Interlaboratory evaluation of an extraction and fluorescence method for the determination of trace beryllium in soils

被引:5
作者
Cronin, John P. [2 ]
Agrawal, Anoop [2 ]
Adams, Lori [2 ]
Tonazzi, Juan C. L. [2 ]
Brisson, Michael J. [3 ]
White, Kenneth T. [4 ]
Marlow, David [1 ]
Ashley, Kevin [1 ]
机构
[1] Natl Inst Occupat Safety & Hlth, Ctr Dis Control & Prevent, US Dept Hlth & Human Serv, Cincinnati, OH 45226 USA
[2] Berylliant Inc, Tucson, AZ 85712 USA
[3] Washington Savannah River Co, Aiken, SC 29808 USA
[4] Consult Serv, Virginia Beach, VA 23462 USA
来源
JOURNAL OF ENVIRONMENTAL MONITORING | 2008年 / 10卷 / 08期
关键词
D O I
10.1039/b804313b
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Analytical methods for the determination of trace beryllium in soils are needed so that anthropogenic sources of this element can be distinguished from native (background) levels of beryllium. In this work, a collaborative interlaboratory evaluation of a new extraction and fluorescence-based procedure for determining beryllium in soil samples was carried out to fulfil method validation requirements for ASTM International voluntary consensus standard test methods. A Canadian reference material, CCRMP Till-1 soil, with a background beryllium concentration of 2.4 mu g g(-1), was selected for study. This certified reference material (CRM) was spiked and homogenized with varying levels of beryllium oxide in order to give batches of material with beryllium concentrations of 4.36 +/- 0.69, 11.5 +/- 0.7, 124 +/- 7 and 246 +/- 16 mu g g(-1) (+/- values are standard deviations). In the interlaboratory study (ILS), which was carried out in accordance with an applicable ASTM International standard practice (ASTM E691), samples of these spiked soils were subjected to extraction in dilute ammonium bifluoride at similar to 90 degrees C for 40 h. Fluorescence measurement of the extracted beryllium was carried out via detection using the high quantum yield fluorophore, hydroxybenzoquinoline sulfonate (HBQS). Interlaboratory precision estimates from six participating laboratories ranged from 0.048 to 0.103 (relative standard deviations) for the five different beryllium concentrations. Pooled bias estimates resulting from this ILS were between -0.049 and 0.177 for the various beryllium levels. These figures of merit support promulgation of the analytical procedure as an ASTM International standard test method.
引用
收藏
页码:955 / 960
页数:6
相关论文
共 19 条
[1]   Extraction and optical fluorescence method for the measurement of trace beryllium in soils [J].
Agrawal, Anoop ;
Cronin, John P. ;
Agrawal, Akshay ;
Tonazzi, Juan C. L. ;
Adams, Lori ;
Ashley, Kevin ;
Brisson, Michael J. ;
Duran, Brandy ;
Whitney, Gary ;
Burrell, Anthony K. ;
McCleskey, T. Mark ;
Robbins, James ;
White, Kenneth T. .
ENVIRONMENTAL SCIENCE & TECHNOLOGY, 2008, 42 (06) :2066-2071
[2]   Validation of a standardized portable fluorescence method for determining trace beryllium in workplace air and wipe samples [J].
Agrawal, Anoop ;
Cronin, John ;
Tonazzi, Juan ;
McCleskey, T. Mark ;
Ehler, Deborah S. ;
Minogue, Edel M. ;
Whitney, Gary ;
Brink, Christopher ;
Burrell, Anthony K. ;
Warner, Benjamin ;
Goldcamp, Michael J. ;
Schlecht, Paul C. ;
Sonthalia, Prerna ;
Ashley, Kevin .
JOURNAL OF ENVIRONMENTAL MONITORING, 2006, 8 (06) :619-624
[3]   Ultrasonic extraction and portable anodic stripping voltammetric measurement of lead in paint, dust wipes, soil, and air: An interlaboratory evaluation [J].
Ashley, K ;
Song, RG ;
Esche, CA ;
Schlecht, PC ;
Baron, PA ;
Wise, TJ .
JOURNAL OF ENVIRONMENTAL MONITORING, 1999, 1 (05) :459-464
[4]  
ASHLEY K, 2005, J ASTM INT, V2, P8, DOI DOI 10.1520/JA113156
[5]  
ASHLEY K, 2006, ASTM STP, V1473
[6]   Ultra-trace determination of beryllium in occupational hygiene samples by ammonium bifluoride extraction and fluorescence detection using hydroxybenzoquinoline sulfonate [J].
Ashley, Kevin ;
Agrawal, Anoop ;
Cronin, John ;
Tonazzi, Juan ;
McCleskey, T. Mark ;
Burrell, Anthony K. ;
Ehler, Deborah S. .
ANALYTICA CHIMICA ACTA, 2007, 584 (02) :281-286
[7]  
*ASTM INT, 2005, E691 ASTM INT
[8]  
ASTM International, 2006, D7202 ASTM
[9]   Trace-level beryllium analysis in the laboratory and in the field: state of the art, challenges and opportunities [J].
Brisson, Michael J. ;
Ashley, Kevin ;
Stefaniak, Aleksandr B. ;
Ekechukwu, Amy A. ;
Creek, Kathryn L. .
JOURNAL OF ENVIRONMENTAL MONITORING, 2006, 8 (06) :605-611
[10]  
*CCRMP, 1995, CERT AN TILL 1 TILL