CT-ED conversion on a GE Lightspeed-RT scanner: influence of scanner settings

被引:24
作者
Ebert, M. A. [1 ,2 ]
Lambert, J. [3 ]
Greer, P. B. [3 ,4 ]
机构
[1] Sir Charles Gairdner Hosp, Dept Radiat Oncol, Nedlands, WA 6009, Australia
[2] Univ Western Australia, Sch Phys, Nedlands, WA 6009, Australia
[3] Univ Newcastle, Sch Math & Phys Sci, Newcastle, NSW 2308, Australia
[4] Calvary Mater Newcastle Hosp, Dept Radiat Oncol, Newcastle, NSW, Australia
关键词
electron density; mA; automatic tube current modulation; Computed tomography; radiotherapy;
D O I
10.1007/BF03178591
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The influence of tube voltage (kV) and current (mA) on the resulting relationship of computed tomography number to electron density (CT-ED) was investigated for a wide-bore GE scanner. The influence of kV and mA scan settings were examined in combination with a 16-bit image reconstruction algorithm made available via the scanner software and which allowed resolution of CT numbers for high density materials. By using titanium and stainless steel inserts in an electron density phantom, mA variation was found to have minimal impact on the CT-ED relationship, whereas variation in kV led to significant differences in CT number for the high density materials. The scanner is also equipped with automatic tube-current modulation capabilities. The influence of automatic tube-current modulation on CT number was investigated for a range of materials in a phantom geometry. It was found that tube current modulation has negligible effect on CT number, though the changing dimension of the phantom did influence CT number of an aluminium insert for scans undertaken with both fixed and modulated tube currents. In light of evidence from other studies examining the influence of CT number on dose calculation, it is recommended that scanner settings and specific CT-ED look-up tables be considered when calculations will be required with high-density materials present.
引用
收藏
页码:154 / 159
页数:6
相关论文
共 19 条
[1]   EFFECT OF A FEMORAL-HEAD PROSTHESIS ON MEGAVOLTAGE BEAM RADIOTHERAPY [J].
BIGGS, PJ ;
RUSSELL, MD .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1988, 14 (03) :581-586
[2]  
Carolan M, 2000, Australas Radiol, V44, P290, DOI 10.1046/j.1440-1673.2000.00816.x
[3]   Applications of simulator computed tomography number for photon dose calculations during radiotherapy treatment planning [J].
Chu, JCH ;
Ni, B ;
Kriz, R ;
Saxena, VA .
RADIOTHERAPY AND ONCOLOGY, 2000, 55 (01) :65-73
[4]   Calibration of CT Hounsfield units for radiotherapy treatment planning of patients with metallic hip prostheses: the use of the extended CT-scale [J].
Coolens, C ;
Childs, PJ .
PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (11) :1591-1603
[5]   Dose reduction in CT by anatomically adapted tube current modulation. 1. Simulation studies [J].
Gies, M ;
Kalender, WA ;
Wolf, H ;
Suess, C ;
Madsen, MT .
MEDICAL PHYSICS, 1999, 26 (11) :2235-2247
[6]   Dose reduction in computed tomography by attenuation-based on-line modulation of tube current: evaluation of six anatomical regions [J].
Greess, H ;
Wolf, H ;
Baum, U ;
Lell, M ;
Pirkl, M ;
Kalender, W ;
Bautz, WA .
EUROPEAN RADIOLOGY, 2000, 10 (02) :391-394
[7]   Dose reduction in CT by anatomically adapted tube current modulation. II. Phantom measurements [J].
Kalender, WA ;
Wolf, H ;
Suess, C .
MEDICAL PHYSICS, 1999, 26 (11) :2248-2253
[8]   Techniques and applications of automatic tube current modulation for CT [J].
Kalra, MK ;
Maher, MM ;
Toth, TL ;
Schmidt, B ;
Westerman, BL ;
Morgan, HT ;
Saini, S .
RADIOLOGY, 2004, 233 (03) :649-657
[9]   Radiation from "extra" images acquired with abdominal and/or pelvic CT: Effect of automatic tube current modulation [J].
Kalra, MK ;
Maher, MM ;
Toth, TL ;
Kamath, RS ;
Halpern, EF ;
Saini, S .
RADIOLOGY, 2004, 232 (02) :409-414
[10]   Comparison of z-axis automatic tube current modulation technique with fixed tube current CT scanning of abdomen and pelvis [J].
Kalra, MK ;
Maher, MM ;
Toth, TL ;
Kamath, RS ;
Halpern, EF ;
Saini, S .
RADIOLOGY, 2004, 232 (02) :347-353