heart" coronary imaging with sensitivity encoding applied in both phase encoding directions (2D-SENSE) was investigated. In order to maximize the signal-to-noise ratio (SNR), coil configuration optimization was performed. To this end, 10 practical coil configurations each consisting of six standard coil elements were investigated and the local SNR was assessed by means of phantom experiments. Based on the experimental data, a symmetric configuration was found to yield the highest SNR. In a volunteer study, 2D-SENSE coronary images were obtained with total reduction factors of 3 and 4. Excellent depiction of the right and left coronary systems was possible in all cases.
引用
收藏
页码:460 / 464
页数:5
相关论文
共 11 条
[1]   Soap-bubble visualization and quantitative analysis of 3D coronary magnetic resonance angiograms [J].
Etienne, A ;
Botnar, RM ;
van Muiswinkel, AMC ;
Boesiger, P ;
Manning, WJ ;
Stuber, M .
MAGNETIC RESONANCE IN MEDICINE, 2002, 48 (04) :658-666
[2]   Prospective adaptive navigator correction for breath-hold MR coronary angiography [J].
McConnell, MV ;
Khasgiwala, VC ;
Savord, BJ ;
Chen, MH ;
Chuang, ML ;
Edelman, RR ;
Manning, WJ .
MAGNETIC RESONANCE IN MEDICINE, 1997, 37 (01) :148-152
[3]  
Pruessmann KP, 1999, MAGNET RESON MED, V42, P952, DOI 10.1002/(SICI)1522-2594(199911)42:5<952::AID-MRM16>3.0.CO
[4]  
2-S
[5]  
Scheidegger MB, 1996, HERZ, V21, P90
[6]   Navigator-echo-base real-time respiratory gating and triggering for reduction of respiration effects in three-dimensional coronary MR angiography [J].
Wang, Y ;
Rossman, PJ ;
Grimm, RC ;
Riederer, SJ ;
Ehman, RL .
RADIOLOGY, 1996, 198 (01) :55-60
[7]   Whole-heart steady-state free precession coronary artery magnetic resonance angiography [J].
Weber, OM ;
Martin, AJ ;
Higgins, CB .
MAGNETIC RESONANCE IN MEDICINE, 2003, 50 (06) :1223-1228
[8]   2D SENSE for faster 3D MRI [J].
Weiger, M ;
Pruessmann, KP ;
Boesiger, P .
MAGNETIC RESONANCE MATERIALS IN PHYSICS BIOLOGY AND MEDICINE, 2002, 14 (01) :10-19
[9]  
Weiger M, 2001, MAGNET RESON MED, V45, P495, DOI 10.1002/1522-2594(200103)45:3<495::AID-MRM1065>3.0.CO
[10]  
2-V