Monitoring nociception during general anesthesia with cardiorespiratory coherence

被引:9
作者
Brouse, Chris J. [1 ]
Karlen, Walter [1 ]
Dumont, Guy A. [1 ]
Myers, Dorothy [2 ]
Cooke, Erin [2 ]
Stinson, Jonathan [2 ]
Lim, Joanne [2 ]
Ansermino, J. Mark [2 ]
机构
[1] Univ British Columbia, Dept Elect & Comp Engn, Vancouver, BC V5Z 1M9, Canada
[2] Univ British Columbia, Dept Anesthesiol Pharmacol & Therapeut, Vancouver, BC V5Z 1M9, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
Cardiorespiratory coherence; Heart rate variability; Respiratory sinus arrhythmia; Nociception; Antinociception; Analgesia; HEART-RATE-VARIABILITY; RESPONSES;
D O I
10.1007/s10877-013-9463-4
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
A novel wavelet transform cardiorespiratory coherence (WTCRC) algorithm has been developed to measure the autonomic state. WTCRC may be used as a nociception index, ranging from 0 (no nociception, strong coherence) to 100 (strong nociception, low coherence). The aim of this study is to estimate the sensitivity of the algorithm to nociception (dental dam insertions) and antinociception (bolus doses of anesthetic drugs). WTCRC's sensitivity is compared to mean heart rate (HRmean) and mean non-invasive blood pressure (NIBPmean), which are commonly used clinical signs. Data were collected from 48 children receiving general anesthesia during dental surgery. The times of dental dam insertion and anesthetic bolus events were noted in real-time during surgeries. 42 dental dam insertion and 57 anesthetic bolus events were analyzed. The change in average WTCRC, HRmean, and NIBPmean was calculated between a baseline period before each event and a response period after. A Wilcoxon rank-sum test was used to compare changes. Dental dam insertion changed the WTCRC nociception index by an average of 14 (82 %) [95 % CI from 7.4 to 19], HRmean by 7.3 beats/min (8.1 %) [5.6-9.6], and NIBPmean by 8.3 mmHg (12 %) [4.9-13]. A bolus dose of anesthetics changed the WTCRC by -15 (-50 %) [-21 to -9.3], HRmean by -4.8 beats/min (4.6 %) [-6.6 to -2.9], and NIBPmean by -2.6 mmHg (3.4 %) [-4.7 to -0.50]. A nociception index based on cardiorespiratory coherence is more sensitive to nociception and antinociception than are HRmean or NIBPmean. The WTCRC algorithm shows promise for noninvasively monitoring nociception during general anesthesia.
引用
收藏
页码:551 / 560
页数:10
相关论文
共 17 条
[1]   Accuracy of the 'Paedfusor' in children undergoing cardiac surgery or catheterization [J].
Absalom, A ;
Amutike, D ;
Lal, A ;
White, M ;
Kenny, GNC .
BRITISH JOURNAL OF ANAESTHESIA, 2003, 91 (04) :507-513
[2]   ECG beat detection using filter banks [J].
Afonso, VX ;
Tompkins, WJ ;
Nguyen, TQ ;
Luo, S .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1999, 46 (02) :192-202
[3]  
Bailón R, 2007, P ANN INT IEEE EMBS, P6675
[4]   AN EFFICIENT ALGORITHM FOR SPECTRAL-ANALYSIS OF HEART-RATE-VARIABILITY [J].
BERGER, RD ;
AKSELROD, S ;
GORDON, D ;
COHEN, RJ .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1986, 33 (09) :900-904
[5]  
Boudreau P, 2012, C P IEEE EN IN PRESS
[6]  
Brouse CJ, 2010, COMPUT CARDIOL CONF, V37, P713
[7]  
Brouse CJ, 2011, IEEE ENG MED BIO, P6114, DOI 10.1109/IEMBS.2011.6091510
[8]   Toward understanding respiratory sinus arrhythmia: Relations to cardiac vagal tone, evolution and biobehavioral functions [J].
Grossman, Paul ;
Taylor, Edwin W. .
BIOLOGICAL PSYCHOLOGY, 2007, 74 (02) :263-285
[9]   Effects of awareness and nociception on heart rate variability during general anaesthesia [J].
Huhle, R. ;
Burghardt, M. ;
Zaunseder, S. ;
Wessel, N. ;
Koch, T. ;
Malberg, H. ;
Heller, A. R. .
PHYSIOLOGICAL MEASUREMENT, 2012, 33 (02) :207-217
[10]  
Johnson J.O., 2009, CLIN ANESTH, P326