A Low-Cost, Normally Closed, Solenoid Valve for Non-Contact Dispensing in the Sub-μL Range

被引:20
作者
Bammesberger, Stefan Borja [1 ]
Kartmann, Sabrina [1 ]
Tanguy, Laurent [2 ]
Liang, Dong [2 ]
Mutschler, Klaus [1 ]
Ernst, Andreas [1 ,3 ]
Zengerle, Roland [1 ,2 ]
Koltay, Peter [1 ,3 ]
机构
[1] Univ Freiburg, Dept Microsyst Engn, IMTEK, D-79110 Freiburg, Germany
[2] HSG IMIT, Inst Mikro & Informat Tech, D-78052 Villingen Schwenningen, Germany
[3] Biofluidix GmbH, D-79110 Freiburg, Germany
关键词
non-contact dispensing; solenoid valve; normally closed; disposable;
D O I
10.3390/mi4010009
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
We present a disposable, normally closed, non-contact dispensing valve for the sub-mu L range. The miniaturized solenoid valve (diameter: 8 mm, height: 27.25 mm) is compatible to standard Luer-Lock interfaces. A highly dynamic actuation principle enables opening times down to 1 ms. The dispensing performance was evaluated for water (eta = 1.03 mPas) and a 66% (w/w) glycerol/water solution (eta = 16.98 mPas), at pressures varying from 200 to 800 mbar. The experimentally determined minimal dispensing volume was 163 nL (CV 1.6%) for water and 123 nL (CV 4.5%) for 66% (w/w) glycerol/water. The low-cost polymer valve enables high precision dispensing of liquid volumes down to the lower end of the sub-mu L range comparable to high-end non-disposable micro-dispensing valves.
引用
收藏
页码:9 / 21
页数:13
相关论文
共 13 条
[1]   Boosted PWM open loop control of hydraulic proportional valves [J].
Amirante, R. ;
Innone, A. ;
Catalano, L. A. .
ENERGY CONVERSION AND MANAGEMENT, 2008, 49 (08) :2225-2236
[2]  
[Anonymous], 2008, LM1949 INJ DRIV CONT
[3]  
Bammesberger S., 2013, DRUG DISCOV IN PRESS
[4]   Challenges and solutions to ultra-high-throughput screening assay miniaturization: submicroliter fluid handling [J].
Dunn, DA ;
Feygin, I .
DRUG DISCOVERY TODAY, 2000, 5 (12) :S84-S91
[5]  
Kong FW, 2009, 2008 IEEE INTERNATIONAL CONFERENCE ON ROBOTICS AND BIOMIMETICS, VOLS 1-4, P480
[6]  
LI D, 2008, ENCY MICROFLUIDICS N, DOI [10.1007/978-0-387-48998-8_1198, DOI 10.1007/978-0-387-48998-8_1198]
[7]   Novel gravimetric measurement technique for quantitative volume calibration in the sub-microliter range [J].
Liang, Dong ;
Steinert, Chris ;
Bammesberger, Stefan ;
Tanguy, Laurent ;
Ernst, Andreas ;
Zengerle, Roland ;
Koltay, Peter .
MEASUREMENT SCIENCE AND TECHNOLOGY, 2013, 24 (02)
[8]  
Mutschler K., MULTIPHYSICS N UNPUB
[9]   Piezo- and solenoid valve-based liquid dispensing for miniaturized assays [J].
Niles, WD ;
Coassin, PJ .
ASSAY AND DRUG DEVELOPMENT TECHNOLOGIES, 2005, 3 (02) :189-202
[10]  
Nordson EFD Optimum Datasheet, NORDS EFD OPT DAT