Early Studies of a Transmission Mechanism for MR-Guided Interventions

被引:0
|
作者
Zhao, Haoran [1 ]
Liu, Xin [2 ]
Zaid, Habib M. [2 ]
Shah, Dipan J. [3 ]
Heffernan, Michael J. [4 ]
Becker, Aaron T. [1 ]
Tsekos, Nikolaos V. [2 ]
机构
[1] Univ Houston, Dept Elect & Comp Engn, Houston, TX 77004 USA
[2] Univ Houston, Dept Comp Sci, Houston, TX 77204 USA
[3] Methodist DeBakey Heart & Vasc Ctr, Cardiovasc MRI Lab, Houston, TX USA
[4] GuidaBot LLC, Houston, TX USA
来源
2017 IEEE 17TH INTERNATIONAL CONFERENCE ON BIOINFORMATICS AND BIOENGINEERING (BIBE) | 2017年
基金
美国国家科学基金会;
关键词
Solid media transmission; MR compatible; Robotic Actuator; COMPATIBLE ROBOTIC SYSTEM; MAGNETIC-RESONANCE; BRACHYTHERAPY; MOTION;
D O I
10.1109/BIBE.2017.00082
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Magnetic resonance imaging (MRI)-guided, manipulator-assisted interventions have the potential to improve patient outcomes. This work presents a force transmission mechanism, called solid-media transmission (SMT), for actuating manipulators inside MRI scanners. The SMT mechanism is based on conduits filled with spheres and spacers made of a nonmagnetic, nonconductive material that forms a backbone for bidirectional transmission. Early modeling and experimental studies assessed SMT and identified limitations and improvements. Simulations demonstrated the detrimental role of friction, which can be alleviated with a choice of low friction material and long spacers. However, the length of spacer is limited by the desired bending of the conduit. A closed-loop control law was implemented to drive the SMT. The 3rd order system fit ratio is 92.3%. A 1-m long SMT was experimentally tested under this closed-loop controller with heuristically set parameters using a customized benchtop setup. For commanded displacements of 1 to 50 mm, the SMT-actuated 1 degree of freedom stage exhibited sub-millimeter accuracy, which ranged from 0.109 +/- 0.057 mm to 0.045 +/- 0.029 mm depending on the commanded displacement. However, such accuracy required long control times inversely proportional to displacement ranging from 7.56 +/- 1.85s to 2.53 +/- 0.11s. This was attributed to friction as well as backlash which is due to suboptimal packing of the media. In MR studies, a 4-m long SMT-actuated 1 DoF manipulator was powered by a servo motor located inside the scanner room but outside the 5 Gauss line of the magnet. With shielding and filtering, the SNR of MR images during the operation of the servo motor and SMT-actuation was found to be 89 +/- 9% of the control case.
引用
收藏
页码:450 / 456
页数:7
相关论文
共 50 条
  • [1] Robotic Systems for MR-guided Interventions
    Bock, M.
    Krafft, A.
    Maier, F.
    Umathum, R.
    Homagk, A. -K.
    Alt, S.
    Zhang, K.
    Pitsaer, C.
    Jenne, J.
    Semmler, W.
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, VOL 25, PT 6, 2009, 25 : 114 - +
  • [2] MR-Guided Interventions of the Prostate Gland a Literature Review
    Zangos, S.
    Eichler, K.
    Thalhammer, A.
    Mack, M. G.
    Marquardt, F.
    Vogl, T. J.
    ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2010, 182 (11): : 947 - 953
  • [3] Interactive MR-guided percutaneous nephrostomy
    Hagspiel, KD
    Kandarpa, K
    Silverman, SG
    JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1998, 8 (06): : 1319 - 1322
  • [4] MR-guided percutaneous interventional surgery robot system: A review
    Zhang, Weixi
    Yu, Jiaxing
    Lu, Xikun
    Xu, Yanjie
    INTERNATIONAL JOURNAL OF ADVANCED ROBOTIC SYSTEMS, 2025, 22 (01):
  • [5] MOLECULAR MR IMAGING AND MR-GUIDED ULTRASOUND THERAPIES IN CANCER
    Grenier, N.
    Quesson, B.
    de Senneville, B. Denis
    Trillaud, H.
    Couillaud, F.
    Moonen, C.
    JBR-BTR, 2009, 92 (01): : 8 - 12
  • [6] Real-Time Active MR-Tracking of Metallic Stylets in MR-Guided Radiation Therapy
    Wang, Wei
    Dumoulin, Charles L.
    Viswanathan, Akila N.
    Tse, Zion T. H.
    Mehrtash, Alireza
    Loew, Wolfgang
    Norton, Isaiah
    Tokuda, Junichi
    Seethamraju, Ravi T.
    Kapur, Tina
    Damato, Antonio L.
    Cormack, Robert A.
    Schmidt, Ehud J.
    MAGNETIC RESONANCE IN MEDICINE, 2015, 73 (05) : 1803 - 1811
  • [7] MR-GUIDED THERMAL ABLATION OF BRAIN-TUMORS
    JOLESZ, FA
    AMERICAN JOURNAL OF NEURORADIOLOGY, 1995, 16 (01) : 49 - 52
  • [8] Clinical implementation of MR-guided vaginal cylinder brachytherapy
    Owrangi, Amir M.
    Jolly, Shruti
    Balter, James M.
    Cao, Yue
    Maturen, Katherine E.
    Young, Lisa
    Zhu, Tong
    Prisciandaroa, Joann I.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2015, 16 (06): : 490 - 500
  • [9] SMART tracking: Simultaneous anatomical imaging and real-time passive device tracking for MR-guided interventions
    Zijlstra, Frank
    Viergever, Max A.
    Seevinck, Peter R.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2019, 64 : 252 - 260
  • [10] 3-T MR-guided brachytherapy for gynecologic malignancies
    Kapur, Tina
    Egger, Jan
    Damato, Antonio
    Schmidt, Ehud J.
    Viswanathan, Akila N.
    MAGNETIC RESONANCE IMAGING, 2012, 30 (09) : 1279 - 1290