Evaluation of Stress Pattern Generated Through Various Thread Designs of Dental Implants Loaded in a Condition of Immediately After Placement and on Osseointegration-An FEA Study

被引:19
作者
Chowdhary, Ramesh [1 ]
Halldin, Anders [1 ]
Jimbo, Ryo [1 ]
Wennerberg, Ann [1 ]
机构
[1] Malmo Univ, Fac Odontol, Dept Prosthodont, SE-20506 Malmo, Sweden
关键词
dental implant; thread design; FINITE-ELEMENT-ANALYSIS; BONE;
D O I
10.1097/ID.0b013e31827daf55
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose: To determine the stress pattern generated through various thread design in experimental simulation models, when loaded immediately after placement and after osseointegration. Methods: Three-dimensional (3D) models were designed using CATIA, computer-aided design modeling software. The study was planned in 2 stages. Eight 2D models were constructed of different thread forms, one set with frictionless and other with bonded for bone to implant interface and loaded vertically with 100 N. In Stage II, 6 3D models of the different threads embedded in the cortical bone were constructed and loaded vertically and obliquely. Results: In 2D models, the von Mises stress concentrated at the crest in the bonded connection thread designs. The stress levels were in the range of 7 to 13 MPa. In the frictional implant bone interface, the thread designs had a clear effect on the stress levels in the bone. In the 3D analysis, the complete implant design affected the stress levels. Conclusions: The thread design affects the magnitude of the stress peak in the bone more effectively in immediately loaded (frictionless) implants than the osseointegrated (bonded) implants. Maximum stress was observed at the first thread in most of the osseointegrated implants. (Implant Dent 2013;22:91-96)
引用
收藏
页码:91 / 96
页数:6
相关论文
共 31 条
  • [1] ADELL R, 1990, International Journal of Oral and Maxillofacial Implants, V5, P347
  • [2] Arramon YP, 1994, T ORTHOP RES SOC, V19, P280
  • [3] Finite element analysis of non-axial versus axial loading of oral implants in the mandible of the dog
    Barbier, L
    Vander Sloten, J
    Krzesinski, G
    Schepers, E
    Van Der Perre, G
    [J]. JOURNAL OF ORAL REHABILITATION, 1998, 25 (11) : 847 - 858
  • [4] Bidez M W, 1992, J Oral Implantol, V18, P264
  • [5] Evaluation of load transfer characteristics of five different implants in compact bone at different load levels by finite elements analysis
    Bozkaya, D
    Muftu, S
    Muftu, A
    [J]. JOURNAL OF PROSTHETIC DENTISTRY, 2004, 92 (06) : 523 - 530
  • [6] Evaluation of design parameters of osseointegrated dental implants using finite element analysis
    Chun, HJ
    Cheong, SY
    Han, JH
    Heo, SJ
    Chung, JP
    Rhyu, IC
    Choi, YC
    Baik, HK
    Ku, Y
    Kim, MH
    [J]. JOURNAL OF ORAL REHABILITATION, 2002, 29 (06) : 565 - 574
  • [7] Clelland N L, 1991, Int J Oral Maxillofac Implants, V6, P391
  • [8] PERSPECTIVES - BONES MECHANICAL USAGE WINDOWS
    FROST, HM
    [J]. BONE AND MINERAL, 1992, 19 (03): : 257 - 271
  • [9] Finite element analysis of four thread-form configurations in a stepped screw implant
    Geng, JP
    Ma, QS
    Xu, W
    Tan, KBC
    Liu, GR
    [J]. JOURNAL OF ORAL REHABILITATION, 2004, 31 (03) : 233 - 239
  • [10] Application of finite element analysis in implant dentistry: A review of the literature
    Geng, JP
    Tan, KBC
    Liu, GR
    [J]. JOURNAL OF PROSTHETIC DENTISTRY, 2001, 85 (06) : 585 - 598